Unomedical Issues Worldwide Recall of Certain Manual Pulmonary Resuscitators
Date:10/6/2009

MCALLEN, Texas, Oct. 6 /PRNewswire/ -- Unomedical Inc., a manufacturer of medical devices, today announced that it is conducting a voluntary recall of certain units of the single-patient use Manual Pulmonary Resuscitator (MPR). The recall only impacts MPRs manufactured from July 2002 - March 2008 and matching the lot numbers listed on the following Unomedical web page: http://www.unomedical.com/?pageid=H3160. This recall is being conducted because of a potential malfunction of certain units of the MPR, which may impair the ability of the device to generate the positive pressure necessary to function properly. The occurrence of such a malfunction may create a situation in which the use of the product could potentially cause serious adverse health consequences or death. This recall does not impact any MPRs manufactured after March 2008.

(Photo: http://www.newscom.com/cgi-bin/prnh/20091006/NY88437 )

Unomedical is contacting customers to arrange for the return and credit of all MPR units subject to this recall by sending notification letters to distributors and customers. In addition, the company has set up a web page with a list of affected lot numbers, guidance (diagram and photos) to allow customers to identify products subject to the recall in the event that the customer has already removed the primary product packaging, and instructions on what actions to take.

In order to distinguish between the recalled product and unaffected product, customers may examine the patient valve housing immediately below the right-angle exhalation port, where the retention ring should be visible. MPR units with a clear or transparent ring, as well as those where no ring can be seen, should be returned to Unomedical as instructed. MPR units with a clearly visible blue retention ring are not affected and do not need to be returned.

The MPR is a single-patient use device used by healthcare professionals. It is intended for patients requiring total or intermittent ventilatory support.

Unomedical notified the U.S. Food & Drug Administration (FDA) of its decision to voluntarily recall the product on October 1, 2009.

Customers with questions are urged to contact Unomedical at 1-800-634-6003. Any adverse reactions experienced with the use of this product, and/or quality problems can also be reported to the FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch website at www.fda.gov/medwatch.

About Unomedical

Unomedical is a provider of medical devices with global operations. For more information, please visit www.unomedical.com.

    Media Contacts:

    Peter Duda or Hallie Bozzi
    Weber Shandwick
    212-445-8287
    unomedical@webershandwick.com
    9am - 6pm (EST)

    Customer Contacts:

    Americas
    Raymond Mendoza
    1-800-634-6003
    8:00am - 5:30pm Mon-Thurs, 8:00am - 1:00pm Fri (CST)

    Europe
    Betina Simonsen
    +45 4810 3036
    8:00am - 4pm Mon-Fri (CET)

    Asia
    Winnie Hui
    + 65 6593 1568
    9:00am - 5:00pm Mon-Fri (SGT)

    Russia
    Michael Martikhin
    +7 49 57487884
    8:30am - 5pm Mon-Fri (MSK)

SOURCE Unomedical Inc.


'/>"/>
SOURCE Unomedical Inc.
Copyright©2009 PR Newswire.
All rights reserved

Related medicine technology :

1. FDA Issues Complete Response Letter for Carisbamate
2. Amgen Issues Statement on Outcomes of Advisory Committee for Reproductive Health Drugs (ACRHD) Meeting
3. Society of Gynecologic Oncologists Issues White Paper on HPV Vaccines Impact on Cervical Cancer Prevention
4. Pennsylvania Agriculture Department Issues Quarantine for Allegheny County Horse Stable
5. FDA Issues Complete Response Letter for INTUNIV(TM) (guanfacine) Extended Release for the Treatment of ADHD in Children and Adolescents
6. Avon Protection Systems Issues Correction/Clarification on ST53 SCBA System
7. Pharmos Issues Business Update on Dextofisopam Trial and Financing
8. U.S. FDA Issues Complete Response Letter for SAPHRIS(TM) (ASENAPINE) in the Acute Treatment of Both Schizophrenia and Bipolar I Disorder
9. FDA Issues a Complete Response Letter for Lillys Olanzapine LAI for Treatment of Schizophrenia in Adults
10. Coalition for 21st Century Medicine Issues Statement in Opposition of Genentech Petition to Restrict Lab Use of Diagnostic Tests for Physicians and Patients
11. MatTeks EpiDerm In Vitro Tissues Headed for Validation in Europe
Post Your Comments:
*Name:
*Comment:
*Email:
(Date:4/20/2017)... April 20, 2017   ZappRx, Inc ., a digital ... prescribing process, today announced it closed $25 million in Series ... capital firm based in Seattle that ... Partners . The Series B round included participation from ... 2014, and GV (formerly Google Ventures). As part ...
(Date:4/20/2017)... -- Eyevensys, a private biotechnology company developing ... technology that enables the safe, local, sustained production of ... range of ophthalmic diseases, announces it has received approval ... (MHRA) to advance its technology into clinical development. ... The EyeCET platform ...
(Date:4/19/2017)... - CRH Medical Corporation (TSX: CRH) (NYSE MKT: CRHM) (the "Company"), announces that ... Conference 2017 at the Sheraton Hotel in Toronto, Ontario ... of the Company is scheduled to present on Tuesday, May 2 ... the Chairman of the Board, Tony Holler will also ... For more ...
Breaking Medicine Technology:
(Date:4/28/2017)... New York, NY (PRWEB) , ... April 28, ... ... Shiel Medical Laboratory to expanded distribution of the GlycoMark test throughout the Northeast ... indicates recent hyperglycemia in patients with diabetes. The GlycoMark test provides a clinically ...
(Date:4/28/2017)... ... April 28, 2017 , ... Ushio America proudly ... offer an instant energy-saving solution for F32T8 fluorescent lamps on most instant-start and ... These 50,000 hour rated lamps utilize the existing electronic ballast, saving labor and ...
(Date:4/28/2017)... ... April 28, 2017 , ... Datta Endoscopic Back Surgery and ... revolutionary endoscopic practice under Dr. Datta. Patel is a Board Certified Anesthesiologist, with ... revolutionary eDiscSculpt Technique created and used by Datta Endoscopic is unlike traditional back ...
(Date:4/28/2017)... (PRWEB) , ... April 28, 2017 , ... The Texas ... in recently to the labor and delivery team at Women’s Hospital at Renaissance in ... who give birth at the hospital and decide to donate. , “Women’s Hospital ...
(Date:4/28/2017)... , ... April 28, 2017 , ... ... hERG liability could substantially improve drug safety and minimize the cost of development. ... validating ion channel inhibition using cell lines and for cardiac toxicity using induced ...
Breaking Medicine News(10 mins):