Selection of Standards for Bio-Plex Cytokine Assays, Rev A

Diana Davis, Aiguo Zhang, Jim Torrence, and Emily Dale, Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive, Hercules, CA 94547 USA


Introduction
Bio-Plex cytokine assays are bead-based quantitative multiplex immunoassays. The concentrations of analytes in these assays are quantitated using a calibration or standard curve. A regression analysis is performed to derive an equation that is used to predict the concentrations of unknown samples. The specific concentrations of the standards used to construct the standard curve have a direct effect on the quality of the results. Therefore, selection of appropriate standards is a critical component of immunoassay performance. In this tech note, the effect of the selection of standards on the accuracy, sensitivity, and dynamic range of the results is discussed. The effect on assay performance of reading assays on a Bio-Plex system set at two different photomultiplier tube (PMT) voltages as well as criteria for selecting a custom set of standards are also addressed.


Regression Analysis for Bio-Plex Assays
Because the typical response of an immunoassay is nonlinear, the nonlinear or logistic regression routine is the preferred method for sandwich immunoassay data analysis (Baud 1993). The four-parameter (4PL) and five-parameter (5PL) logistic equations contain parameters (the midpoint of the curve, the slope where the curve is linear, and the upper and lower points) that are used to predict the response of the assay (relative median fluorescence intensity, or MFI; Figure 1). Use of logistic regression yields the widest range of predictable concentrations for Bio-Plex immunoassays (Nix and Wild 2001).


Standard Curve Ranges for Bio-Plex Assays
Two sets of standards are recommended for Bio-Plex cytokine assays, a high-sensitivity range (0.23,200 pg/ml) and a broad range (1.9532,000 pg/ml). The specific concentrations for the two sets of standards are shown in Table 1. The highsensitivity standards are recommended for use when the concentrations of the unknowns fall within the physiological range of cytokines in serum. The highsensitivity standards are suitable for quantitation of low levels of cytokines and provide the greatest sensitivity. The broad range standards yield a wide dynamic range and are used when quantitation of a wide range of cytokine concentrations is desired or when the range of expected concentrations of cytokines is unknown. The broad range standards enable the detection of the widest range of concentrations and yield a wide dynamic range. A comparison of the high-sensitivity and broad range standard curves plotted as concentration versus fluorescence intensity is shown in Figure 2.


Effect of Range on Standards Recovery
The overall dynamic range, accuracy, and sensitivity of results obtained using high-sensitivity vs. broad range standards are quite different. These differences may be illustrated using a method known as backcalculation of standards, or standards recovery (Davies 2001). In this method, a regression model is applied to a set of data and the resulting equation is used to calculate the concentrations of the standards using the response values (Nix and Wild 2001). The analysis is performed by calculating the concentration of each standard and then comparing it to the actual concentration using the following formula: (observed concentration) / (expected concentration) x 100. If the regression analysis is accurate, the calculated concentrations will be very close to the actual concentrations of the standards. This method yields information about the dynamic range, sensitivity, and accuracy of calculated concentrations of an assay (Baud 1993). The recovery value for each standard is an indication of the accuracy of the calculated concentration. A value within the range of 70130% is considered accurate recovery. The dynamic range of the assay, or the range where accurate concentrations may be quantitated, is therefore indicated by the range of standards yielding a recovery of 70130%.

A comparison of the standards recovery using broad range and high-sensitivity standards for three different cytokines is shown in Tables 2 and 3. The broad range standard concentrations where recovery is 70130% is 1.95 to 8,000 or 32,000 pg/ml, depending on the cytokine assayed. The high-sensitivity standards yield a narrower dynamic range of 0.2 or 0.78 to 3,200 pg/ml. In all three assays, the broad range standards enable the quantitation of a much wider range of concentrations.

A general indication of assay sensitivity is given by the lowest concentration in a set of standards that yields accurate recovery (70130%). Since the lowest concentration of the high-sensitivity standards is lower than the broad range standards (i.e., 0.2 pg/ml vs. 1.95 pg/ml, respectively), it would be expected that the high-sensitivity standards would yield better recovery at a lower concentration, and that the high-sensitivity standards should be chosen to maximize sensitivity in an assay. Tables 2 and 3 illustrate some typical recovery values for broad range and high-sensitivity standards.


Effect of PMT Voltage on Assay Performance
Bio-Plex assays are analyzed on a Bio-Plex system that contains a PMT. The PMT detects assay signals in the reporter or RP1 channel. The PMT voltage on the Bio-Plex system is automatically adjusted during CAL2 calibration using the RP1 target value from a bead with stable fluorescence. This method yields a reproducible signal output from day to day and between instruments. Two RP1 target values are recommended for use with Bio-Plex assays, a low target value for the broad range standards and a high target value for the high-sensitivity standards. Following calibration, the low RP1 target value yields a low PMT voltage and the high RP1 target value yields a higher PMT voltage. Table 4 compares the resulting PMT voltage on a Bio-Plex system following calibration with the low and high target values.

The PMT voltage is increased in assays in which higher sensitivity is desired. However, when using the broad range standards, the signals of the high concentration standards on certain cytokines may become saturated and result in a reduced dynamic range. Figures 3 and 4 show a broad range standard IL-8 assay analyzed with the high and low PMT settings. The standard points are labeled with the corresponding recovery. In Figure 3, the recovery at the two highest points is beyond the range of the regression, due to a saturated signal at the high end of the standard curve, and thus OOR (out of range) is displayed. When the same assay is analyzed at the low PMT setting, there is no saturation and the standards recovery is accurate across the entire range of standards (Figure 4).

The high-sensitivity standards yield a much lower fluorescence intensity signal since the concentration range is much lower. When using the low range standards, reading the assay with the high PMT voltage may improve the sensitivity. Table 5 shows a comparison of the recovery of a low range GM-CSF assay run using the low and high PMT voltage settings. The low PMT setting yielded acceptable recovery down to only 3.13 pg/ml. When the same assay was run at the high PMT setting, accurate recovery down to 0.2 pg/ml was achieved.


Guidelines for Running Cytokine Assays
The overall assay performance must be taken into consideration when determining how to set up both the cytokine assay standards and the Bio-Plex array reader. In Table 6, the guidelines for optimizing the Bio-Plex system (including assays and the array reader) are shown. The standards as well as the PMT voltage chosen for reading the assay may significantly affect results.


Selecting a Custom Set of Standards
The two sets of standards and RP1 target values discussed in this tech note have been validated using the Bio-Plex system and have been shown to yield optimal results. Efficient assay performance may also be achieved using a custom set of standards as long as a few simple rules are followed.

1. Select a minimum of 8 standard curve points.
The logistic regression model requires a minimum of 5 standards for the 4PL and 6 standards for the 5PL to fulfill the analysis requirements (Motulsky 1996). A total of 8 points are recommended to ensure an accurate fit, with a maximum of 12 points (Baud 1993). This practice allows the flexibility to remove standard points with erroneous results due to errors in assay preparation.

2. Choose standards evenly distributed across the entire range of concentrations.
As noted earlier, the response of a Bio-Plex assay is nonlinear; there is a flattening of the response near zero and at the high end of the curve with a linear portion in the middle as shown in Figure 2. Due to the nature of this response, the chosen standards concentrations should be evenly distributed across the entire range (Nix and Wild 2001). The regression routine will then yield the most accurate representation of the assay response.

3. Choose a 5PL regression routine.
When conditions are optimal, the upper and lower portions of the curve are symmetrical, and the 4PL logistic routine is appropriate. In reality, many conditions, such as a high-dose hook effect and inadequate washing, can cause asymmetry in the curve (Nix and Wild 2001). The 5PL regression contains an additional parameter to accommodate asymmetry in the curve, and thus is the more robust type of logistic regression for Bio-Plex assays.

4. Prepare standards using serial dilutions.
When selecting the range of a custom set of standards, determine the lowest and highest desired concentrations, then determine what dilution series should be used to yield a total of 8 points within the range. For example, if the low and high desired values are 5 and 10,000 pg/ml respectively, a threefold dilution series would yield 8 points (10,000, 3,333, 1,111, 370, 123.4, 41.1, 13.7, and 4.6). A dilution series is simple to prepare and less prone to errors compared to preparation of individual concentrations.

5. Determine which PMT setting is optimal.
Read the assay on the Bio-Plex array reader after calibrating with the low RP1 target value, then the high value, to determine which setting yields the best results. An assay plate may be read a second time by aspirating the contents of the wells after the first reading and resuspending in 125 l of buffer.


Summary
Bio-Plex assays may be run with a variety of standard curve concentrations. Two sets of concentrations are recommended, a high-sensitivity range for maximum sensitivity or for samples such as serum that fall within the biological range of cytokines, and a broad range for samples where a wide range of cytokine concentrations are expected. The voltage of the reporter PMT may also be varied to maximize assay performance. The high-sensitivity standards should be run following calibration with the high RP1 target value and the broad range standards perform best following calibration with the low RP1 target value. Overall, the highsensitivity standards yield better sensitivity, while maximum dynamic range is achieved using the broad range standards. Excellent assay performance may also be achieved using a unique set of standards if a few simple rules are followed: 1) select a minimum of 8 standard curve concentrations, 2) select a set of concentrations that are evenly distributed across the entire selected range, 3) select a 5PL regression in Bio-Plex Manager software for the most robust assay analysis, 4) prepare the standard curve points using a serial dilution method, and 5) determine which PMT setting is optimal.


References
Baud M, Data analysis, mathematical modeling, pp 656671 in (Masseyeff RF et al., eds), Methods of Immunological Analysis Volume 1: Fundamentals , VCH Publishers, Inc., New York, NY (1993)

Davies C, Concepts, pp 78110 in (David Wild, ed), The Immunoassay Handbook, 2nd ed, Nature Publishing Group, New York, NY (2001)

Motulsky H, The GraphPad guide to nonlinear regression, in GraphPad Prism Software User Manual, GraphPad Software Inc., San Diego, CA (1996)

Nix B and Wild D, Calibration curve-fitting, pp 198210 in (David Wild, ed), The Immunoassay Handbook, 2nd ed , Nature Publishing Group, New York, NY (2001)


back to top
'"/>

Source:


Page: All 1 2 3 4 5 6 7 8

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Plasmid Miniprep Kit Selection
3. Universal Guide to DNA Standards
4. Bio-Plex Cytokine Immunoassays and ELISA: Comparison of Two Methodologies in Testing Samples From Asthmatic and Healthy Children, Rev A
5. Detection of Phosphorylated and Total ERK and p38 MAPK by Bio-Plex Assay, ELISA, and Western Blotting, Rev A
6. Correlation of Bio-Plex Suspension Array Reader Validation With Multiplex Cytokine Assay Performance, Rev A
7. Application of the Bio-Plex Cell Lysis Kit in Western Blot Analysis of Phosphoproteins, Rev A
8. Development of a Multiplex Bead-Based Assay for Antibody Screening of a Nonhuman Primate Colony on the Bio-Plex System
9. Feasibility of Multiplexing Bio-Plex Total Target and Phosphoprotein Assays
10. Simultaneous Detection of 8 Cytokines in Mouse or Human Sera
11. The Gradient Feature: Use in Optimization of Allelic Discrimination Assays, Rev B
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Selection Standards for Bio Plex Cytokine Assays Rev

(Date:8/31/2015)... -- Research and Markets ( http://www.researchandmarkets.com/research/53wxp3/israel_biomedical ) has announced ... - Growth, Trends & Forecasts (2014-2020)" report to ... market is estimated at $0.19 billion by 2018 at ... The near future will bring biomedical sensors that ... These sensors can be set to trigger alerts when ...
(Date:8/31/2015)... , Aug. 31, 2015 /PRNewswire/ - Portage Biotech ... Securities Exchange: PBT.U), is pleased to announce that its ... advancing its lead candidate, PPL-003, to an Investigational New ... eye disease and uveitis.  PPL recently completed a study ... topical PPL-003 solution achieved highly significant efficacy and a ...
(Date:8/31/2015)... HAYWARD, Calif. and SHANGHAI ... Inc., a privately-held biopharmaceutical company developing next-generation antibiotics, ... Phase 2 clinical study for its lead drug ... designed to treat drug-resistant bacteria such as MRSA ... safer and better tolerated therapeutic option than currently ...
(Date:8/31/2015)... -- Trimb Healthcare AB ("Trimb") today announced that it ... ("YouMedical"), a Dutch OTC company with operations in key ... EUR 16 million, with solid growth and leading brands ... to build a leading international OTC company, and this ... gives us a stronger North European OTC platform, from ...
Breaking Biology Technology:Israel Biomedical Sensors Market 2015 - Growth, Trends & Forecasts to 2020 2Portage's PPL will advance its lead candidate PPL-003 to an IND for dry eye disease and uveitis 2Portage's PPL will advance its lead candidate PPL-003 to an IND for dry eye disease and uveitis 3MicuRX Reports Positive Top-Line Results In Phase 2 Clinical Trial For Novel Antibiotic MRX-I In Complicated Skin And Soft Tissue Infections 2
... into why and how nanowires take the form they ... future electronic components. PhD student Peter Krogstrup from the ... the sensational new theoretical model, which is developed in ... One of the most important components in future electronic ...
... How would you like to store all the films ever ... Magnets made of just a few metallic atoms could make ... also recently been proposed as components for spintronics devices. There,s ... been seen to work at temperatures a few hairs above ...
... Group, Inc.  (OTC Bulletin Board: MDXG ), ... biomaterial-based products and bioimplants processed from human amniotic membrane, ... 31, 2010. Full Year 2010 ResultsIn ... with the calendar year. Its prior fiscal year was ...
Cached Biology Technology:Major advance in understanding how nanowires form 2Heavy metals open path to high temperature nanomagnets 2MiMedx Group Announces 2010 Results 2MiMedx Group Announces 2010 Results 3MiMedx Group Announces 2010 Results 4MiMedx Group Announces 2010 Results 5MiMedx Group Announces 2010 Results 6MiMedx Group Announces 2010 Results 7
(Date:7/31/2015)... , 31 de julio de 2015 ... (ICG-10,  www.icg-10.org ) del 22 al 25 de octubre ... La conferencia celebra su décimo aniversario este ... convertido en una de las reuniones anuales más influyentes ... es una de las conferencias científicas más dinámicas, entusiastas ...
(Date:7/31/2015)... 2015 La 10e Conférence internationale sur ... par le BGI du 22 au 25 octobre ... Chine. La conférence célèbre son 10e ... l,ICG est devenue l,une des réunions annuelles mondiales ... et c,est aussi la plus dynamique, la plus ...
(Date:7/31/2015)... , 31. Juli 2015 Die 10. internationale ... von BGI veranstaltet und findet vom 22. bis ... , statt. Die Konferenz ... ihrem Start 2006 ist die ICG weltweit zu ... geworden. Sie ist eines der dynamischsten, enthusiastischsten und ...
Breaking Biology News(10 mins):La 10ª Conferencia Internacional sobre Genómica (ICG-10) comienza en octubre 2La 10ª Conferencia Internacional sobre Genómica (ICG-10) comienza en octubre 3La 10e International Conference on Genomics (ICG-10) doit s'ouvrir en octobre 2La 10e International Conference on Genomics (ICG-10) doit s'ouvrir en octobre 3Die 10. internationale Konferenz zu Genomik (ICG-10) beginnt im Oktober 2Die 10. internationale Konferenz zu Genomik (ICG-10) beginnt im Oktober 3
... rubbed on the skin to reduce signs of aging and ... the first evidence of one of vitamin E,s normal body ... tears in the plasma membranes that protect cells from outside ... University researchers report in the journal Nature Communications ., ...
... is an adorable doe-eyed bat with a dark side it ... Tsang, a PhD student in ecology and evolutionary biology at The ... to a revolutionary way to help fund her research into how ... viruses can spread from flying foxes to humans by tracing the ...
... Researchers at the University of Delaware are examining tiny ... not only how these organisms adapt to the severe cold, ... The National Science Foundation study, led by Adam Marsh, ... Ocean, and Environment, also will compare the process of temperature ...
Cached Biology News:Scientists identify an innate function of vitamin E 2Crowdfunding science: Student raises cash online to follow a flying fox 2Crowdfunding science: Student raises cash online to follow a flying fox 3Will Antarctic worms warm to changing climate? 2Will Antarctic worms warm to changing climate? 3
Our Histology, Immunohistochemistry and tissue arrays services include samples processing, different type of staning, histollogical stains, image analysis, tissue arrays, immunohistochemistry, etc....
Mouse monoclonal antibody to PDK2 - pyruvate dehydrogenase kinase, isoenzyme 2...
... of an N-terminal FLAG® fusion and a ... from bicistronic mRNA. Two genes can ... transcription of a single message from the ... protein-protein interaction studies, multi-subunit proteins, and cloning ...
Mouse monoclonal antibody raised against a partial recombinant CALCOCO2. NCBI Entrez Gene ID = CALCOCO2...
Biology Products: