Serial Dilution with the Bravo Liquid Handling Platform


Introduction
Many laboratory protocols require the serial dilution of reagents or compounds. IC50 assays are commonly used to evaluate drug efficacy, and assay development procedures as well as standard curve generation involve the serial dilution of proteins, compounds, or other detection agents. These processes can be streamlined with automated liquid handling equipment with serial dilution capabilities.


Velocity11s Bravo Liquid Handling Platform can be utilized to perform serial dilutions. It has nine standard deck positions that can hold any SBSstandard microplate. These deck positions can also be configured for heating, cooling, shaking, and tip washing. The design of the head, along with various tip and head options, allow serial dilutions to be performed by column or row, from 300 nL to 200 μL dilutions.

Operators can easily write protocols with the included VWorks4 software package. One key feature is the Serial Dilution Wizard. The Serial Dilution Wizard lets the operator easily program the instrument for making serial dilutions. The wizard walks the operator through the steps of a serial dilution task: tip changes, mix parameters, and volume transfers. The operator can either determine the transfer based on initial volumes (transfer by volume) or by a concentration gradient (transfer by concentration).

This technical note explores the parameters that lead to an efficient, precise, and accurate serial dilution protocol. The goal was to produce good precision and accuracy for each dilution in the least amount of time. The mix parameters are essential for a robust serial dilution task. The parameters explored include:
The Number of mix cycles
Mix height
Liquid class settings
Tip retraction/extension


Materials Used
Velocity11 Bravo with a 96 Channel LT Disposable Tip Head
Velocity11 96LT 200 μL tips (product no. 06880.002)
Velocity11 96-well manual fill reservoirs (product no. 08105.001)
96 well polystyrene, black flat clear bottom plates (Greiner 655087)
Fluorescein solution (3 μM in 50 mM Tris-HCl, pH 8.0)
SPECTRAFluor Fluorescence Spectrophotometer (Tecan)


Summary
Accurate serial dilution with a full channel head
Efficient mixing with as few as three mix cycles
No tip changes necessary
No wash steps necessary


Method
A reservoir containing 60 mL of fluorescein solution is placed on position 2 of the Bravo (fluorescein reservoir for serial dilution by Row is placed on position 7). A reservoir containing 100 mL of 50 mM Tris-HCl is placed on position 3. A 96-well polystyrene plate is placed on position 5. 200 μL tip boxes are placed on positions 6 and 9. A Works4 liquid class for 51-200 μL dispense is utilized.


1. Tips are pressed onto the head from position 9.

2. 150 μL Tris-HCl solution is transferred from position 3 to position 5. Aspirate parameters are 6 mm from the bottom of the reservoir with a 2 μL pre-aspirate volume. Dispense parameters are 2 mm from the bottom of the plate, with a 2 μL blowout volume.

3. Tips are unloaded back to position 9.

4. Eight tips are pressed onto the last column of the head from position 6.

5. Fluorescein solution is mixed at position 2. Mix parameters are 150 μL of , 3 mix cycles 6 mm from the bottom of the plate, with a 2 μL air gap.

6. 150 μL fluorescein solution is transferred from position 2 to Column 1 of the plate at position 5. Aspirate parameters are 6 mm from the bottom of the reservoir with a 2 μL pre-aspirate volume. Dispense parameters are 2 mm from the bottom of the plate, with a 2 μL blowout volume.

7. 190 μL was mixed in Column 1 of the 96-well plate. Mix parameters vary on each experiment performed throughout the study as described below.

8. A 1:2 serial dilution (150 μL) was performed from Column 1-10. All aspirations have a 2 μL pre-aspirate volume at 2 mm from the bottom of the plate. All dispenses occur 2 mm from the bottom of the plate, with a 2 μL blowout volume. Mix parameters vary on each experiment performed throughout the study as described below.

9. 150 μL of excess volume Column 10 is transferred to waste.

10. Tips are unloaded back to position 6.

11. Plates are centrifuged at 1800 rpm for 60 seconds to ensure consistent well menisci.

12. Fluorescence absorption is read at 485nm excitation and 535nm emission wavelength with a gain of 63, 5 flash cycles, and a 40 ms integration time.


A similar serial dilution method by row is performed, with the exception that the fluorescein reservoir is placed on position 7.

Fluorescence measurements from each well are used to determine the precision of the transfer. Coefficient of Variance (CV) calculations are made by dividing the standard deviation by the mean for each column/row. Accuracy is calculated based on an equation derived from a fluorescein/Tris-HCl calibration curve consisting of data points at 2, 1.5, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01 and 0.005 μM, compared to the actual fluorescence value in each well.

Results may vary, depending on individual experimental methods and liquid class optimization.


Results
Number of mix cycles

The first experiment was to determine the number of mixes required to produce a precise and accurate 1:2 serial dilution. The volume of the mix was just under the maximum volume of the tip, and the number of mixes was varied from 3-20 cycles. Table 1 shows the resulting CV of each column for each mix cycle protocol. The average precision (averaging CVs for columns 1-10) shows that the precision improves asymptotically as the number of mix cycles is increased (Figure 2). Three mixes yielded an average CV of 11.8%, while 20 mixes had a considerably better CV of 1.7%. The precision in all cases worsens as the serial dilution proceeds across the plate; this is expected as the error in the earlier columns is propagated with each transfer.



In addition, the accuracy ratio improves as the number of mix cycles increases (Figure 3). The accuracy ratio is an average of the concentration of the diluted column compared to the previous columna perfect serial dilution would have an accuracy ratio of 1:2.00 across the entire plate. The accuracy ratio of the plate improves with more mix cycles, varying from 1:1.85 to 1:2.01. This is expected as the fluorescein would be more equally distributed in the well the more the well contents are mixed, ensuring that transfers are made from a uniformly mixed solution.


While the precision and accuracy with 20 mix cycles is very close to a perfect serial dilution, it also undoubtedly takes too much time for an automated process. The 20 mix cycle protocol takes 20 minutes to perform, while a 3 mix cycle protocol takes less than six minutes. Efforts were then focused on the factors that can improve the 3 mix cycle protocol to produce accuracy and precision results that were closer to the achieved 20 mix cycle protocol.

Mix Height
The mix height was adjusted in order to determine the effect of distributing the liquid at a different location in the well. As the mix height was increased, the average precision improved. Figure 4 shows that at a height of 3mm, the average precision is 3.9%, significantly lower than 15% at a height of 0.1mm. Following the same trend as previously observed, accuracy tracks with precision, so the higher mix height also improves our accuracy ratio to 1.95 (Table 2). This trend is observed because the higher dispense height at this dispense speed ensures that more of the sample is circulated by the mix cycle: in a mix far from the bottom of the well, dispensed liquid is forced closer to the bottom while dispensing, and aspirated liquid is pulled from the top of the well. If the mix occurs closer to the bottom of the plate, the dispensed liquid is pulled back into the tip during the aspiration.



Mix Liquid Class Setting
The next parameter explored was the mix liquid class settings. Each liquid class has a set of velocities and accelerations for the aspirate and dispense. This experiment was to determine if increasing the aspirate and dispense speeds of the mix would result in improved serial dilution results.

The original liquid class speed for the mix was 100 μL/s velocity and 500 μL/s2 acceleration. Based on the average precision of the data (Table 3), mixing became more effective as the mix speed is increased. Accuracy is also improved as the mix speed increases, because of consistency of the mixes across the plate. Under these conditions, a mix liquid class setting of a velocity 300μL/s and an acceleration 1000μL/s2 was utilized. There was no appreciable difference in time or performance by increasing the velocity from 300 μL/s to 500 μL/s.


Dynamic Tip Retraction/Extension
To further explore improving precision and accuracy, dynamic tip retraction/extension was used on the mix parameter. Utilizing this particular parameter in the mix cycle, the tip moves into the well during each aspiration and retracts during each dispense. The following retraction/extension test heights were calculated based on the optimal mix height on experiment 2 to give an effective mixing. Mix area was also considered on this test (Figure 5). This will show whether having the tips retracts/extends during the mixing gives a noticeable change in the CV and amount transferred.


In this experiment, the average precision improved slightly as mix area increased (Table 4). Mixing in half the well was marginally worse than mixing in the entire well, and mixing in the lower half of the well was marginally worse than mixing in the top half. In fact, mixing in half the well was worse than not moving the tips during the mix at all. There was little to no variation in accuracy as we changed the tip retraction, suggesting that this parameter did not have an effect, and/or we had optimized the mix parameters to the best degree possible.


Number of mix cycles with the improved mix parameters
Based on the experiments above, the conclusion was that the best mix parameters should have the following characteristics:
Three mix cycles
2 mm mix height
300 μL/s velocity and 1000 μL/s2 acceleration
0.02 mm/μL tip retraction

To verify this conclusion, the first experiment (varying the number of mix cycles) was repeated with the improved mix parameters. Table 5 compares the old mix parameters with the improved mix parameters. The improved mix parameters (Figure 6) provides increased precision and accuracy because the mix was more effective. Additionally, the improved mix parameters decreased the time required to run a serial dilution protocol by nearly a minute. Most of this improvement was due to increased liquid class speeds. Accuracy also improved in comparison to the original mix settings.



Serial Dilution by Row
The Bravo can perform serial dilutions by column and by row. While all of the previous experiments were serial dilution by column, it was necessary to confirm that the serial dilution by row would work equally well. Table 6 shows that by applying the improved mix parameters, high precision and accuracy was possible in another serial dilution method. Average precision was improved mostly due to the fact that there were seven dilutions instead of 10; error was propagated over a fewer number of dilutions. With fewer dilutions, the overall serial dilution time decreased as well.



Conclusions
This study shows that the Bravo can quickly and reliably serial dilute across columns or rows of a plate. Based on these experiments, the following settings gave us low CV and an accurate overall dispense. It is possible to achieve accurate and precise serial dilution results in a short process time with:
No tip changes
A minimum number of mix cycles
No wash steps

The ability of the Bravo to do serial dilution with a 96 or 384 channel disposable tip head means that serial dilution can be performed with one instrument with one head type. A head swap is not necessary. A parameter setting in the protocol allowed a change from transferring liquid with the entire head to transferring with a single row or column.

Serial dilution experiments can be optimized through a number of different parameter changes, and based on the results shown here, the keys to obtaining accurate and precise mixing are the following, in order of decreasing contribution to overall mixing efficiency:
Maximize volume of mixing
Maximize mixing speed
Maximize the area traversed by tip retraction/extension.



'"/>

Source:


Page: All 1 2 3 4 5 6 7 8

Related biology technology :

1. Serial Analysis of Gene Expression Using ABI PRISM DNA Sequencers and BigDye Terminators
2. Luciferase Reporter Gene Cells and Recombinant Luciferase Dilutions Studied with the CLIPR Luciferase Assay Kit and the LMax Microplate Luminometer (MaxLine Application Note #38)
3. Ergonomics and Liquid Handling How can BrinkmannTM Help ?
4. Determination of N-nitrosamines in Baby Bottle Rubber Teats by Liquid Chromatography-Atmospheric Pressure Chemical Ionization Mass Spectrometry
5. Sciclone ALH 3000 Liquid Handler: Functional Characterization of the High Volume Head - Part 1: Plate-to-Plate Precision Testing
6. Precision Testing for the Sciclone Advanced Liquid Handler Z8 Option
7. 384 Well Pipetting Precision and Accuracy Testing Using the Sciclone Advanced Liquid Handler
8. Cancer-Related miRNAs Uncovered by the mirVana miRNA Microarray Platform
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Serial Dilution with the Bravo Liquid Handling Platform

(Date:10/10/2008)...ss., Oct. 10 /- Boston Scie...t has received approval from the,U.S. Food and Dru...tion,TAXUS(R) Liberte(R) Paclitaxel-Eluting Corona...AXUS Liberte stent early next month in the United,...f its TAXUS(R),Expresss2(TM) Atom(TM) Paclitaxel-E...
(Date:10/10/2008)...iangsu, CHINA, Oct. 10 /Xinhua-- Beik...ch.com ) a global leader in,stem cell research an...nced,today it has signed eight separate mutual coo...ld,s most outstanding stem cell research organizat...tablished a "Stem Cell Expert,Advisory Committee" ...
(Date:10/10/2008)..., Del., Oct. 10 /- AstraZen... and Drug Administration (FDA) has,approved once-d...-Release,Tablets for the acute treatment of the de..., the manic and mixed episodes associated with bip...bipolar I disorder as adjunctive,therapy to lithiu...
(Date:10/10/2008)... Md., Oct. 10 /- EntreMed, ...cal company developing,therapeutics for the treatm... today that Thomas H. Bliss, Jr., EntreMed,s Senio...nt, will present a Company,overview at the 16th An...ng,held October 12-14, 2008 in London, United King...
Breaking Biology Technology:Video: Boston Scientific Announces FDA Approval of Second-Generation TAXUS(R) Liberte(R) Drug-Eluting Stent 2Video: Boston Scientific Announces FDA Approval of Second-Generation TAXUS(R) Liberte(R) Drug-Eluting Stent 3Stem Cell Leader Beike Signs 8 New Cooperation Agreements, Establishes Safety Monitoring Boards 2Stem Cell Leader Beike Signs 8 New Cooperation Agreements, Establishes Safety Monitoring Boards 3FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 2FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 3FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 4FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 5FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 6FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 7FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 8FDA Approves AstraZeneca's SEROQUEL XR(R) for the Treatment of Bipolar Depression and Bipolar Mania 9EntreMed to Present at BioPartnering Europe Conference 2EntreMed to Present at BioPartnering Europe Conference 3
...artered in Carlsbad, California. , Dynal : 400+ em...n Deer. Headquartered in Oslo, Norway. Brown Deer... to acquire Dynal Biotech ASA , whose U.S. operat...d of March for $380.4 million. , ,Invitrogen has ...pany, which sells supplies to the biotechnology an...
...rtant impact that generic pharmaceutical drugs are...world. Their rise and growth spurs Big Pharma to i...llion Americans not having any kind of medical hea...o cheaper medicines from Canada being rapidly shut...eaper medicines is becoming more and more apparent...
...ople in a significant way (like finding them a job...also appreciate little acts of kindness (like maki...and paying for meals). , ,While these details may ... how people perceive you and your networking style...cause you want to make a good impression. Over tim...
...ple high-energy neutrino particles from deep space...ctic ice. Photo courtesy of Daan Hubert/UW-Madison...conditions, an international team of scientists, e...t critical elements of a massive neutrino telescop...- in a 1.5 mile-deep hole drilled into the Antarct...
Other Biology Technology:Invitrogen to buy Norwegian biotech with operations in Brown Deer 2Pharmas will be forced into generic-drug strategy 2Pharmas will be forced into generic-drug strategy 3Little things can be big in networking 2Little things can be big in networking 3First pieces of Antarctic deep-space particle detector are in place 2First pieces of Antarctic deep-space particle detector are in place 3First pieces of Antarctic deep-space particle detector are in place 4First pieces of Antarctic deep-space particle detector are in place 5
(Date:10/10/2008)...ty: From Local to Global, will be the focus of the...ference, but this single event is just one small s...s a top priority around the world and requires imm...poverty, poor nutrition, hunger and deficiencies. ...es its members and the public to get involved and ...
(Date:10/9/2008)...RA, Calif., Oct. 9 /PRNewswire-FirstCall/ -- Synap...eveloper of human interface,solutions for mobile c...s,announced today that it will report financial r...rsday, October 23, 2008 after the close of market....call for analysts and,investors at 2:00 p.m. PT (5...
(Date:10/9/2008)...rchers at the European Molecular Biology Laborator... - the first complete developmental blueprint of a...entists could for the first time track all cells f...The data was reconstructed into a three-dimensiona..., published in the current online issue of Scienc...
(Date:10/9/2008)...yme may play a major role in anencephaly, offering...are fatal birth defect in which the brain fails to... the University of Illinois at Chicago College of ...ue of the journal of Molecular Endocrinology . ,...ith anencephaly each year. Most do not survive mor...
Breaking Biology News(10 mins):World Food Day brings attention to food security around the globe 2Synaptics to Report First Quarter Results on October 23 2Digital zebrafish embryo provides the first complete developmental blueprint of a vertebrate 2Clue to genetic cause of fatal birth defect 2Biomonitoring 3055 2/ 1Mentice Inc Expands Its North American Sales and Clinical Presence With the Hiring of Jeff Sirek as Director of Educational Sales East and Sue Lands 17743 1Mentice Inc Expands Its North American Sales and Clinical Presence With the Hiring of Jeff Sirek as Director of Educational Sales East and Sue Lands 17743 2Multi generational Diversity Top Challenge for Businesses Today 3A Leading Company to Share Insights June 4 6 17739 1Multi generational Diversity Top Challenge for Businesses Today 3A Leading Company to Share Insights June 4 6 17739 2Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 1Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 2Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 3Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 4Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 5Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 6Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 7Universal Health Realty Income Trust Reports 2008 First Quarter Financial Results 17738 8
...r Emad Tajkhorshid and graduate student James Gumb... of I.,s National Institutes of Health Resource fo...instakingly model a critical part of a mechanism b...Photo by L. Brian Stauffer Using X-ray data and a...ware, researchers at the University of Illinois ha...
...oined the battle to save Australia’s iconic Tasman...ating devil populations. , Researchers from CSIRO’...aboratory (AAHL), Textiles and Fibre Technology an...n the cause of Devil Facial Tumour Disease (DFTD) ...nimals. , DFTD is infectious and is thought to...
... it does to an alpine ibex. , According to a team... ibex demonstrate a correlation between horn growt...e shown that greater genetic diversity correlates ... of the horns reliably advertises the genetic qual...Dr. David Coltman, an evolutionary geneticist at t...
...mson Family Cancer Research Institute of the Unive...ene important in embryo development leads to prema...lt mice. This gene, ATR, is essential for the body...ins in the DNA damage response underlie certain ty...work appears in the inaugural issue of Cell Stem C...
Other Biology News:Simulations unravel outer membrane transport mechanism 2Simulations unravel outer membrane transport mechanism 3Simulations unravel outer membrane transport mechanism 4CSIRO scientists join fight to save 'Tassie devil' 2Bigger horns equal better genes 2Loss of stem cells correlates with premature aging in animal study 2
Rabbit polyclonal to Dishevelled 2 ( Abpromise for all tested applications). entrezGeneID: 1856 SwissProtID: O14641
... Unconjugated Applications: ELISA: Use at an ...endent dilution. WB: Use at a concentration of 5 -...a whole extract of rat brain. Predicted molecular ... Optimal dilutions/concentrations should be determ...
SHP-2 Antibody
... antiserum has been tested on tissue sections from...obtained with AB1773 corresponds to the pattern de... GLYT2 mRNA and to what has been described for gly...ith the immunogen peptide completely abolishes the...
Biology Products: