The Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by LC/MS/MS Using ,,, an Atmospheric Pressure Photoionization Source

The atmospheric pressure photoionization source, PhotoSpray ion source (Robb, Covey and Bruins1) uses photons to ionize large quantities of a dopant molecule added along with the vaporized mobile phase. Analyte molecules are efficiently ionized through secondary reactions initiated by the charged dopant. This new technique was used to develop an analytical method for polycyclic aromatic hydrocarbons (PAHs), comparing the difference under reverse phase and normal phase chromatographic conditions. A total of 16 PAHs were analysed including the determination of unknown quantities in atmospheric samples from both rural and urban areas.

Introduction

Polycyclic aromatic hydrocarbons (PAHs) are a group of over 100 different chemicals, typically formed during incomplete combustion, which are present in a variety of environmental matrices. Airborne PAHs, usually found in aerosol particles, are implicated in heterogeneous atmospheric processing and as important indicators of gas-phase atmospheric processes. Certain carcinogenic PAHs may be carried by ultrafine particles in ambient air into the lungs, and are often the targets of rigorous environmental monitoring. Currently, environmental PAHs are most commonly analysed by HPLC or GC/EI-MS; techniques which require long, high-resolution separations, and complicated MS methods. Atmospheric pressure photoionization is a new technique that uses the energy of photons to generate ions from the vaporized eluent of an LC. This study compares the analysis of PAHs by LC/MS/MS using both reverse phase and normal phase chromatographic conditions.

Materials and methods

A 20 ppm stock standard solution containing 16 common PAHs (see Figure 1). Under optimized LC/MS/MS conditions for both reversed phase and normal phase chromatographic conditions, the performance, linearity and LOD levels were determined. The most abundant single positive transition for each PAH compound was used to construct an +MRM method. Concentrations from ambient air for several PAHs were determined from samples collected on high volume filters at both urban and rural sites. Standard solutions were prepared by serial dilution in acetonitrile/ water and iso-octane for reversed phase and normal phase conditions, respectively. Spectrometer: API 3000 LC/MS/MS System with Analyst Software PhotoSpray Source (see Figure 2 and 3) HPLC: Perkin-Elmer 200 Series Autosampler and Micro pumps

LC conditions: Acetonitrile/water (gradient mode) flow at 1.0 mL/min Column: Hypersil PAH, 100 x 4.6 mm; 5 m, 50:50 post column split Normal phase conditions: 100% iso-octane (isocratic mode) flow at 0.5 mL/min Column: Spherisorb Silica, 250 x 4.6 mm; 5 m Dopant: Toluene at a flowrate of 100 l/min.

Results

The source uses a modified housing of a conventional APCI source. The corona discharge needle has been removed and replaced with a connection for the lamp gas (nitrogen) which protects the Magnesium Fluoride optical window from the vaporised mobile phase.

Reversed phase PAH separation was accomplished by gradient flow (ACN/H2O) through a Hypersil PAH column, 100 x 4.6 mm; 5 m. Individual MRM transitions were monitored simultaneously for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 4 is a 20 L injection of a 5 ng/L solution (100 ng on column).

Normal phase PAH separation was accomplished with 100% iso-octane through a Spherisorb Silica column, 250 x 4.6 mm; 5 m. Individual MRM transitions were monitored for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 5 is a 20 L injection of a 5 ng/L solution (100 ng on column)

The observed response for PAHs under normal phase conditions show an increase on the order of 10 to 70 times of that observed under reversed phase conditions (Figure 6).

The post-column addition of dopant flow can have an effect on the overall response. This will vary from compound to compound. As shown in Figure 7, the response of naphthalene is increased by a factor of 7, while the response of benzo(a)pyrene is relatively unchanged.

Under normal phase conditions, 5000 pg/L fell out of the calibration range for most PAHs, except Indeno(1,2,3,c,d) pyrene and Dibenz(a,h)anthracene. Reversed phase conditions could be calibrated as high as 5000 pg/L (Figure 8).

Ambient air samples were collected on high volume filters(subsequently solvent extracted, dried to 0.2 ml) from three locations, rural, urban and a traffic tunnel. Sample chromatogramsare shown in Figure 9, and calculated ambient concentrations are shown in Table 2.

Conclusions

  • The new PhotoSpray source provides a sensitive ionization technique for analysis of polycyclic aromatic hydrocarbons by LC/MS/MS.
  • Normal phase conditions provide an increase in sensitivity by a factor of at least 10 and as much as 70 over reversed phase conditions.
  • The APPI LC/MS/MS normal phase method provides adequate sensitivity to be useful in the analysis of atmospheric aerosol samples for PAHs.

References

1Robb, D. B., Covey, T. R., Bruins, A. P., Anal. Chem. 2000, 72, 3653-3659.

Authors

Gary Impey, Byron Kieser, Jean-Franois Alary, Applied Biosystems/MDS SCIEX

Acknowledgements

The authors wish to thank Satoshi Irei (York University, Toronto, Canada) for supplying PAH standards and the ambient air samples for analysis.


'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
(Date:12/17/2014)... 17, 2014 Audacity, a leading California-based ... member of the Huntsworth Health Group, is proud to ... team. In her role, Gonzales will serve as Senior ... incredibly honored to welcome a talent of Jamie’s caliber ... Gaëtan Fraikin. “She is a proven healthcare innovation leader ...
(Date:12/17/2014)... 2014  United Therapeutics Corporation (NASDAQ: UTHR ... as changes to Martine Rothblatt , Ph.D,s compensation ... United Therapeutics announced the promotion of Roger ... and David Zaccardelli , Pharm.D. to Executive Vice ... promotions, Dr. Rothblatt,s title will change from Chairman and ...
(Date:12/17/2014)... 17, 2014 Tigercat Pharma, Inc. today ... evaluating the investigational oral NK-1 receptor antagonist VPD-737 ... chronic itch who are unresponsive or inadequately responsive ... steroids and antihistamines. The primary ... Scale (VAS) itch score from baseline, comparing drug ...
(Date:12/15/2014)... (PRWEB) December 15, 2014 METTLER ... inline PVM tool, ParticleView V19 with PVM ... situ probe-based particle vision and measurement tool continuously ... process conditions. ParticleView V19 then automatically prepares a ... tracking particle size and concentration changes. This compelling ...
Breaking Biology Technology:Audacity Welcomes Healthcare Communications Expert Jamie Gonzales as Senior Vice President of Client Services 2United Therapeutics Corporation Announces Executive Promotions and CEO Compensation Change 2United Therapeutics Corporation Announces Executive Promotions and CEO Compensation Change 3United Therapeutics Corporation Announces Executive Promotions and CEO Compensation Change 4Velocity Pharmaceutical Development, LLC and Tigercat Pharma, Inc. Announce Phase 2 Results for VPD-737 in Patients With Chronic Pruritus 2Velocity Pharmaceutical Development, LLC and Tigercat Pharma, Inc. Announce Phase 2 Results for VPD-737 in Patients With Chronic Pruritus 3Velocity Pharmaceutical Development, LLC and Tigercat Pharma, Inc. Announce Phase 2 Results for VPD-737 in Patients With Chronic Pruritus 4New Real-Time In Situ Probe-Based Video Microscope from METTLER TOLEDO 2
... will seek additional venture funding this year in ... president and CEO Loren G. Peterson said. , ,Peterson ... the Innovation Monday program. He said ZyStor, which is ... filing the application for its lead product in early ...
... Wis. - For the second time in six ... giant, claiming patent infringement. , ,This time Hyperphrase Technologies ... Internet search engine company, for allegedly using technology that ... other sites or advertisements. Hyperphrase claims that Google Toolbar ...
... enjoyed another year of impressive growth, according to the ... a quarterly publication on the trademark industry published by ... and online brand monitoring services. , ,The report provides ... trademark filing activity, a ranking of leading trademark law ...
Cached Biology Technology:Milwaukee area biotech ZyStor seeks more funding 2Milwaukee area biotech ZyStor seeks more funding 3
(Date:12/15/2014)... , Dec. 12, 2014 Research and Markets ... addition of the "Global Facial Recognition Market ... http://photos.prnewswire.com/prnh/20130307/600769 Facial ... of individuals. Facial recognition system measures the overall ... jaw edges, mouth, and the distance between eyes. ...
(Date:12/11/2014)... SALEM, N.C. , Dec. 10, 2014  That blood ... for quite a while. Hypertension – the medical term for ... in the early 1800s, and the inflatable cuff that,s used ... That doesn,t, however, mean there,s nothing new about hypertension, its ... some long-held beliefs about the condition and the best ways ...
(Date:12/10/2014)... 9, 2014  Valencell, a leader in performance biometric ... staggering demand from its licensees for highly accurate, clinically ... not solely coming from fitness and health sectors, but ... "A wearable is only as useful as the biometric ... ultimate driver in long-term mass consumer adoption of wearable ...
Breaking Biology News(10 mins):Global Facial Recognition Market 2015-2019: Key Vendors are 3M Cogent, Cognitec Systems, NEC and Safran Group 2Global Facial Recognition Market 2015-2019: Key Vendors are 3M Cogent, Cognitec Systems, NEC and Safran Group 3Research points to need for new approaches to treatment of high blood pressure 2Research points to need for new approaches to treatment of high blood pressure 3Research points to need for new approaches to treatment of high blood pressure 4Wearable Technology Products Demand Highly Accurate Biometric Technology 2
... egg question. Where do the infectious protein particles called ... cause neurodegenerative disorders, such as mad cow/Creutzfeld-Jakob disease, in ... of their properly folded protein counterparts, but they usually ... Biochemists at Emory University School of Medicine have identified ...
... Institute of Molecular and Cell Biology (IMCB) has discovered ... deadly plant toxin Ricin, widely feared for its bioterrorism ... for the tens of thousands of hospital-acquired infections in ... this first ever genome-wide study to understand how the ...
... sometimes have its advantages. That,s according to a study of ... native Chilean forest. The data show that the pollinators those ... of resources found in agricultural lands. As a result, trees ... successfully when separated by resource-poor pine plantations than by those ...
Cached Biology News:Cellular stress can induce yeast to promote prion formation 2A*Star scientists discover how to combat hospital-acquired infections and life-threatening toxins 2A*Star scientists discover how to combat hospital-acquired infections and life-threatening toxins 3As agricultural riches waylay pollinators, an endangered tree suffers 2As agricultural riches waylay pollinators, an endangered tree suffers 3
... Costar strip ejector is a unique product ... strips from their holder. Linkage: This CLS ... to easily match Cornings product number. If ... the old Sigma-Aldrich number (Z71,314-7) or contact ...
...
... The ideal modular cleanroom for biological ... that require cleanliness (to Class 10/ISO ... precision-joined to create ultra-smooth surfaces that ... disinfect. One-inch radius corners simplify cleaning. ...
This free-standing work station safely ventilates fumes upwards, away from the work surface. Optional Exhaust Purification System features bonded charcoal filters that remove organic chemical vapors ...
Biology Products: