Navigation Links
Selection of Standards for Bio-Plex Cytokine Assays, Rev A

Diana Davis, Aiguo Zhang, Jim Torrence, and Emily Dale, Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive, Hercules, CA 94547 USA


Introduction
Bio-Plex cytokine assays are bead-based quantitative multiplex immunoassays. The concentrations of analytes in these assays are quantitated using a calibration or standard curve. A regression analysis is performed to derive an equation that is used to predict the concentrations of unknown samples. The specific concentrations of the standards used to construct the standard curve have a direct effect on the quality of the results. Therefore, selection of appropriate standards is a critical component of immunoassay performance. In this tech note, the effect of the selection of standards on the accuracy, sensitivity, and dynamic range of the results is discussed. The effect on assay performance of reading assays on a Bio-Plex system set at two different photomultiplier tube (PMT) voltages as well as criteria for selecting a custom set of standards are also addressed.


Regression Analysis for Bio-Plex Assays
Because the typical response of an immunoassay is nonlinear, the nonlinear or logistic regression routine is the preferred method for sandwich immunoassay data analysis (Baud 1993). The four-parameter (4PL) and five-parameter (5PL) logistic equations contain parameters (the midpoint of the curve, the slope where the curve is linear, and the upper and lower points) that are used to predict the response of the assay (relative median fluorescence intensity, or MFI; Figure 1). Use of logistic regression yields the widest range of predictable concentrations for Bio-Plex immunoassays (Nix and Wild 2001).


Standard Curve Ranges for Bio-Plex Assays
Two sets of standards are recommended for Bio-Plex cytokine assays, a high-sensitivity range (0.23,200 pg/ml) and a broad range (1.9532,000 pg/ml). The specific concentrations for the two sets of standards are shown in Table 1. The highsensitivity standards are recommended for use when the concentrations of the unknowns fall within the physiological range of cytokines in serum. The highsensitivity standards are suitable for quantitation of low levels of cytokines and provide the greatest sensitivity. The broad range standards yield a wide dynamic range and are used when quantitation of a wide range of cytokine concentrations is desired or when the range of expected concentrations of cytokines is unknown. The broad range standards enable the detection of the widest range of concentrations and yield a wide dynamic range. A comparison of the high-sensitivity and broad range standard curves plotted as concentration versus fluorescence intensity is shown in Figure 2.


Effect of Range on Standards Recovery
The overall dynamic range, accuracy, and sensitivity of results obtained using high-sensitivity vs. broad range standards are quite different. These differences may be illustrated using a method known as backcalculation of standards, or standards recovery (Davies 2001). In this method, a regression model is applied to a set of data and the resulting equation is used to calculate the concentrations of the standards using the response values (Nix and Wild 2001). The analysis is performed by calculating the concentration of each standard and then comparing it to the actual concentration using the following formula: (observed concentration) / (expected concentration) x 100. If the regression analysis is accurate, the calculated concentrations will be very close to the actual concentrations of the standards. This method yields information about the dynamic range, sensitivity, and accuracy of calculated concentrations of an assay (Baud 1993). The recovery value for each standard is an indication of the accuracy of the calculated concentration. A value within the range of 70130% is considered accurate recovery. The dynamic range of the assay, or the range where accurate concentrations may be quantitated, is therefore indicated by the range of standards yielding a recovery of 70130%.

A comparison of the standards recovery using broad range and high-sensitivity standards for three different cytokines is shown in Tables 2 and 3. The broad range standard concentrations where recovery is 70130% is 1.95 to 8,000 or 32,000 pg/ml, depending on the cytokine assayed. The high-sensitivity standards yield a narrower dynamic range of 0.2 or 0.78 to 3,200 pg/ml. In all three assays, the broad range standards enable the quantitation of a much wider range of concentrations.

A general indication of assay sensitivity is given by the lowest concentration in a set of standards that yields accurate recovery (70130%). Since the lowest concentration of the high-sensitivity standards is lower than the broad range standards (i.e., 0.2 pg/ml vs. 1.95 pg/ml, respectively), it would be expected that the high-sensitivity standards would yield better recovery at a lower concentration, and that the high-sensitivity standards should be chosen to maximize sensitivity in an assay. Tables 2 and 3 illustrate some typical recovery values for broad range and high-sensitivity standards.


Effect of PMT Voltage on Assay Performance
Bio-Plex assays are analyzed on a Bio-Plex system that contains a PMT. The PMT detects assay signals in the reporter or RP1 channel. The PMT voltage on the Bio-Plex system is automatically adjusted during CAL2 calibration using the RP1 target value from a bead with stable fluorescence. This method yields a reproducible signal output from day to day and between instruments. Two RP1 target values are recommended for use with Bio-Plex assays, a low target value for the broad range standards and a high target value for the high-sensitivity standards. Following calibration, the low RP1 target value yields a low PMT voltage and the high RP1 target value yields a higher PMT voltage. Table 4 compares the resulting PMT voltage on a Bio-Plex system following calibration with the low and high target values.

The PMT voltage is increased in assays in which higher sensitivity is desired. However, when using the broad range standards, the signals of the high concentration standards on certain cytokines may become saturated and result in a reduced dynamic range. Figures 3 and 4 show a broad range standard IL-8 assay analyzed with the high and low PMT settings. The standard points are labeled with the corresponding recovery. In Figure 3, the recovery at the two highest points is beyond the range of the regression, due to a saturated signal at the high end of the standard curve, and thus OOR (out of range) is displayed. When the same assay is analyzed at the low PMT setting, there is no saturation and the standards recovery is accurate across the entire range of standards (Figure 4).

The high-sensitivity standards yield a much lower fluorescence intensity signal since the concentration range is much lower. When using the low range standards, reading the assay with the high PMT voltage may improve the sensitivity. Table 5 shows a comparison of the recovery of a low range GM-CSF assay run using the low and high PMT voltage settings. The low PMT setting yielded acceptable recovery down to only 3.13 pg/ml. When the same assay was run at the high PMT setting, accurate recovery down to 0.2 pg/ml was achieved.


Guidelines for Running Cytokine Assays
The overall assay performance must be taken into consideration when determining how to set up both the cytokine assay standards and the Bio-Plex array reader. In Table 6, the guidelines for optimizing the Bio-Plex system (including assays and the array reader) are shown. The standards as well as the PMT voltage chosen for reading the assay may significantly affect results.


Selecting a Custom Set of Standards
The two sets of standards and RP1 target values discussed in this tech note have been validated using the Bio-Plex system and have been shown to yield optimal results. Efficient assay performance may also be achieved using a custom set of standards as long as a few simple rules are followed.

1. Select a minimum of 8 standard curve points.
The logistic regression model requires a minimum of 5 standards for the 4PL and 6 standards for the 5PL to fulfill the analysis requirements (Motulsky 1996). A total of 8 points are recommended to ensure an accurate fit, with a maximum of 12 points (Baud 1993). This practice allows the flexibility to remove standard points with erroneous results due to errors in assay preparation.

2. Choose standards evenly distributed across the entire range of concentrations.
As noted earlier, the response of a Bio-Plex assay is nonlinear; there is a flattening of the response near zero and at the high end of the curve with a linear portion in the middle as shown in Figure 2. Due to the nature of this response, the chosen standards concentrations should be evenly distributed across the entire range (Nix and Wild 2001). The regression routine will then yield the most accurate representation of the assay response.

3. Choose a 5PL regression routine.
When conditions are optimal, the upper and lower portions of the curve are symmetrical, and the 4PL logistic routine is appropriate. In reality, many conditions, such as a high-dose hook effect and inadequate washing, can cause asymmetry in the curve (Nix and Wild 2001). The 5PL regression contains an additional parameter to accommodate asymmetry in the curve, and thus is the more robust type of logistic regression for Bio-Plex assays.

4. Prepare standards using serial dilutions.
When selecting the range of a custom set of standards, determine the lowest and highest desired concentrations, then determine what dilution series should be used to yield a total of 8 points within the range. For example, if the low and high desired values are 5 and 10,000 pg/ml respectively, a threefold dilution series would yield 8 points (10,000, 3,333, 1,111, 370, 123.4, 41.1, 13.7, and 4.6). A dilution series is simple to prepare and less prone to errors compared to preparation of individual concentrations.

5. Determine which PMT setting is optimal.
Read the assay on the Bio-Plex array reader after calibrating with the low RP1 target value, then the high value, to determine which setting yields the best results. An assay plate may be read a second time by aspirating the contents of the wells after the first reading and resuspending in 125 l of buffer.


Summary
Bio-Plex assays may be run with a variety of standard curve concentrations. Two sets of concentrations are recommended, a high-sensitivity range for maximum sensitivity or for samples such as serum that fall within the biological range of cytokines, and a broad range for samples where a wide range of cytokine concentrations are expected. The voltage of the reporter PMT may also be varied to maximize assay performance. The high-sensitivity standards should be run following calibration with the high RP1 target value and the broad range standards perform best following calibration with the low RP1 target value. Overall, the highsensitivity standards yield better sensitivity, while maximum dynamic range is achieved using the broad range standards. Excellent assay performance may also be achieved using a unique set of standards if a few simple rules are followed: 1) select a minimum of 8 standard curve concentrations, 2) select a set of concentrations that are evenly distributed across the entire selected range, 3) select a 5PL regression in Bio-Plex Manager software for the most robust assay analysis, 4) prepare the standard curve points using a serial dilution method, and 5) determine which PMT setting is optimal.


References
Baud M, Data analysis, mathematical modeling, pp 656671 in (Masseyeff RF et al., eds), Methods of Immunological Analysis Volume 1: Fundamentals , VCH Publishers, Inc., New York, NY (1993)

Davies C, Concepts, pp 78110 in (David Wild, ed), The Immunoassay Handbook, 2nd ed, Nature Publishing Group, New York, NY (2001)

Motulsky H, The GraphPad guide to nonlinear regression, in GraphPad Prism Software User Manual, GraphPad Software Inc., San Diego, CA (1996)

Nix B and Wild D, Calibration curve-fitting, pp 198210 in (David Wild, ed), The Immunoassay Handbook, 2nd ed , Nature Publishing Group, New York, NY (2001)


back to top
'"/>

Source:


Page: All 1 2 3 4 5 6 7 8

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Plasmid Miniprep Kit Selection
3. Universal Guide to DNA Standards
4. Bio-Plex Cytokine Immunoassays and ELISA: Comparison of Two Methodologies in Testing Samples From Asthmatic and Healthy Children, Rev A
5. Detection of Phosphorylated and Total ERK and p38 MAPK by Bio-Plex Assay, ELISA, and Western Blotting, Rev A
6. Correlation of Bio-Plex Suspension Array Reader Validation With Multiplex Cytokine Assay Performance, Rev A
7. Application of the Bio-Plex Cell Lysis Kit in Western Blot Analysis of Phosphoproteins, Rev A
8. Development of a Multiplex Bead-Based Assay for Antibody Screening of a Nonhuman Primate Colony on the Bio-Plex System
9. Feasibility of Multiplexing Bio-Plex Total Target and Phosphoprotein Assays
10. Simultaneous Detection of 8 Cytokines in Mouse or Human Sera
11. The Gradient Feature: Use in Optimization of Allelic Discrimination Assays, Rev B
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Selection Standards for Bio Plex Cytokine Assays Rev

(Date:8/28/2014)... August 28, 2014 Supplementing their ... a promotion for Eppendorf Safe Lock Centrifuge ... purchase the highest quality tubes at an affordable ... for customers doing chemical, medical, pharmaceutical, and life ... are not regularly able to afford Eppendorf products. ...
(Date:8/27/2014)... Green & Grow Inc. (GGI) has raised an ... Otter Capital as a significant new partner. Otter Capital’s ... Agriplier™ technology, building on recent compelling field trial results ... meeting, we have been impressed with Otter Capital’s knowledge ... Sobba, President and CEO of GGI. “We look forward ...
(Date:8/27/2014)... , Aug. 27, 2014 Reportlinker.com ... is available in its catalogue: ... http://www.reportlinker.com/p02316661/Global-Biodegradable-Packaging-Materials-Market-2014-2018.html ... materials that are readily decomposed by the ... conventional non-degradable packaging in terms of raw ...
(Date:8/27/2014)... TX (PRWEB) August 27, 2014 The ... and in-depth study on the current state of the ... report provides a basic overview of the industry, including ... and Chinese domestic market analysis are provided with a ... the market. A comparison between the international and Chinese ...
Breaking Biology Technology:Eppendorf Safe Lock Centrifuge Tube Promotion Now Available at Major Pipette Distributor Pipette.com 2Eppendorf Safe Lock Centrifuge Tube Promotion Now Available at Major Pipette Distributor Pipette.com 3Green & Grow Inc. Secures $6M Series B Funding from Otter Capital 2Green & Grow Inc. Secures $6M Series B Funding from Otter Capital 3Global Biodegradable Packaging Materials Market 2014-2018 2Global Biodegradable Packaging Materials Market 2014-2018 3Global Biodegradable Packaging Materials Market 2014-2018 4Global Biodegradable Packaging Materials Market 2014-2018 5Global Biodegradable Packaging Materials Market 2014-2018 6Global Biodegradable Packaging Materials Market 2014-2018 7Trifluoroacetic Acid Industry (Asia, Europe, North America, ROW) Analysis Report Now Available at ReportsnReports.com 2
... Madison, Wis . - If manufacturers are interested in new ... a conference that provides a learning and technology transfer opportunity. ... already puts on an annual design event at the ... expanded the program with the intention of opening doors to ...
... some one thousand theme-based cities being developed at a phenomenal ... name identities of such projects become nightmares. In size, they ... large dwelling. , ,Now this requires a new definition of ... other traditional metropolises. , ,For example, the introduction of ...
... reported net income of $113.5 million during the first quarter ... rising 11 percent over the same quarter in 2006. , ... industry, said net income was down from the $116.2 million ... share were slightly higher than the $0.64 in the same ...
Cached Biology Technology:Branding around the community brand concept 2Branding around the community brand concept 3
(Date:8/28/2014)... are good examples of complex systems. Their infrastructure ... water for cooling, transport to supply fuel, and ... in the network chain is interconnected with a ... A team of UK-based scientists has studied various ... findings have been published in EPJ B by ...
(Date:8/28/2014)... health risk to people with asthma according to a new ... Immunology . , By critically reviewing the findings from 17 ... the presence of several types of mould can lead to ... likelihood of developing the condition. , The research has been ... School and is the first time all of the information ...
(Date:8/28/2014)... important crop-producing countries will be fully saturated with pests ... continue, according to a new study led by the ... can already be found in around half the countries ... at its current rate, scientists fear that a significant ... pests within the next 30 years. , Crop pests ...
Breaking Biology News(10 mins):Inter-dependent networks stress test 2Indoor mold poses health risk to asthma sufferers 2New study charts the global invasion of crop pests 2
... new study published in the Journal of Trauma ... (DP) stimulation in spinal cord-injured patients is successful not ... in bridging patients to independent respiration, where they could ... ventilator or stimulation. The stimulation is provided by ...
... , (Santa Barbara, Calif.) As the Endangered Species Act (ESA) ... agency, the U.S. Fish and Wildlife Service (USFWS), was mired in ... ( Canis lupus ) from the List of Endangered and Threatened ... baileyi ). As a result, the USFWS sought an ...
... the right shape might be just as important in a ... an important role in the effectiveness of cells grown to ... design new structures that enable cells to "shape up," researchers ... come up with a way to measure, and more importantly, ...
Cached Biology News:Diaphragm pacing in spinal cord injury successful in weaning patients from ventilators 2Diaphragm pacing in spinal cord injury successful in weaning patients from ventilators 3Panel issues report on gray wolf science 2Shape-sifting: NIST categorizes bio scaffolds by characteristic cell shapes 2
... These handy dispensers save space and ... and other garb. They are the ... parts used in labs, processing and ... charges and the particles they attract. ...
...
... Stoelting, selects single cells, or clusters, with ... at all, a single cell can be ... is collected into the pipette, and automatically ... microtiter plate. Quixells unique aspiration and expulsion ...
... for genome-wide microRNA expression profiling and built ... on-chip synthesis platform. These microarrays are available ... Profiling Service. Probe Content ... cow microRNAs (108 unique mature sequences). The ...
Biology Products: