Datascope Corp./MAQUET Issues Worldwide Voluntary Recall of the System CS100, CS100i and CS300 Intra-Aortic Balloon Pumps for Potential Electrical Test Failure Code
Date:6/16/2017

WAYNE, N.J., June 16, 2017 /PRNewswire/ -- Datascope Corp. is voluntarily performing a worldwide field correction of certain Intra-Aortic Balloon Pumps (IABPs) sold by Datascope Corp. for a potential electrical test failure code.    

AFFECTED PRODUCT

PART NUMBER

CS100i IABP
CS100 IABP
CS300 IABP

0998-UC-0446HXX; 0998-UC-0479HXX
0998-00-3013-XX;  0998-UC-3013-XX
0998-00-3023-XX;  0998-UC-3023-XX

This field correction also applies to any System 98 or System 98XT IABP that was converted to a CS100i or CS300 IABP.

Overall Action for User
Patients receiving IABP therapy are in critical condition and sudden interruption of therapy could result in unsafe, hemodynamic instability. Maquet/Getinge indicates that the risk-benefit of using an affected CS100i, CS100 or CS300 IABP should be assessed by the medical team for each patient when no alternative IABP or alternative therapy is available. Please adhere to the following instructions when using affected devices:

1) Pursuant to the USER Instruction WARNINGS, clinicians are instructed not to leave the patient unattended during IABP therapy.
2) An additional hazard associated with a sudden shutdown is related to the static condition (no inflating or deflating) of the balloon during the interruption of therapy. It is important to note the following WARNING in the CS100i, CS100 or CS300 IABP Operating Instructions Manual:

WARNING: The patient balloon should not remain inactive in the patient (i.e., not inflating or deflating) for more than 30 minutes due to the potential for thrombus formation.

3) Until the service is performed, we recommend powering on the IABP prior to inserting the IAB catheter to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the event the IABP fails to successfully complete the self-test and exhibits electrical test failure code 58, please remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office.

Datascope has received a complaint that has been associated with a patient death due to the failure of the device to initiate therapy. This complaint involved a CS300 IABP that did not pump due to an electrical test failure code #58 (power up vent tests fail), maintenance code #3, and an autofill failure. An electrical test failure code #58 is caused by a solenoid valve requiring more power than the solenoid driver board can deliver to open the valve.

There are approximately 12,000 affected units sold globally. The affected IABP units were distributed in the U.S. and worldwide (in over 100 countries). Affected units were distributed between March 23, 2003, and December 11, 2013. Units distributed after December 11, 2013, are not affected by the field correction.

A service representative from Datascope will be replacing the defective solenoid driver boards.  Customers having affected IABP unit(s) will be contacted by a representative of the Maquet/Getinge Service Team to schedule on-site service.

For additional information regarding this field correction, please contact the Customer Service Department at 1-888-627-8383 and Press 2 (Monday through Friday from 8:00 a.m. to 6:00 p.m. EDT).

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

About Getinge
Getinge is a global provider of innovative solutions for operating rooms, intensive-care units, sterilization departments and for life science companies and institutions. Based on our first-hand experience and close partnerships with clinical experts, healthcare professionals and medtech specialists, we are improving the every-day life for people, today and tomorrow.

 

To view the original version on PR Newswire, visit:http://www.prnewswire.com/news-releases/datascope-corpmaquet-issues-worldwide-voluntary-recall-of-the-system-cs100-cs100i-and-cs300-intra-aortic-balloon-pumps-for-potential-electrical-test-failure-code-300475511.html


'/>"/>
SOURCE Getinge
Copyright©2017 PR Newswire.
All rights reserved


Related medicine technology :

1. Omron Healthcare Issues Reminder for Stroke Awareness Month: Monitor Your Blood Pressure
2. U.S. FDA Issues Complete Response Letter for Baricitinib
3. Research and Markets - Raw Material Requirements (Health Canada/USP/EP) in a cGMP Environment: Issues and Solutions: 2-day In-person Seminar (San Francisco, CA, United States - October 5-6, 2017)
4. Edison Issues Update on Quantum Genomics (ALQGC)
5. Edison Issues Update on Immunovia (IMMUNOV)
6. Edison Issues ADR Update on Prima BioMed (PBMD)
7. Edison Issues Outlook on ReNeuron Group (RENE)
8. Medical Device Safety Issues Prompt Regulator Inspections
9. Biocepts CEO Issues Letter to Stockholders
10. USPTO Issues 2nd Patent to ImmersiveTouch Inc. on Its Surgical VR Technology
11. Luminex Announces Preliminary Revenue for Fourth Quarter 2016 and Issues 2017 Revenue Guidance; To Present at 35th Annual J.P. Morgan Healthcare Conference
Post Your Comments:
*Name:
*Comment:
*Email:
(Date:9/25/2017)... Sept. 25, 2017  EpiVax, Inc., a leader ... and immune-engineering today announced the launch of EpiVax ... of personalized therapeutic cancer vaccines. EpiVax has provided ... access to enabling technologies to the new precision ... lead EpiVax Oncology as Chief Executive Officer. Gad ...
(Date:9/19/2017)... 2017   ZirMed Inc ., a recognized leader in ... it has been ranked #1 by its users for the ... 2017 User Survey. ZirMed was recognized as the top-ranked end-to-end ... medical centers over 200 beds and holds one of the ... survey history. ...
(Date:9/12/2017)... , Sept. 12, 2017  Consumer reviews on the independent review ... as the number one company for hearing aids, ranking it higher ... other brands. ... Named #1 by Consumers For Hearing Aids ... Embrace Hearing is an online store that provides high performance, state-of-the-art, ...
Breaking Medicine Technology:
(Date:10/13/2017)... ... ... Talented host, actor Rob Lowe, is introducing a segment ... of "Success Files," which is an award-winning educational program broadcasted on PBS Member ... with passion and integrity. , Sciatica occurs when the sciatic nerve in the ...
(Date:10/13/2017)... (PRWEB) , ... October 13, 2017 , ... “America On ... Christian identity. “America On The Brink” is the creation of published author, William ... several great-grandchildren. As a WWII veteran, he spent thirty years in the Navy. ...
(Date:10/12/2017)... LA (PRWEB) , ... October 12, 2017 , ... Planet ... in the U.S., announced today its plans to open a flagship location in Covington, ... occupy the former Rooms To Go store next to Office Depot in the Holiday ...
(Date:10/12/2017)... ... ... The American College of Medical Informatics (ACMI) will present the 2017 Morris F. ... AMIA’s Annual Symposium in Washington, D.C. AMIA’s Annual Symposium is taking place ... the field of medical informatics, this prestigious award is presented to an individual whose ...
(Date:10/12/2017)... ANGELES (PRWEB) , ... October 12, 2017 , ... ... Parsa Mohebi Hair Restoration, has recently contributed a medical article to the newly ... on cosmetictown.com. Dr. Mohebi’s article spotlights the hair transplant procedure known as ...
Breaking Medicine News(10 mins):