Speciation of Arsenic by LC-ICP-MS

Michael Leist Varian

Analytical Instruments Mulgrave, Victoria 3170, Australia E-mail: icpms@varianinc.com

Introduction

As toxicity, mobility, and bioavailability can differ greatly between the various chemical species in which an element occurs, reporting only the total concentrations can often be misleading. Arsenic is one such example where the various species differ; the inorganic trivalent form (As III) is the most toxic, followed by the inorganic pentavalent form (As V). Other common forms of arsenic include monomethyl arsenic (MMA), dimethyl arsenic (DMA) and arsenobetaine (AsB), which have signifi cantly reduced toxicities. Therefore, separation and detection of these species can greatly assist risk-based toxicity assessments.

When Liquid Chromatography (LC) is interfaced with Inductively Coupled Plasma Mass Spectrometry (ICP-MS), species elute one by one from the LC column directly to the ICP-MS for detection by elemental speciation. The coupling of LC to ICP-MS is a straight forward task; the LC column is connected to the nebulizer of the ICP-MS by a piece of PEEK tubing, and no hardware changes are required to either the LC or the ICP-MS. Coupling an LC to a Varian ICP-MS has the added advantage of offering high sensitivity, with over 80% of the analyte ions passing through the skimmer cone being transferred to the quadrupole1. The Varian ICP-MS, with its patented 90 degree ion mirror2, is the worlds fi rst ICP-MS with tunable gigahertz sensitivity.

The following experiment shows the sensitivity and detection capabilities of the high sensitivity Varian ICP-MS for element speciation.

Experimental

Instrumentation

A schematic of the overall system is shown in Figure 1.

The LC system used was a Varian ProStar 230 Tertiary Solvent Delivery Module, a Rheodyne injector with a 50 μL sample loop, and a Hamilton PRP-X100 (4.1 mm x 250 mm, 10 μm) anion exchange column.

The Varian ICP-MS was used for all arsenic measurements. ICP-MS setup was fully controlled by Varians ICP-MS Expert software, which provides one-step instrument setup, optimization and method development. Prior to connecting the ICP-MS to LC, it was tuned to high sensitivity mode automatically by using the auto optimization routine included in the software. The ICP-MS nebulizer was then connected to the LC column using a 20 cm length of PEEK tubing (0.010 I.D.). See Tables 1-3 for the ICP-MS and LC conditions used.

Reagents and samples

Deionized water (18 MΩ cm, Millipore MilliQ, Billerica, MA, USA) was used for all solution preparations (calibration solutions and mobile phase).

Calibration standards

Arsenic pentoxide and arsenic trioxide were obtained from SPEX CertiPrep (Metuchen, NJ USA). Arsenobetaine (AsB, C5H11AsO2) was obtained from Sigma Aldrich (Castle Hill, NSW Australia). Cacodylic acid (dimethylarsenic acid, DMA) and monosodium acid methane arsonate (MMA) were purchased from Chem Service (West Chester, PA USA). All calibration solutions were prepared daily.

Mobile phase LC

Ammonium carbonate (Merck, Kilsyth, Victoria, Australia) and either Suprapur ammonia solution obtained from Merck (Kilsyth, Victoria, Australia) or AR Select Plus nitric acid, (Mallinckrodt Baker, Phillipsburg, NJ, USA) were used for pH adjustments of the mobile phase. The mobile phase was prepared daily.

Reference materials

Standard Reference Material 1640, Trace Elements in Natural Water was obtained from the National Institute of Standards and Technology (Gaothersburg, MD, USA). Trace Metals in Fish (HPS CRM-TMF Lot #123921), Trace Metals in Drinking Water (HPS CRM-TMDW Lot#222518) and Oyster Tissue (HPS CRM-OT Lot #209510) were all obtained from High-Purity Standards (Charleston, SC, USA). The urine reference material (Seronorm Lot#No2525) was obtained from Sero As, (Billingstad, Norway).

Results and Discussions

The time resolved data from Varian ICP-MS Expert was exported as a .CDF fi le and opened in the Galaxie software, where program data analysis was performed. For reviewing acquired data, Galaxie provides an easy way to view single or multiple chromatograms, and compare results in a single screen.

Elemental speciation

Speciation of the fi ve inorganic and organic arsenic species can be performed using either an isocratic or gradient LC method (see Figures 2 and 3). Using either method, excellent calibrations for each of the arsenic species were obtained with calibration coeffi cients ≥ 0.999 achieved when calibrating on solutions ranging from 0.01 to 50 μg/L. An example calibration curve for MMA is shown in Figure 4. Note the low calibration solution in this calibration is 0.01 μg/L (10 ng/L).

While the gradient method increases the analysis time by approximately 9 minutes (4 minutes increased analysis time and 5 minutes column re-equilibration), it does provide better separation of As(III) from DMA, and improved separation from potential interferences. For example, samples with a high Cl content can result in an additional peak (75ArCl). As shown in Figure 5, the gradient method adequately separates the Cl peak from the neighbouring MMA and AsV peaks.

Typical detection limits

Table 4 shows the typical detection limits (DLs) for the fi ve common organic and inorganic forms of arsenic. All the measurements were made under routine analytical laboratory conditions, not clean-room conditions. This work indicates typical DL values that can be routinely achieved outside a clean-room in a clean laboratory.

Although previous experiments have reported signal improvements when 13% methanol is aded to the mobile phase4, this application note demonstrates that the Varian ICP-MS can produce low detection limits without this step. The addition of methanol can also cause problems if other elements such as Cr (VI) are measured, as the formation of 40Ar13C leads to substantial degradation of the signal-to-noise ratio, and higher limits of detection. Additions above 1% methanol can also cause problems, as the chromatographic separation between AsB and As(III) may be degraded to such an extent that base line resolution is not achievable.

Analysis of reference materials

Given the lack of reference materials that are certifi ed for all fi ve arsenic species, a number of reference materials were spiked with known quantities of each species and the % recovery recorded, to test for robustness of the Gradient method.

The reference materials analyzed included two water reference materials, NIST 1640 and HPS TMDW, two biological tissue reference materials, HPS OT and HPS TMF, and a urine reference material. Each reference material was spiked with 11 μg/L of each As compound and analyzed. The HPS TMF and the Seronorm urine reference material were diluted by a factor of 100 prior to injection. The percentage spike recoveries for each reference material can be viewed in Table 5. Very good recoveries were achieved, especially considering that no internal standard was used.

Conclusions

This work has demonstrated that the Varian ICP-MS is an excellent detector when coupled with liquid chromatography for elemental speciation. Low ng/L detection limits can be routinely obtained and very good recoveries were obtained on a variety of spiked reference materials.

References

1. S. Elliott, M. Knowles and I., Kalinitchenko, A New Direction in ICP-MS, Spectroscopy , 19(1), 30 (2004)

2. I. Kalinitchenko, Ion Optical System for a Mass Spectrometer , Australian Patent 750860, 14 November 2002, US Patent 6,614,021B1, 2 September 2003.

3. K. Ackley, C. BHymer, K. Sutton, J. Caruso, Speciation of Arsenic in Fish Tissue Using Mircowave-Assited Extraction Followed by HPLC-ICP-MS, Journal of Analytical Atomic Spectrometry , 14, 845 (1999)

4. Y. Martinez-Bravo, A.F. Roig-Navarro, F.J. Lopez, F. Hernandez, Multielemental Determination of Arsenic, Selenium and Chromium(VI) Species in Water by High-Performance Liquid Chromatography Inductively Coupled Plasma Mass Spectrometry, Journal of Chromatography A , 926, 265 (2001).


'"/>

Source:


Page: All 1 2 3 4 5

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Speciation Arsenic ICP

(Date:9/2/2014)... 2014  Xencor, Inc. (NASDAQ: XNCR ... antibodies for the treatment of autoimmune diseases, asthma ... appointment of Debra Zack , M.D., Ph.D., ... Lloyd Rowland , senior vice president, chief ... "Debra,s expertise in biologics clinical development and medical ...
(Date:9/2/2014)... Winston-Salem, NC and Rocky Mount, NC (PRWEB) September 02, ... (PMG) announced today the establishment of its 11th clinical ... 1, PMG Research has assumed ownership and operation of ... in Rocky Mount, NC, further expanding PMG’s presence in ... to an additional 130,000 patients. For the patients in ...
(Date:9/1/2014)... 1, 2014 Reportlinker.com announces that a ... catalogue: India Solid Waste Management Vehicles ... http://www.reportlinker.com/p02350445/India-Solid-Waste-Management-Vehicles-Market-Forecast-and-Opportunities-2019.html In ... governing bodies responsible for complete handling and regulation ... average, around 135,000 MT solid waste is generated ...
(Date:8/31/2014)... Available spectra is similar for all L-series products ... a wide range of plants. The spectra is ... and pending wide spectrum technology. The single chip ... in contrast to growth spectra made from monochromatic ... large variety of differently sized and shaped growing ...
Breaking Biology Technology:Xencor Appoints Debra Zack, M.D., Ph.D., Vice President, Clinical Development and Lloyd Rowland, Senior Vice President, Chief Compliance Officer and General Counsel 2Xencor Appoints Debra Zack, M.D., Ph.D., Vice President, Clinical Development and Lloyd Rowland, Senior Vice President, Chief Compliance Officer and General Counsel 3PMG Research Announces the Formation of its Eleventh Site 2PMG Research Announces the Formation of its Eleventh Site 3India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 2India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 3India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 4India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 5India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 6India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 7India Solid Waste Management Vehicles Market Forecast and Opportunities, 2019 8
... ... Qteros of Marlborough, Mass., and Applied CleanTech, based in Israel, today ... from municipal and agricultural liquid waste into ethanol fuel for cars. ... feedstock has been found to boost ethanol production, address sewage sludge ...
... CryoLife, Inc., (NYSE: CRY ) ... today announced that it has received a Humanitarian Use Device ... its CryoValve® SG aortic human heart valve. The HUD ... Exemption (HDE) for the CryoValve SG aortic human heart valve, ...
... Calif., Oct. 6 Varian, Inc. (NasdaqGS: VARI) announced that at ... its stockholders adopted the merger agreement entered into with Agilent ... per share in cash. Approximately 84% of the shares ... the record date for the special meeting, voted to adopt ...
Cached Biology Technology:Cellulose from Wastewater Can Now be Turned into Fuel for Cars 2Cellulose from Wastewater Can Now be Turned into Fuel for Cars 3Cellulose from Wastewater Can Now be Turned into Fuel for Cars 4CryoLife Receives Humanitarian Use Device Designation for SynerGraft(R) Processed Human Aortic Heart Valves 2CryoLife Receives Humanitarian Use Device Designation for SynerGraft(R) Processed Human Aortic Heart Valves 3CryoLife Receives Humanitarian Use Device Designation for SynerGraft(R) Processed Human Aortic Heart Valves 4
(Date:9/1/2014)... CAMBRIDGE, Mass-- Conventional wisdom has long held that ... of coral reefs are passive organisms that ... substances, such as nutrients and oxygen. But now ... Science (WIS) in Israel have found that they ... sweep water into turbulent patterns that greatly enhance ...
(Date:9/1/2014)... nocturnal animals, bats are perfectly adapted to a life without ... the reflected echoes to measure distance to obstacles or prey. ... echo delays. A study carried out by researchers at Technische ... this map dynamically adapts to external factors. , ... flies in too close to an object, the number of ...
(Date:8/31/2014)... in modern medicine is developing drugs that are highly ... side-effects to the patient. Such properties are directly related ... the drug should have a shape that is perfectly ... it with high specificity. Publishing in Nature Chemistry ... that can impact the 3D structure of bioactive peptides ...
Breaking Biology News(10 mins):Nature's tiny engineers 2Nature's tiny engineers 3Zooming in for a safe flight 2A new synthetic amino acid for an emerging class of drugs 2A new synthetic amino acid for an emerging class of drugs 3
... VIRGINIA KEY, Fla. Seawater is a complex, ... materials that despite scientists best efforts, presents difficulties ... energy. Like the water itself, the calculations scientists ... revisions and clarifications over the years as research ...
... is available in Spanish . , TECNALIA ... multispectral artificial vision systems to enhance the value of electronic ... The aim of this project, known as SORMEN, is ... from electronic waste based on a system of multispectral vision ...
... part in protein folding to ensure cells function properly, ... with information critical to designing vaccines and helps them ... and Parkinson,s, that result from molecular motion gone awry. ... large computer clusters to run simulations. Or they had ...
Cached Biology News:New equation of state of seawater 2New equation of state of seawater 3TECNALIA uses artificial vision to improve recycling of electronic scrap metal 2TECNALIA uses artificial vision to improve recycling of electronic scrap metal 3TECNALIA uses artificial vision to improve recycling of electronic scrap metal 4New open-source software permits faster desktop computer simulations of molecular motion 2New open-source software permits faster desktop computer simulations of molecular motion 3
... Clone/PAD: ZMD.358. Immunogen: Synthetic ... region of human mouse and ... with human and mouse LRP5 ... (positive controls: mouse LRP5-pSEC-tag-transfected 293T ...
MarrowMAX Bone Marrow Medium...
... recognizes murine IRF7 protein, and ... related IRF proteins. Reactivity of ... protein was confirmed by western ... cells transfected with IRF7. Reactivity ...
Rabbit anti-phospho-EGFr (Tyr992) - Concentrate...
Biology Products: