Quantitative Analysis of GFP in Plant Extracts Using the VersaFluor Fluorometer System

Adrian R. Elliott, Peer M. Schenk and Christopher P.L. Grof
CSIRO Tropical Agriculture, Brisbane, Queensland


Introduction
The Green Fluorescent Protein (GFP) from Aequorea victoria is a useful reporter gene in higher organisms, including plants and animals. The fluorescence emitted is easy to detect and requires no substrate or cofactors. Furthermore, improved GFP versions have been created with higher fluorescence coefficients, improved expression characteristics for eukaryotes, or altered spectral properties. One form, GFP-S65T incorporates a threonine amino acid substitution in place of serine at position 65, which results in red-shifted fluorescence characteristics.1 This form is one that has been proven to be very beneficial for use in plants.

Quantification of GFP is unlike widely used enzyme-based reporters such as β-glucuronidase (GUS) and firefly luciferase (LUC). We show here that fluorescence detection using the VersaFluor fluorometer system with appropriate narrow band filters can be used to detect and quantify GFP-S65T protein levels in plant extracts against a standard curve made from purified GFP-S65T protein.


Method
The test samples consisted of sugarcane cv Q117 callus expressing Ubi1-sgfpS65T-nos3. Approximately 500 mg of tissue was ground in extraction buffer (10 mM Tris-EDTA, pH 8.0, 0.02% [w/v] sodium azide) at 4 C. This extract was centrifuged in a microcentrifuge at 13,000 rpm for 20 min. and the supernatant removed for analysis. Fluorescence microscopy showed that all visible GFP fluorescence was located in the soluble fraction. Protein concentrations of the plant extracts were deter mined using a Bio-Rad Protein Assay Kit. Plant samples were diluted to 400 g/ml of extractable protein in buffer. Two milliliters of diluted extract were used for each reading. A dilution series (0.10.4 g/ml) of purified rGFP-S65T protein (Clontech), was made in either extraction buffer or 400 g/ml untransformed callus extract.

The VersaFluor fluorometer was fitted with a 490 5 nm excitation filter and a 510 5 nm emission filter so that the reflective face of the filters faced towards the incident beam. The VersaFluor fluorometer was used according to the supplied instructions and blanked using 2 ml of extraction buffer. The range (on medium gain) was set using the highest plant sample. All samples and standards were measured by taking the mean of measurements at 20, 40 and 60 seconds.


Results and Discussion
Figure 1 shows the RFUs for the GFP dilution series, quantified using the VersaFluor fluorometer. The presence of other fluorescent compounds in the plant extracts did not increase the RFUs significantly, indicating that the 10 nm bandpass emission filter adequately omitted fluorescence at other wavelengths.

The curve for RFUs against rGFP-S65T concentration was linear over the range of plant GFP concentrations measured. Measurements were made after the optical sensor in the VersaFluor had stabilized. An example of stabilization of the fluorescence readings is shown for a 0.2 g/ml rGFP-S65T protein standard in Figure 2. For most readings, the signal stability varied less than 10% of the original value over 180 seconds. Plant samples were then read as for the GFP-S65T dilution series. GFP concentration was extrapolated from the standard curve corresp onding to the dilution series in untransformed calli extract to account for the background fluorescence. GFP levels in calli expressing Ubi-sgfpS65Tnos3 ranged up to 0.525 g rGFP-S65T per mg of extractable protein.

The ability to quantify GFP in extracts from transgenic organisms facilitates the comparative assessment of expression levels directed and influenced by attached genetic sequences, such as promoters, introns and terminators. The ability to test gene sequences in this way is a prerequisite for the correct deployment of useful gene sequences in transgenic plants.


Reference
1. Heim, R., Cubitt, A. B. and Tsien, R. Y. Nature, 373, 663-664 (1995).


back to top
'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. Mx4000 Multiplex Quantitative PCR System*,PPP
2. Improved Quantitative Selectivity of Clenbuterol in Human Urine using High Resolution on the TSQ Quantum Mass Spectrometer
3. The LightCycler is the only System for Quantitative PCR that is as Flexible as You are!
4. Quantitative Measurement of Cell Proliferation Using the BrdU ELISA: A Comparison Between Colorimetric and Chemiluminescent Detection
5. Quantitative end-point RT-PCR gene expression measurement using the Agilent 2100 Bioanalyzer and standardized RT-PCR
6. Detection of Genetically Modified Soybean in Processed Foods Using Real-Time Quantitative PCR with SYBR Green I Dye on the DNA Engine Opticon 2 System
7. Qualitative and Quantitative Assessment of Plasmid DNA Isolation, Rev A
8. Quantitative Analysis of Protein-DNA Associations in Vivo Using Real-Time PCR, Rev A
9. Rapid, Reproducible Real-Time Quantitative RT-PCR Using the iCycler iQ Real-Time PCR Detection System and iQ Supermix, Rev A
10. Multiplexed Quantitative Peptide Assays for Protein Biomarkers of Cardiovascular Disease in Human Plasma
11. Quantitative Analysis of the Cell Cycle in Imaging Applications
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:6/23/2016)... June 22, 2016  Amgen (NASDAQ: AMGN ... QB3@953 life sciences incubator to accelerate the ... shared laboratory space at QB3@953 was created to help ... obstacle for many early stage organizations - access to ... sponsorship, Amgen launched two "Amgen Golden Ticket" awards, providing ...
(Date:6/22/2016)... , June 22, 2016 Research and ... Global Markets" report to their offering. ... billion in 2014 from $29.3 billion in 2013. The market is ... of 13.8% from 2015 to 2020, increasing from $50.6 billion in ... projected product forecasts during the forecast period (2015 to 2020) are ...
(Date:6/22/2016)... ... June 22, 2016 , ... Quantitative Radiology Solutions, ... and current participant in the Phase 1 Ventures program, is leveraging regional and ... Quantitative Radiology Solutions helps physicians make better treatment decisions by quantifying medical imaging ...
(Date:6/22/2016)... ... 22, 2016 , ... The Immigrant Journey Awards , ... to North Texas and the nation, recently held its annual luncheon program. ... civic and economic vitality of North Texas. Proceeds from the event are used ...
Breaking Biology Technology:
(Date:5/16/2016)...   EyeLock LLC , a market leader of ... an IoT Center of Excellence in Austin, ... of embedded iris biometric applications. EyeLock,s iris ... security with unmatched biometric accuracy, making it the most ... EyeLock,s platform uses video technology to deliver a fast ...
(Date:5/3/2016)... , May 3, 2016  Neurotechnology, a provider ... MegaMatcher Automated Biometric Identification System (ABIS) , ... multi-biometric projects. MegaMatcher ABIS can process multiple complex ... any combination of fingerprint, face or iris biometrics. ... SDK and MegaMatcher Accelerator , which ...
(Date:4/26/2016)... and LONDON , April ... part of EdgeVerve Systems, a product subsidiary of ... today announced a partnership to integrate the Onegini ...      (Logo: http://photos.prnewswire.com/prnh/20151104/283829LOGO ) ... their customers enhanced security to access and transact ...
Breaking Biology News(10 mins):