Quality control of antibodies using the,,,2100 bioanalyzer and the Protein 200,,,Plus assay

Antibodies are commonly used in diagnostics, as research tools (for example protein arrays) or as biopharmaceutical therapeutics. In all cases the successful production of antibodies requires precise testing and quality control. The key quality criteria include: protein identification, quantitation and the monitoring of purity and stability. Agilent's 2100 bioanalyzer, the first commercial lab-on-a-chip analysis tool, developed in collaboration with Caliper Technologies Corporation, was used to verify the suitability of this technology for antibody quality control. The system allows sizing and analysis proteins of 5 to 200 kDa, depending on the application. In addition, it determines the relative quantitation based on internal standards for each sample, or absolute quantitation based on user-defined calibration standards. The microfabricated chips with distinct microfluidic channels allows antibody analysis under both reducing and non-reducing conditions in a single run. In this Application Note we demonstrate the use of the system for the evaluation of a stress stability test of two different antibodies.
Experiments and Results
Antibody samples were kindly provided from a customer in the pharmaceutical industry. For the first sample a stress stability test was performed for one month at 40 C, for the second sample the stress test was done for 12 weeks at 40 C. The standard samples and the stress test samples were analyzed under reducing conditions on the Agilent 2100 bioanalyzer using a Protein 200 Plus LabChip kit and the dedicated Protein 200 Plus software assay. All chips were prepared according to the protocol provided with the Protein 200 Plus LabChip kit. The kit includes 25 chips, spin filters and all the reagents needed for the experiments including the Protein 200 Plus ladder as well as the upper and lower marker premixed in the sample buffer.
To validate the performance of the instrument three individual users measured the samples on two separate instruments using two lots of reagents on different days. Thus a total number of twelve chips were run. The samples were diluted 1:3 in PBS buffer (Life Technologies GmbH, Karlsruhe, Germany) in order to adjust the concentration of the samples to approximately 5 mg/ml total protein concentration. Sample preparation was performed individually for each chip. All samples were loaded twice on each chip. Therefore, 24 analyses per sample were evaluated using the following Protein 200 Plus assay settings for peak integration: minimum peak height of 0.1, minimum peak width of 0.1 and slope threshold of 4.0.

An example of the performed analysis is depicted in figure 1. Figure 1A shows the gel-like image of the four different antibody samples (2 standards, 2 test samples) aligned to the ladder sample. The ladder includes proteins of known sizes, was analyzed on each chip and was used for size determination. Figure 1B displays the overlay of the corresponding electropherograms. Both standard samples show almost only two peaks corresponding to the light (LC) and heavy chain (HC) of the antibody respectively. After incubation of the antibodies at elevated temperatures additional peaks appear in the electropherograms (marked with an arrow) due to degradation of the antibody or the formation of aggregates, larger products. Table 1 shows the results of the validation study. On average 98.76 % of standard antibody sample 1 corresponds to the light and heavy chain of the antibody. After incubation for one month at 40C the two chains represent only 93.78% of the whole protein a mount detected. Antibody 2 was incubated for 12 weeks at 40C and the proportion of light and heavy chain decreases from an average of 98.38 % to 85.34 %. An excellent reproducibility below 2 % CV was achieved

Conclusion
In this study we show that the Agilent 2100 bioanalyzer can be used as an ideal tool to study the stability of antibodies in stress tests with excellent reproducibility. This application is commonly used as a quality control step in QA/QC departments in order to trigger typical degradation and aggregation patterns for a specific antibody. Those patterns will be used in the down stream processes of antibody production and formulations.
Traditional denaturing sodium dodecylsulfate-polyacrylamide gel electrophoreses (SDS-PAGE) still serves as one of the standard methods for the QC of antibodies. SDS-PAGE is labor-intensive, timeconsuming and difficult to standardize (1-3). In contrast the Agilent 2100 bioanalyzer provides a fast, standardized method with automated and detailed data analysis.



'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. A Simple and Rapid Technique to Purify High-Quality Midiprep DNA
2. Generate High-Quality, Directional Plasmid cDNA Libraries
3. Quality assurance of RNA derived from laser microdissected tissue samples obtained by the PALM MicroBeam System using the RNA 6000 Pico LabChip kit.
4. High Quality RNA Isolation Services
5. Isolate High Quality Total RNA from LCM Samples Suitable for Microarray and qRT-PCR Analysis
6. Preparation of PCR Quality Template DNA from a Wide Range of Tissues Using InstaGene Matrix
7. Using the Agilent 2100 bioanalyzer for quality control of protein samples prior to MS-analysis
8. Advancing the quality control methodology to assess isolated total RNA and generated fragmented cRNA
9. Evaluation of RNA isolated from human tissues: quality control for surgical and postmortem samples
10. Outstanding transfection rates using the Multiporator
11. Outstanding transfection rates using the Multiporator
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:1/19/2017)... Basel, Switzerland (PRWEB) , ... ... ... leading provider of advanced software solutions for pharmaceutical research and development (R&D), ... project-based expertise in omic data analysis and interpretation for the rapidly evolving ...
(Date:1/19/2017)... (PRWEB) , ... January 18, 2017 , ... ... National Institutes of Health (NIH) to update its Data Sharing Policy. Specifically, the ... element of grant applications subject to the existing policy. AMIA recommended that NIH ...
(Date:1/19/2017)... -- Northwest Biotherapeutics, Inc. (OTCQB: NWBO) ("NW Bio"), a U.S. ... and inoperable solid tumor cancers, announced today that Dr. ... Bio, will present at the Phacilitate Immunotherapy World Conference ... Hotel in Miami, Florida . ... Therapeutic Approaches – Expanding the Reach of Cancer Immunotherapy ...
(Date:1/18/2017)... 2017   Boston Biomedical , an industry leader ... cancer stemness pathways, will feature data from two clinical ... 2017 ASCO Gastrointestinal Cancers Symposium, held from January 19-21, ... Napabucasin is an orally-administered investigational agent designed to ... Cancer stem cells (CSCs) possess the property of stemness ...
Breaking Biology Technology:
(Date:1/6/2017)... Jan. 5, 2017  Delta ID Inc., a leader ... technology for automotive at CES® 2017. Delta ID has ... to demonstrate the use of iris scanning as a ... the driver in a car, and as a way ... experience. Delta ID and Gentex will demonstrate ...
(Date:12/22/2016)... MOUNTAIN VIEW, Calif. , Dec. 20, 2016  As ... all levels, 23andMe, the leading personal genetics company, recently released ... Only Me . The book focuses on the topics ... the Next Generation Science Standards (NGSS) taught in elementary school ... the second in a series by illustrator Ariana Killoran ...
(Date:12/19/2016)... TORONTO , 19 de diciembre de 2016  Mosaic ... permitirá el desarrollo acelerado de MSC-1, un anticuerpo humanizado que se ... tumor en 2017, con múltiples sitios previstos a lo largo de ... ... el factor inhibidor de leucemia (LIF), una citoquina pleiotrópica que se ...
Breaking Biology News(10 mins):