Abbott Diabetes Care Announces Recall of Certain Lots of Precision Xtra®, Precision Xceed Pro®, MediSense® Optium™, Optium™, OptiumEZ and ReliOn® Ultima Blood Glucose Test Strips in the United States and Puerto Rico

ALAMEDA, Calif., Dec. 22, 2010 /PRNewswire/ -- Abbott Diabetes Care today announced that it has initiated a recall of 359 lots (approximately 359 million strips) of Precision Xtra®, Precision Xceed Pro®, MediSense® Optium™, Optium™, OptiumEZ and ReliOn® Ultima Blood Glucose Test Strips in the United States and Puerto Rico.

The test strips are used with Abbott's Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium and OptiumEZ blood glucose monitoring systems. ReliOn Ultima test strips are used with the ReliOn Ultima blood glucose monitoring system. The blood glucose monitors are not being recalled and customers can continue to use the blood glucose monitors.

The affected test strips may give falsely low blood glucose results, which can lead users to try to raise their blood glucose when it is unnecessary and to fail to treat elevated blood glucose due to a falsely low reading. The problem appears to be related to longer than expected blood fill times. This may be dependent on the age of the strips and if the strips have been stored in (or exposed to) higher temperatures (above 72˚F and not to exceed 86˚F) for an extended period of time.

Customers who have test strips from these affected lots should discontinue use of the product. Abbott Diabetes Care will replace affected test strips at no charge. Abbott Diabetes Care initiated this recall following a routine internal quality review that indicated certain lots of Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima blood glucose test strips may experience longer than normal blood fill times which may cause falsely low blood glucose readings.

Customers are advised to contact their health care provider to determine testing options if they do not have any unaffected test strips to use.

However, if consumers must use affected test strips while in the process of obtaining new strips, they should:

  • Check the amount of time it takes for their blood glucose meter to start the "countdown" after they first apply blood to the test strip. They should start timing immediately after blood first makes contact with the test strip.  If the meter takes longer than five seconds to start the countdown, that test strip is affected and the result should not be used.  Users should check the time for each test strip they use because all of the strips in a package may not be affected to the same degree.  If any reading appears lower than they would expect or does not seem to correlate with the way they are feeling, the user should contact their health care provider immediately.
  • If they do not immediately have access to unaffected strips and are unable to test their blood sugar, users need to be aware of symptoms of high blood sugar (hyperglycemia) and low blood sugar (hypoglycemia).
  • Symptoms of high blood sugar include excessive thirst, excessive urination, blurred vision, weakness, nausea, vomiting, and abdominal pain. If a user is experiencing any of these symptoms or is not feeling well, they should contact their health care professional immediately.
  • Symptoms of low blood sugar may include trembling, excessive sweating, weakness, hunger, confusion, and headache. Some individuals may have no symptoms at all before they develop unconsciousness or seizures. It is important to treat low blood sugars promptly to avoid loss of consciousness or a seizure.  If a user is unable to obtain unaffected strips, they should contact their health care provider for advice on how to treat these symptoms before they occur.  

  • The recommendation is that hospitals and user facilities stop using the recalled test strips and follow the steps below:

    1. Use test strips from unaffected lots.
    2. If the health care facility does not have any test strips from unaffected lots but has an alternative point-of-care blood glucose testing system, begin using the alternative system until new, unaffected test strips can be obtained.

    If the health care facility does not have any test strips from unaffected lots or immediate access to an alternative point-of-care blood glucose testing system, health care providers should take the following steps:

  • Verify any critical glucose test results (e.g., test results that may be used to adjust insulin therapy in vulnerable patient populations) generated on the Precision Xceed Pro Blood Glucose Monitoring System using a central laboratory blood glucose method.  Clinical judgment should be applied when deciding whether to act on results prior to verification.
  • Verify any Precision Xceed Pro Blood Glucose Monitoring System results that do not match a patient's symptoms, or seem unexpected for the patient's clinical status, using a central laboratory blood glucose method.
  • Monitor the amount of time it takes for the Precision Xceed Blood Glucose Meter to start the "countdown" after blood is first applied to the test strip.  If the amount of time exceeds five seconds, discard that test strip immediately to prevent the risk of receiving a falsely low blood glucose result (if this occurs, health care providers should note the lot number of that test strip, discard all the test strips from that lot, and notify Abbott Diabetes Care immediately).

  • Customers can check if they have test strips from the affected lots by visiting and looking up their product lot number.

    Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima customers can also get information by calling Abbott Diabetes Care customer service at 1-800-448-5234 (English) and 1-800-709-7010 (Espanol).

    The website and customer service phone numbers listed above will also provide customers with information about returning affected test strips.

    Unaffected Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima Blood Glucose Test Strip lots may be used by customers to test their blood glucose as directed by their health care providers.

    Abbott Diabetes Care has reported this situation to the U.S. Food and Drug Administration and is working to make the necessary corrective actions to prevent this situation from recurring.

    URGENT PRODUCT RECALL – DISCONTINUE USE AND RETURNAFFECTED PRODUCT LOTSOptiumEZ Blood Glucose Test Strips45358, 45369, 45377, 45504, 45677, 45678, 45681, 45789, 45841, 45848, 45849, 45999, 46000, 46008, 46061, 46067, 46102, 46104, 46117, 46159, 46179, 46201, 46202, 46211, 46230, 46247, 46311, 46312, 46341, 46368, 46369, 46371, 46395, 46399, 46401, 46436ReliOn Ultima Blood Glucose Test Strips45358, 45369, 45377, 45379, 45466, 45504, 45511, 45608, 45613, 45641, 45642, 45644, 45672, 45674, 45675, 45676, 45677, 45721, 45734, 45736, 45739, 45742, 45743, 45789, 45790, 45800, 45801, 45840, 45841, 45849, 45850, 46008, 46009, 46061, 46067, 46069, 46072, 46103, 46105, 46148, 46201, 46202, 46214, 46215, 46232, 46313, 46332, 46336, 46338, 46341, 46368MediSense Optium Blood Glucose Test Strips45001A159, 45001A195, 45001A228, 45001A233, 45001A236, 45001A243, 45001A247, 45001A249,

    45001A250, 45001A255, 45001A279, 45001A285, 45001A295, 45001A298, 45001A306, 45001A317,

    45001A318, 45001A327, 45001A346, 45001A398, 45001A449, 45001A469, 45001A578, 45001A693,

    45001A697, 45001A747, 45001A790, 45001A795, 45001A839, 45001A917Optium Blood Glucose Test Strips45001A484, 45001A486, 45001A512, 45001A514, 45001A517, 45001A519, 45001A527, 45001A541,

    45001A544, 45001A552, 45001A558, 45001A564, 45001A572, 45001A598, 45001A665, 45001A677,

    45001A709, 45001A713, 45001A732, 45001A771, 45001A819, 45001A85245455, 45516, 45708, 45709, 45737, 45738,  45777, 45778, 45783Precision Xceed Pro Blood Glucose Test Strips44535H, 44545H, 44565H, 445A5H, 44645H, 44665H, 446E5H, 44725H, 44755H, 44925H, 44945H, 44945H, 44975H, 44995H, 449A5H, 44A15H, 44A35H, 44A75H, 44A85H, 44C65H, 44C75H, 44CE5H, 44CH5H, 44CP5H, 44CR5H, 45025H, 45035H, 450A5H, 450C5HPrecision Point of Care Blood Glucose Test Strips45515, 45517, 45701, 45791, 45963, 45964, 46367Precision Xtra Blood Glucose Test Strips45001A330, 45001A467, 45001A481, 45001A482, 45001A483, 45001A487, 45001A488, 45001A496, 45001A497, 45001A499, 45001A500, 45001A503, 45001A505, 45001A506, 45001A510, 45001A511, 45001A513, 45001A519, 45001A520, 45001A521, 45001A530, 45001A532, 45001A534, 45001A536, 45001A537, 45001A545, 45001A556, 45001A565, 45001A566, 45001A568, 45001A573, 45001A577, 45001A584, 45001A594, 45001A595, 45001A597, 45001A605, 45001A607, 45001A608, 45001A609, 45001A613, 45001A614, 45001A626, 45001A634, 45001A637, 45001A641, 45001A644, 45001A645, 45001A656, 45001A660, 45001A664, 45001A669, 45001A670, 45001A671, 45001A677, 45001A678, 45001A682, 45001A685, 45001A694, 45001A700, 45001A705, 45001A706, 45001A707, 45001A709, 45001A712, 45001A730, 45001A732, 45001A739, 45001A742, 45001A744, 45001A751, 45001A753, 45001A757, 45001A762, 45001A777, 45001A778, 45001A786, 45001A788, 45001A790, 45001A798, 45001A806, 45001A814, 45001A815, 45001A816, 45001A820, 45001A840, 45001A842, 45001A847, 45001A852, 45001A856, 45001A860, 45001A861, 45001A872, 45001A876, 45001A882, 45001A910, 45001A912, 45001A914, 45001A918, 45001A920, 45001A921, 45001A940, 45001A947, 45001A948, 45001A949, 45001A971, 45001A976, 45001A979, 45001A982, 45001A983, 45001C008, 45001C011, 45001C01245384, 45388, 45396, 45407, 45418, 45455, 45456, 45463, 45464, 45512, 45514, 45516, 45521, 45611, 45612, 45614, 45619, 45627, 45632, 45639, 45645, 45646, 45653, 45670, 45679, 45680, 45682, 45707, 45708, 45724, 45729, 45731, 45732, 45733, 45735, 45737, 45738, 45755, 45777, 45782, 45783, 45784Precision Xtra Blood Glucose & Ketone Monitoring System01P107D, 01P128D, 01P139D, 02P140D, 01P153D, 02P201D, 02P223D, 02P229D, 03P107D, 03P155D, 03P158D, 03P194D, 03P209D, 03P222D, 03P223D, 03P242D, 04P154D, 04P194D, 04P201D, 04P315D, 05P106D, 05P107D, 05P127D, 05P130D, 05P138D, 05P139D, 05P140D, 05P152D, 05P201D, 06P180D, 06P322D, 07P141D, 08P106D, 08P138D, 08P279D, 09P106D, 09P140DPrecision G3b Smartblue Blood Glucose Test Strips45001A710MediSense Optium Blood Glucose Electrodes45001A133, 45001A137, 45001A226, 45001A252, 45001A257, 45001A277, 45001A285, 45001A817, 45001A835, 45001A921, 45001C001Abbott Diabetes Care Customer Service(800) 448-5234 (Eng)(800) 709-7010 (Esp)

    SOURCE Abbott
    Copyright©2010 PR Newswire.
    All rights reserved

    Related medicine technology :

    1. Abbott Named One of Americas Most Generous Companies, Sets Company Records for Employee Giving
    2. Abbott Remains on the Dow Jones Sustainability Indexes for the Sixth Consecutive Year in Recognition of Leading Corporate Citizenship Performance
    3. Abbott Joins Clinton Health Access Initiative in Effort to Improve Infant Testing for HIV in Africa
    4. Abbott and Enanta Present Positive Initial Results from Phase 2 Study of ABT-450/r for Treatment of Hepatitis C
    5. Abbott Again Recognized for Environmental Initiatives in Illinois
    6. Abbott Statement: Abbott Provides Support for Cholera Relief Efforts in Haiti
    7. Abbott to Voluntarily Withdraw Meridia® (Sibutramine) in the U.S.
    8. Abbott Laboratories Agrees to Withdraw Its Obesity Drug Meridia
    9. Abbott Named One of the Top Employers in the Biotech and Pharmaceutical Industry by Science Magazine
    10. Abbott Receives FDA Market Clearance for Next-Generation Contact Lens Disinfecting Solution
    11. Abbotts XIENCE V® Shows Continued Strong Performance in Key Safety and Efficacy Measures at Two Years in Investigator-Initiated COMPARE Trial
    Post Your Comments:
    (Date:10/9/2015)...  Eli Lilly and Company (NYSE: LLY ) ... rights to Locemia,s intranasal glucagon, a potential treatment for ... Intranasal glucagon, which is currently in Phase III clinical ... for severe hypoglycemia. --> ... nasal powder formulation that is delivered in an emergency ...
    (Date:10/9/2015)... CHAPEL HILL, N.C. , Oct. 9, 2015 ... Pharmaceutical or Medical Device environments, corporate communications is ... external communications, corporations and professionals must continuously adjust ... the effectiveness of these adjustments is difficult, but ... to today,s communication options. ...
    (Date:10/9/2015)... , Oct. 9, 2015 Regulatory affairs groups ... --> --> These specialized groups ... the plethora of global regulations pertaining to the development and ... role, regulatory groups largely rely on their own internal effectiveness ... --> --> ...
    Breaking Medicine Technology:
    ... Minn., May 4, 2011 American Medical Systems® ... devices and therapies for male and female pelvic health, ... Ministry of Health, Labor and Welfare (MHLW) to market ... the treatment of Benign Prostatic Hyperplasia (BPH), or enlarged ...
    ... 4, 2011 The most recent release of Via ... decision support for advance care planning to encourage early ... treatment intent and likely outcomes. Recent ... effect that attention to planning end of life care ...
    Cached Medicine Technology:
    (Date:10/10/2015)... Clarkston, Metamora, Michigan (PRWEB) , ... October 10, 2015 , ... ... Up to 10% of cancer patients have an inherited gene mutation. Fortunately, knowing family ... stage. Women with certain genetic mutations have:, · up to 44% lifetime risk of ...
    (Date:10/10/2015)... ... October 10, 2015 , ... Adenomyosis is ... uterus, the endometrium, grows inside the muscle layer of the uterus. It's a ... of adenomyosis are similar to the symptoms of endometriosis and may include pelvic ...
    (Date:10/10/2015)... Chicago, IL (PRWEB) , ... October 10, 2015 , ... With the winter season coming ... manufacturer of low-watt heaters, Cozy Products is familiar with the problem and discusses ... a few degrees to the inside of the chicken coop in the winter, utilize a ...
    (Date:10/10/2015)... ... ... have to walk with crutches for the rest of my life," said an inventor from ... that would give me more independence and freedom would be amazing. This inspired me to ... developed the STRAP IN to enable an individual to walk with crutches without having to ...
    (Date:10/10/2015)... ... , ... The sixth annual cost of cybercrime study the Ponemon Institute released ... companies. The New York Times broke down some of those numbers in this ... among the companies surveyed, it only proves that regular threat assessments, penetration testing and ...
    Breaking Medicine News(10 mins):
    ... Sales and Marketing Veteran to Drive International Retail Expansion, ... Group and New WrinkleFree Brand Line ExtensionIRVINE, Calif., Jan. ... developer and distributor of medically developed and efficacy-based skin ... mass market retail locations throughout the U.S., announced today ...
    ... The following statement can be attributed to Michael ... has consistently said two things: We are not ... But we have urged public policy makers to ... merger without conditions, because without conditions the merger ...
    ... (Nasdaq: OMCL ), a leading provider of system solutions to acute healthcare ... quarter 2008 financial results. , What: Omnicell ... webcast, When: January ... Lipps, chairman, president and chief executive officer, ...
    ... Maine, Jan. 21 IDEXX Laboratories, Inc. (Nasdaq: ... quarter and full-year financial results for Friday, January 30, at ... conference call beginning at 9:00 a.m. (eastern) on that day. ... through a link on the IDEXX Web site, ...
    ... can lead to failure, , , WEDNESDAY, Jan. 21 (HealthDay ... a common cause of kidney transplant failure may help ... polyomavirus nephropathy, which affects about 9 percent of kidney ... adults but can cause serious problems for people with ...
    ... CAMBRIDGE, Mass. -- Schizophrenia may blur the boundary between ... that is involved in self-reflection, and thus causing an ... brain imaging study has found. , The traditional ... and emotions that characterize the disease are caused by ...
    Cached Medicine News:
    Disposable tracheostomy care tray....
    Disposable tracheostomy care tray....
    ... the Right Solution For Your Tracheostomy Care ... care tray features the latest components for ... infection control task. Designed by clinicians, these ... of each patient and caregiver. All of ...
    ... For Your Tracheostomy Care Needs, ,Each ... the latest components for efficient and convenient ... Designed by clinicians, these latex-free trays provide ... and caregiver. All of our trays also ...
    Medicine Products: