Wenda Ribeiro and Margaret Saha, Department of Biology, College of William and Mary, Williamsburg, Virginia 23187
Introduction
DNA sequence analysis is an integral part of genetic research. Recently,
the use of automated sequencers has become increasingly more common in
small research laboratory settings. However, the success of automated
long read sequencing reactions require precise picomolar concentrations
of DNA. We have successfully employed the VersaFluor fluorometer (Bio-Rad)
to accurately and routinely quantify DNA for sequencing with the LiCor
4200 automated sequencer.
Materials and Methods
A genomic clone for xHif (Xenopus hypoxia inducible factor) was isolated
by a homology screen of a Xenopus muscle genomic library using a 32P-dCTP
labeled mouse probe (a gift of Steven McKnight). Phage DNA was isolated
using a standard protocol, digested with EcoRI and HindIII and run on
a 1.2% agarose gel (Figure 1). Bands corresponding to insert DNA were
excised and the DNA isolated on glass milk. The DNA fragments were subcloned
into BlueScript SK+ (Stratagene) and transformed into DH5a cells. Plasmid
DNA was isolated using anion exchange chromatography. A λZap (Stratagene)
cDNA clone for Xenopus xHif was also isolated using a homology screen
of a tadpole library. Rescued plasmid DNA was obtained using the manufacturers
instructions and was subsequently transformed into DH5α cells. Plasmid
DNA was isolated using anion exchange chromatography.
Quantitation of the DNA was achieved using the Fluorescent DNA Quantitation Kit (Bio-Rad) with a 360 nm excitation filter and a 460 nm emission filter. The standard curve was established for quantification of DNA concentrations ranging from 1005,000 g/ml by plotting seven points of varying DNA concentrations made by serial dilution of control calf thymus DNA. All dilutions were made in 2 ml of 1 g/ml Hoechst 33258 dye and included 0 ng DNA as a blank and 10,000 ng DNA as the upper range. The values were recorded in relative fluorescence units (RFUs) and were plotted on a graph with the x axis = DNA concentration (ng) and the y axis = RFU (Figure 2). The slope of the line was determined to be 1.0.
By employing the equation y = mx+b, we established the value of b to be equal to 0. Therefore, by setting the upper range value at 10,000 RFU using 10 l of 1 mg/ml calf thymus DNA and using 2 l of the sample DNA in 2 ml of the 1 g/ml Hoechst dye, the RFU reading is equal to ng/ml for our sample. We are then able to calculate the volume of DNA needed to obtain a 250 femtamolar concentration per reaction.
Results and Discussion
Using the Excel II sequencing kit (Epicenter) and quantifying our DNA
with the VersaFluor fluorometer, we have been able to achieve highly readable
and reproducible sequencing runs of up to 1,200 bp with the LiCor 4200
automated sequencer. This has enabled us to sequence the entire 2.2 Kb
cDNA for Xenopus hypoxia inducible factor, xHIF, in several runs, and
will allow us to sequence our genomic clone easily and efficiently.
The correct concentration of DNA in each reaction is the key to obtaining good reads on the LiCor 4200. This entails precise quantitation of the DNA. The VersaFluor fluorometer has proven to be easy to calibrate and provides excellent and reproducible results.
This work is supported by National Institute of Health Grant 1815HD32080 and National Science Foundation Grant 1BN9553176.