The Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by LC/MS/MS Using ,,, an Atmospheric Pressure Photoionization Source

The atmospheric pressure photoionization source, PhotoSpray ion source (Robb, Covey and Bruins1) uses photons to ionize large quantities of a dopant molecule added along with the vaporized mobile phase. Analyte molecules are efficiently ionized through secondary reactions initiated by the charged dopant. This new technique was used to develop an analytical method for polycyclic aromatic hydrocarbons (PAHs), comparing the difference under reverse phase and normal phase chromatographic conditions. A total of 16 PAHs were analysed including the determination of unknown quantities in atmospheric samples from both rural and urban areas.

Introduction

Polycyclic aromatic hydrocarbons (PAHs) are a group of over 100 different chemicals, typically formed during incomplete combustion, which are present in a variety of environmental matrices. Airborne PAHs, usually found in aerosol particles, are implicated in heterogeneous atmospheric processing and as important indicators of gas-phase atmospheric processes. Certain carcinogenic PAHs may be carried by ultrafine particles in ambient air into the lungs, and are often the targets of rigorous environmental monitoring. Currently, environmental PAHs are most commonly analysed by HPLC or GC/EI-MS; techniques which require long, high-resolution separations, and complicated MS methods. Atmospheric pressure photoionization is a new technique that uses the energy of photons to generate ions from the vaporized eluent of an LC. This study compares the analysis of PAHs by LC/MS/MS using both reverse phase and normal phase chromatographic conditions.

Materials and methods

A 20 ppm stock standard solution containing 16 common PAHs (see Figure 1). Under optimized LC/MS/MS conditions for both reversed phase and normal phase chromatographic conditions, the performance, linearity and LOD levels were determined. The most abundant single positive transition for each PAH compound was used to construct an +MRM method. Concentrations from ambient air for several PAHs were determined from samples collected on high volume filters at both urban and rural sites. Standard solutions were prepared by serial dilution in acetonitrile/ water and iso-octane for reversed phase and normal phase conditions, respectively. Spectrometer: API 3000 LC/MS/MS System with Analyst Software PhotoSpray Source (see Figure 2 and 3) HPLC: Perkin-Elmer 200 Series Autosampler and Micro pumps

LC conditions: Acetonitrile/water (gradient mode) flow at 1.0 mL/min Column: Hypersil PAH, 100 x 4.6 mm; 5 m, 50:50 post column split Normal phase conditions: 100% iso-octane (isocratic mode) flow at 0.5 mL/min Column: Spherisorb Silica, 250 x 4.6 mm; 5 m Dopant: Toluene at a flowrate of 100 l/min.

Results

The source uses a modified housing of a conventional APCI source. The corona discharge needle has been removed and replaced with a connection for the lamp gas (nitrogen) which protects the Magnesium Fluoride optical window from the vaporised mobile phase.

Reversed phase PAH separation was accomplished by gradient flow (ACN/H2O) through a Hypersil PAH column, 100 x 4.6 mm; 5 m. Individual MRM transitions were monitored simultaneously for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 4 is a 20 L injection of a 5 ng/L solution (100 ng on column).

Normal phase PAH separation was accomplished with 100% iso-octane through a Spherisorb Silica column, 250 x 4.6 mm; 5 m. Individual MRM transitions were monitored for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 5 is a 20 L injection of a 5 ng/L solution (100 ng on column)

The observed response for PAHs under normal p hase conditions show an increase on the order of 10 to 70 times of that observed under reversed phase conditions (Figure 6).

The post-column addition of dopant flow can have an effect on the overall response. This will vary from compound to compound. As shown in Figure 7, the response of naphthalene is increased by a factor of 7, while the response of benzo(a)pyrene is relatively unchanged.

Under normal phase conditions, 5000 pg/L fell out of the calibration range for most PAHs, except Indeno(1,2,3,c,d) pyrene and Dibenz(a,h)anthracene. Reversed phase conditions could be calibrated as high as 5000 pg/L (Figure 8).

Ambient air samples were collected on high volume filters(subsequently solvent extracted, dried to 0.2 ml) from three locations, rural, urban and a traffic tunnel. Sample chromatogramsare shown in Figure 9, and calculated ambient concentrations are shown in Table 2.

Conclusions

  • The new PhotoSpray source provides a sensitive ionization technique for analysis of polycyclic aromatic hydrocarbons by LC/MS/MS.
  • Normal phase conditions provide an increase in sensitivity by a factor of at least 10 and as much as 70 over reversed phase conditions.
  • The APPI LC/MS/MS normal phase method provides adequate sensitivity to be useful in the analysis of atmospheric aerosol samples for PAHs.

References

1Robb, D. B., Covey, T. R., Bruins, A. P., Anal. Chem. 2000, 72, 3653-3659.

Authors

Gary Impey, Byron Kieser, Jean-Franois Alary, Applied Biosystems/MDS SCIEX

Acknowledgements

The authors wish to thank Satoshi Irei (York University, Toronto, Canada) for supplying PAH standards and the ambient air samples for analysis.


'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:5/19/2016)... (PRWEB) , ... May 19, 2016 , ... ... research organization (CRO) has welcomed Abu Siddiqui as Director, Large Molecule & Biomarker ... executing biologics, vaccine and translational biomarker discovery studies for preclinical and clinical safety ...
(Date:5/18/2016)... 2016 The Biotech industry continues to ... that there are no opportunities ahead. Today, ActiveWallSt.com has on ... THLD ), Seattle Genetics Inc. (NASDAQ: SGEN ... Corp. (NASDAQ: OPHT ). Sign up now to ... http://www.activewallst.com/ Threshold Pharmaceuticals Inc.,s shares gained ...
(Date:5/17/2016)... ZURICH, Switzerland , May 18, 2016 ... their GONAL-f® 2.0 prefilled pen following approval by EMA, ... Merck, a leading science and technology company, the new ... and bring an increased level of confidence to patients ... the new GONAL-f® prefilled pen easier to handle with ...
(Date:5/17/2016)... Switzerland , May 17, 2016 ... company located in Basel, Switzerland ... investigational oral inhibitor of P38 mitogen-activated protein kinase. ... , Strekin will build the necessary ... in indications in which MAP Kinases play fundamental ...
Breaking Biology Technology:
(Date:5/3/2016)... 2016  Neurotechnology, a provider of high-precision biometric ... Biometric Identification System (ABIS) , a complete system ... ABIS can process multiple complex biometric transactions with ... fingerprint, face or iris biometrics. It leverages the ... MegaMatcher Accelerator , which have been used ...
(Date:4/28/2016)... First quarter 2016:   , ... the first quarter of 2015 The gross margin was ... 18.8) and the operating margin was 40% (-13) Earnings ... flow from operations was SEK 249.9 M (21.2) , ... SEK 7,000-8,500 M. The operating margin for 2016 is ...
(Date:4/26/2016)... April 27, 2016 Research ... Multi-modal Biometrics Market 2016-2020"  report to their offering.  ... The analysts forecast the global multimodal ... 15.49% during the period 2016-2020.  Multimodal ... sectors such as the healthcare, BFSI, transportation, automotive, ...
Breaking Biology News(10 mins):