The atmospheric pressure photoionization source, PhotoSpray ion source (Robb, Covey and Bruins1) uses photons to ionize large quantities of a dopant molecule added along with the vaporized mobile phase. Analyte molecules are efficiently ionized through secondary reactions initiated by the charged dopant. This new technique was used to develop an analytical method for polycyclic aromatic hydrocarbons (PAHs), comparing the difference under reverse phase and normal phase chromatographic conditions. A total of 16 PAHs were analysed including the determination of unknown quantities in atmospheric samples from both rural and urban areas.
Introduction
Polycyclic aromatic hydrocarbons (PAHs) are a group of over 100 different chemicals, typically formed during incomplete combustion, which are present in a variety of environmental matrices. Airborne PAHs, usually found in aerosol particles, are implicated in heterogeneous atmospheric processing and as important indicators of gas-phase atmospheric processes. Certain carcinogenic PAHs may be carried by ultrafine particles in ambient air into the lungs, and are often the targets of rigorous environmental monitoring. Currently, environmental PAHs are most commonly analysed by HPLC or GC/EI-MS; techniques which require long, high-resolution separations, and complicated MS methods. Atmospheric pressure photoionization is a new technique that uses the energy of photons to generate ions from the vaporized eluent of an LC. This study compares the analysis of PAHs by LC/MS/MS using both reverse phase and normal phase chromatographic conditions.Materials and methods
A 20 ppm stock standard solution containing 16 common PAHs (see Figure 1). Under optimized LC/MS/MS conditions for both reversed phase and normal phase chromatographic conditions, the performance, linearity and LOD levels were determined. The most abundant single positive transition for each PAH compound was used to construct an +MRM method. Concentrations from ambient air for several PAHs were determined from samples collected on high volume filters at both urban and rural sites. Standard solutions were prepared by serial dilution in acetonitrile/ water and iso-octane for reversed phase and normal phase conditions, respectively. Spectrometer: API 3000 LC/MS/MS System with Analyst Software PhotoSpray Source (see Figure 2 and 3) HPLC: Perkin-Elmer 200 Series Autosampler and Micro pumps LC conditions: Acetonitrile/water (gradient mode) flow at 1.0 mL/min Column: Hypersil PAH, 100 x 4.6 mm; 5 m, 50:50 post column split Normal phase conditions: 100% iso-octane (isocratic mode) flow at 0.5 mL/min Column: Spherisorb Silica, 250 x 4.6 mm; 5 m Dopant: Toluene at a flowrate of 100 l/min.Results
The source uses a modified housing of a conventional APCI source. The corona discharge needle has been removed and replaced with a connection for the lamp gas (nitrogen) which protects the Magnesium Fluoride optical window from the vaporised mobile phase.
Reversed phase PAH separation was accomplished by gradient flow (ACN/H2O) through a Hypersil PAH column, 100 x 4.6 mm; 5 m. Individual MRM transitions were monitored simultaneously for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 4 is a 20 L injection of a 5 ng/L solution (100 ng on column).
Normal phase PAH separation was accomplished with 100% iso-octane through a Spherisorb Silica column, 250 x 4.6 mm; 5 m. Individual MRM transitions were monitored for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 5 is a 20 L injection of a 5 ng/L solution (100 ng on column)
The observed response for PAHs under normal p hase conditions show an increase on the order of 10 to 70 times of that observed under reversed phase conditions (Figure 6).
The post-column addition of dopant flow can have an effect on the overall response. This will vary from compound to compound. As shown in Figure 7, the response of naphthalene is increased by a factor of 7, while the response of benzo(a)pyrene is relatively unchanged.
Under normal phase conditions, 5000 pg/L fell out of the calibration range for most PAHs, except Indeno(1,2,3,c,d) pyrene and Dibenz(a,h)anthracene. Reversed phase conditions could be calibrated as high as 5000 pg/L (Figure 8).
Ambient air samples were collected on high volume filters(subsequently solvent extracted, dried to 0.2 ml) from three locations, rural, urban and a traffic tunnel. Sample chromatogramsare shown in Figure 9, and calculated ambient concentrations are shown in Table 2.
Conclusions
References
1Robb, D. B., Covey, T. R., Bruins, A. P., Anal. Chem. 2000, 72, 3653-3659.Authors
Gary Impey, Byron Kieser, Jean-Franois Alary, Applied Biosystems/MDS SCIEXAcknowledgements
The authors wish to thank Satoshi Irei (York University, Toronto, Canada) for supplying PAH standards and the ambient air samples for analysis.