Quality assurance of RNA derived from,,,laser microdissected tissue samples,,,obtained by the PALM MicroBeam System,,,using the RNA 6000 Pico LabChip kit.


Gene expression profiling of material isolated by microdissection has become a very popular method for analyzing cellular behavior in a micro scale and is used in a wide range of research and clinical applications. Laser-assisted microdissection allows isolation of RNA from specific cell populations out of their surrounding tissue environment. Usually, the sample yield that can be obtained by this method is too low for quality control of RNA using a standard technique. On the other hand, extraction of high quality RNA is crucial for all subsequent steps and the success of the overall experiment. Since the introduction of the Agilent RNA 6000 Pico LabChip kit, it is possible for the first time to analyze total RNA samples in concentrations down to 200 pg/l, which is in the range of the sample concentration one can expect from microdissected tissue or cells. In this study we analyzed RNA samples that were obtained by laser microdissection and pressure catapulting (LMPC; PALM MicroBeam System) from tissue sections. The influence of different staining procedures and RNA extraction kits on RNA quality was investigated. RNA was tested for quality and approximate yield, as well as the reproducibility within homologous sample replicates. The data revealed clear differences between the different isolation kits and also between the staining procedures that were used.
The analysis of gene expression has become one of the major scientific tools for understanding the behavior of cells in vivo and in vitro and is used in many biological and medical applications. Using microarray hybridization, the abundance of a large number of transcripts can be analyzed in a single experiment. Comparison of different expression patterns enables scientists to correlate the changes in the transcriptome to various external factors influencing the cellular behavior. Especially in clinical research there is the need for methods that allow expression analysis from very specific cells. Prerequisite for meaningful results are samples of high purity without contamination from unwanted cells that could potentially interfere with detection of important transcripts. Laser-assisted microdissection with the PALMMicroBeam system allows isolation of individual cells in a simple and fast way. Using this technology, researchers are able to procure pure cell populations from specific areas of tissue sections for isolation of DNA, RNA, and proteins with visual control using a microscope. An UV-A-laser-mediated process dissects selected specimen from various sources and transfers them contact free directly into collection vessels for subsequent extraction (see figure 1). Together with novel methods for linear RNA amplification it is possible to perform gene expression profiling analyses even from very limited amount of cell material. The most

critical parameter for the success of such an experiment is the integrity and purity of the RNA.
Experimental case study
Tissue preparation and microdissection
Snap-frozen mouse liver tissue stored at 80 C was cut in 7 m serial sections on a cryotome at 25 C. The sections were transferred to PALM MembraneSlides (1mm glass slides covered with a 1.35 m thin Polyethylene-naphthalate- membrane to facilitate the laser pressure catapulting procedure) and air dried for 10 seconds. After a fixation step of 5 minutes in 70% ethanol at -20 C and a short wash in RNase free water (10 seconds) the sections were further processed according to standard histochemical procedures. Depending on the subsequent tests either Hematoxylin/Eosin (HE), Nuclear Fast Red (NFR), Methylgreen (MG) or Methylene Blue (MB) staining was applied. After staining and a short increasing ethanol series the sections were air-dried and either used immediately or refrozen at 20 C for up to 2 days. The PALM Micro- Beam System was used to precisely excise the selected tissue areas with the UV-A-laser and then to catapult the areas of interest contact free against gravity into collecting caps filled with 12 l of RNase-free water. For lysis the isolated material was then immediately mixed by inversion with the respective lysis buffers for RNA extraction (see below).
RNA Purification
For this study RNA isolation kits from 3 different manufacturers were used. Two are based on a column-purification, while the third one uses an extraction-/precipitation- based approach. RNA cleanup was carried out according to the manufacturers protocols with the only variation that uniformly a volume of 30 l of elution buffer or water was used in the final step to collect the purified RNA. For the column-procedures the elution process was repeated once with the primary eluate of 30 l to improve the yield. The scope of this work was the demonstration of feasibility to perform RNA sample quality control from microdissected material and not the evaluation of commercially available RNA extraction kits. Therefore, the manufacturers are not explicitly named here, but instead referred to as manufacturer A, B, and C.
RT-PCR
Selected RNA samples were reversely transcribed with the First Strand cDNA Synthesis Kit for RT PCR (AMV) cDNA Synthesis Kit for RT-PCR (Roche) according to the manufacturers protocol. Briefly, 5-8 l of each RNA solution were transcribed by AMV-RTase with random primers in a total volume of 20 l for one hour at 42 C. For the subsequent quantitative RT-PCR analyses 1-2 l of each cDNA-solution were used as templates. The PCR amplification of the cDNA was performed in a LightCycler instrument (Roche) in 20 l reactions using protocols and components of the QuantiTect SYBR Green PCR-Kit (Qiagen). cDNA-specific primers for murine PBGD (porphobilinogen deaminase) were used as model system producing a PCR-fragment of 154 bp. Resulting crossing-point values were used to compare extraction efficiencies of different kits or influences of different staining procedures.
RNA 6000 Pico Assay Protocol
All chips were prepared according to the instructions provided with the RNA 6000 Pico LabChip Kit. In brief, 550 l of the RNA 6000 Pico gel matrix were placed on a spin filter, centrifuged at 1500 g and divided in 65 l aliquots. After addition of 1l RNA 6000 Pico dye concentrate to one gel aliquot, the mixture was vortexed and centrifuged at 13000 g for 10 min. A RNA 6000 Pico chip was filled with gel-dye mix using the chip priming station, followed by addition of Conditioning Solution and Marker. 1 l of RNA 6000 ladder (Ambion) and RNA samples were added in the designated wells, the chip was vortexed for 1 minute and run on an Agilent 2100 bioanalyzer.
Results and Discussion
RNA quality and yield obtained by different isolation kits
All 3 kits tested in this study appeared to work well together with the RNA 6000 Pico assay (figure 2). For the comparison of the different kits we used a standard HE tissue staining and dissected areas corresponding to 1000, 2000 and 3000 cells. Quality and yield of the isolated RNA showed a high variability with respect to the different kits and also within the replicates of one specific method. This may be caused either by differences in the tissue samples (different areas were isolated from the same section) or by some variances in sample handling. On average the extraction-/precipitation-based method (manufacturer C) showed a higher RNA recovery than the two column-based methods, which is most probably caused by irreversible RNA-binding at the column material. The average yield (manufacturer C) obtained from a 1000 cell area was approximately 700 pg/l when resolving the pellet in 30 l of water. This would result roughly in a theoretical average RNA recovery rate of approx 21 pg per cell. In contrast to this the yield obtained by the two column-based kits varied between approx. 130 and 380 pg/l (1000 cells; 30 l elution volume). Note that the quantitative results from an analysis with the RNA 6000 Pico kit are somewhat dependent on the salt contentof a sample. Only a rough estimation of the RNA concentration of a sample can be obtained.

RNA quality and yield obtained from tissue stained with different methods
In this test, we compared RNA quality and yield when using different standard histochemical staining methods. Therefore, we dissected areas corresponding to 2000 cells and purified the RNA using the kit of manufacturer C (3 replicates for each sample). Serial tissue sections were stained using either Hematoxylin/Eosin (HE), Nuclear Fast Red (NFR), Methylgreen (MG) or Methylene Blue (MB). In general, all four staining procedures were compatible with the RNA Pico Assay, but significant differences in RNA yield and quality were observed. Best results were obtained using the MG stain (average yield: approx. 3100 pg; highly intact RNA; see fig. 3A). HE (figure 3B) and NFR staining (figure 3C) gave very similar results in terms of RNA yield and quality (average yield: approx. 1600 pg; slightly degraded RNA). In the case of MB staining, ambiguous results were observed (average yield: approx. 730 pg; partially degraded RNA; figure 3D). The 28 S band is almost entirely missing, which could be due to RNA degradation or due to a salt content that exceeds the salt specifications of the Pico kit. RNA-extraction with the kit of manufacturer B from the same sections with all four staining methods yielded slightly different relations (see in RT-PCR results below), and overall lower concentration levels as expected from the comparison of the kits from above. It should be noted that this short investigation does not put a value on the use of certain cell stains for work with microdissected material but rather points out the value of the RNA 6000 Pico kit for optimization of experimental protocols.

Correlation of the results obtained by the RNA Pico Assay to real-time RT-PCR
Comparison of RNA extraction kits:
All 3 sample extraction kits that were tested yielded good quality RNA that could be used for RT-PCR (figure 4). For the comparison we used HE-stained sections to microdissect and catapult areas of 1000 cells each (3 replicates per sample). After RNA isolation with the three different kits real-time RT-PCR with the LightCycler was performed. The amount of cDNA used as template was half of the RNA used for the 2100 bioanalyzer analysis (fig. 4A). The specificity of the PCR-fragments was proven by the melting curve (figure 4B). The crossing points (Cp)* of the growth curves (figure 4C) being a scale for the initial specific RNA-amount were close together (about one cycle distance each) correlating to the Bioanalyzer values although Manufacturer A was overestimated there probably due to the DNA-contamination (figure.1).

Comparison of staining procedures
In this test serial tissue sections were stained using either Hematoxylin/Eosin (HE), Nuclear Fast Red (NFR), Methylgreen (MG) or Methylene Blue (MB). Areas corresponding to 2000 cells were dissected, the RNA purified using Manufacturer B (3 replicates each) and analyzed with the 2100 bioanalyzer (figure 5A) and the Light-Cycler as above. The cDNA input for the PCR corresponded to a fourth of the RNA amount used for the 2100 bioanalyzer run. The melting curve proves the specificity of the products (figure 5B), while the growth curve reflects the starting amount of RNA (figure 5C). In contrast to the test with Manufacturer C (see above) NFR was the optimal stain, though the Cp-values lay very close together (only half a cycle difference) except for MB which showed worse results. (CpNFR: 34,7; CpHE: 35,2; CpMG: 35,8; CpMB: 37,6). This real-time RT-PCR results again correlated well with the 2100 bioanalyzer values.

Conclusion
RNA derived from laser-microdissected tissue isolated by the PALMMicroBeam system is of high quality and can be analyzed efficiently using the RNA 6000 Pico assay with the Agilent 2100 bioanalyzer. RNA-purification kits of different manufacturers and various common staining procedures have been tested and were compatible with the Pico Assay. Nevertheless, it seems to be advisable that any extraction protocol and staining method should be tested in combination to find the optimal procedure for tissue microdissection experiments. The RNA 6000 Pico kit is well suited to show differences in RNA quality and, therefore, is an ideal tool to optimize experimental conditions. Its unprecedented sensitivity allows for the first time quality control in the context of microdissection experiments ensuring successful gene expression profiling experiments.



'"/>

Source:


Page: All 1 2 3 4 5 6 7 8

Related biology technology :

1. A Simple and Rapid Technique to Purify High-Quality Midiprep DNA
2. Generate High-Quality, Directional Plasmid cDNA Libraries
3. Quality control of antibodies using the 2100 bioanalyzer and the Protein 200 Plus assay
4. High Quality RNA Isolation Services
5. Isolate High Quality Total RNA from LCM Samples Suitable for Microarray and qRT-PCR Analysis
6. Preparation of PCR Quality Template DNA from a Wide Range of Tissues Using InstaGene Matrix
7. Protocol for GenomePlex Whole Genome Amplification from Formalin-Fixed Parrafin-Embedded (FFPE) tissue
8. 2D-LC analysis of phosphopeptides in brain tissue using Ettan MDLC and Finnigan LTQ
9. Sensitive identification of phosphopeptides in brain tissue using Ettan MDLC and Finnigan LTQ
10. GenomiPhi amplification of FFPE tissue DNA for array CGH
11. High sensitivity quantitation of metabolites of nitrofuran antibiotics in animal tissue using LC/MS/MS
Post Your Comments:
(Date:10/19/2014)... October 19, 2014 The Latin American ... market in Latin America with analysis and forecast of ... 2013, and is expected to reach $2,366.8 million by ... 2018. , Browse through the TOC of the Latin ... of the in-depth analysis provided. It also provides a ...
(Date:10/19/2014)... (PRWEB) October 19, 2014 The Asian ... in Asian with analysis and forecast of revenue. The ... grow to around $463.9 million by 2018, at a ... Browse through the TOC of the Asian Automatic patient ... analysis provided. This also provides a glimpse of the ...
(Date:10/19/2014)... The report "Unmanned Ground Vehicle Market ... Payload, Type & by Geography - Forecasts & ... ground vehicle market into various sub segments with ... The report also identifies the factors driving this ... with the adoption trends. , The unmanned ground ...
(Date:10/18/2014)... 2014 The report "1,6-Hexanediol Market ... Forecasts to 2019", analyzes the global market on ... understand the industry details such as the key ... geographic regions. , Browse 81 Market Data Tables ... in-depth TOC on “1,6-Hexanediol Market by Application and ...
Breaking Biology Technology:The Latin American hardware encryption display market is expected to reach $2,366.8 million by 2018 - New Report by MicroMarket Monitor 2The Latin American hardware encryption display market is expected to reach $2,366.8 million by 2018 - New Report by MicroMarket Monitor 3The Asian Automatic patient billing market is estimated to grow to around $463.9 million by 2018 - New Report by MicroMarket Monitor 2The Asian Automatic patient billing market is estimated to grow to around $463.9 million by 2018 - New Report by MicroMarket Monitor 3Unmanned Ground Vehicle (UGV) Market Estimated to Reach $8.26 Billion by 2020 - New Research Report by MarketsandMarkets 2Unmanned Ground Vehicle (UGV) Market Estimated to Reach $8.26 Billion by 2020 - New Research Report by MarketsandMarkets 3Unmanned Ground Vehicle (UGV) Market Estimated to Reach $8.26 Billion by 2020 - New Research Report by MarketsandMarkets 4Unmanned Ground Vehicle (UGV) Market Estimated to Reach $8.26 Billion by 2020 - New Research Report by MarketsandMarkets 51,6-Hexanediol Market Projected to Reach $916.1 Million at a CAGR of 8.3% by 2019 – Report by MarketsandMarkets 21,6-Hexanediol Market Projected to Reach $916.1 Million at a CAGR of 8.3% by 2019 – Report by MarketsandMarkets 31,6-Hexanediol Market Projected to Reach $916.1 Million at a CAGR of 8.3% by 2019 – Report by MarketsandMarkets 41,6-Hexanediol Market Projected to Reach $916.1 Million at a CAGR of 8.3% by 2019 – Report by MarketsandMarkets 5
... , LONDON, Nov. 16 The "omics" revolution of the ... will test the frontiers of biomedical research in the coming ... development promises excellent growth potential for the biomarker analysis market. ... assay development, there is a huge lag between the time ...
... WASHINGTON The Project on Emerging Nanotechnologies (PEN) ... and iPod Touch that lets users discover and ... emerging technology of using materials by engineering them ... from consumer electronics to improved drug delivery systems. ...
... SHANGHAI, China, Nov. 13 /PRNewswire-Asia-FirstCall/ ... a leading Chinese firm,specializing in the research, development, manufacture, ... will host a,conference call at 11:00 a.m. EST on ... for the second quarter of its 2010 fiscal year. ...
Cached Biology Technology:Frost & Sullivan Reveals Strategies on Turning Challenges into Opportunities in the European Biomarkers Market 2findNano app puts nanotech in your pocket 2China-Biotics, Inc. Announces Conference Call to Discuss Second Quarter Fiscal 2010 Financial Results 2
(Date:10/17/2014)... German . To ... treating overdoses, doctors are often limited to supportive therapy such ... a combination of drugs involved. So what can be done ... pills? ETH professor Jean-Christophe Leroux from the Institute of Pharmaceutical ... this question. "The task was to develop an agent that ...
(Date:10/16/2014)... challenged conventional thinking on how the bowel lining develops ... how bowel cancer starts. , The researchers produced evidence ... the ,crypts, that are a feature of the bowel ... bowel cancer development, a controversial finding as scientists are ... 3D imaging technologies, Dr Chin Wee Tan and Professor ...
(Date:10/16/2014)... -- NXT-ID, Inc. (NASDAQ: NXTD and ... focused on the growing mobile commerce market announces that ... Smart Wallet, at the 13 th annual Future ... conference on the intersection of technology and the economy. ... founder and publisher of the Strategic News Service™, the ...
Breaking Biology News(10 mins):Emergency aid for overdoses 2Emergency aid for overdoses 3Cryptic clues drive new theory of bowel cancer development 2NXT-ID Inc. to Present at 13th Annual FiRe Conference 2NXT-ID Inc. to Present at 13th Annual FiRe Conference 3
... of Interventional Radiology Foundation,s Discovery Campaign, which seeks to ... areas of discovery, announced a major corporate pledge to ... for vascular access and the diagnosis and treatment of ... pledged $500,000 and was named a "Futurist" supporter. ...
... Researchers from the Wildlife Conservation Society have stumbled upon ... a natural stone arch spanning more than 200 feet ... of Afghanistan, the recently discovered Hazarchishma Natural Bridge is ... level, making it one of the highest large natural ...
... is not a sexual behaviour commonly discussed, but a ... 200 young women has found one in three (32.9 ... QUT Associate Professor Robert Schweitzer,s research, published ... , looked at the prevalence of postcoital dysphoria or ...
Cached Biology News:Fueling interventional radiology growth: SIR Foundation announces Navilyst Medical Pledge 2Newly discovered natural arch in Afghanistan one of world's largest 21 in 3 women suffer post-sex blues 2
... involves the optimization of immunostaining. ... of the primary antibody, selection ... (e.g. Alk. Phos., HRP, fluorescence, ... detection, evaluation by light or ...
Request Info...
... Quality ISO 9001:2000 Certified ... All Precedures Validated, SOP's Training ... Process for All Procedures GLP Documentation Upon ... one stop shopping point for all your ...
Request Info...
Biology Products: