Navigation Links
PathCheck Applied to Measurement of Protein Solutions in the SpectraMax Plus Microplate Spectrophotometer (MaxLine Application Note #26)

Protein concentration is commonly estimated by measuring the absorbance at 280 nm and calculating the concentration using the extinction coefficient of the protein. The procedure is feasible because extinction coefficients are typically obtained using standard cuvettes with a fixed 1 cm pathlength. Extinction-based protein assays can also be done in microplates, but they have not been widely-used because of the inconvenience of compensating for variable pathlengths of samples in microplate wells. The introduction of the Molecular Devices SPEC-TRAmax PLUS microplate spectrophotometer marks the first time that photometric measurements made in a microplate format can be automatically normalized such that the resultant values are identical to those obtained with the corresponding solutions in a standard 1 cm pathlength cuvette. The following application note outlines the use of the pathlength correction feature Pathcheck, compares absorbance results obtained in cuvettes and microplate, and gives an example where Pathcheck is applied to column chromatographic fractions and the elution profile displayed.


For details of the principles of pathlength correction, refer to MAXline application note #25, Optical density measurements automatically normalized to a 1 cm pathlength in the SPECTRAmax PLUS microplate spectrophotometer.

1. SPECTRAmax PLUS microplate spectrophotometer

2. UV-transparent microplates; e.g.:
SPECTRAplate Quartz, Molecular Devices Corp. (catalog # R8024)
SPECTRAplate Quartz half-area microplate, Molecular Devices Corp., (catalog # R8028)
Quartz microtest 8-well strip, Hellma (catalog # 730.010QS); use with a frame for 8-well strips, E & K Scientific Products (catalog # 564101). Note: the 8-well strip frame sold by Evergreen Scientific does no t fit properly.
Disposable UV Plate, Corning Costar Corp. (catalog # 3635)
UV-MAX, (Polyfiltronics, Inc., catalog # VO32SCXP8)

3. Pipettor and tips or transfer pipets suitable for use with microplates.

4. Samples (100-300 L each)

5. Quartz cuvette

Set up to make pre-read measurements
Step 1 Launch SOFTmax PRO, then open a Plate section or, if necessary, create a new Plate section.

Step 2 Set up the Instrument Settings dialog box as shown in Figure 1. In the Instrument Settings dialog box, select to perform an endpoint read at 280 nm. Select PathCheck, (Pre-read plate will automatically be checked) and select Cuvette Reference. (Alternatively, you can use the factory-installed Water Constant as a shortcut, though it is potentially less accurate.)

Step 3 Use the Template Editor to create a template showing where standards, blanks, and unknowns will be located on the microplate.

Step 4 Set up the Reduction dialog box as shown in Figure 2. Set the wavelength combination to L1 and the data mode to Absorbance. Make sure that Use pre-read plate and Apply Pathcheck are selected.

Making pre-read measurements
The purpose of the pre-read is to subtract the background absorbance of the plate. In contrast, blanks (plate, sample or group) are designated in the template and are subtracted after pathlength correction. In some cases, you may choose to use stored pre-read values (See below).
Step 1 Select a clean microplate and pipet distilled water into the wells in approximately the same volumes as the intended samples. Accurate pipetting is not necessary.

Step 2 Put the plate into the drawer of the SPECTRAmax PLUS, then click the Read button in SOFTmax PROs toolbar to make the pre-read measurements.

Step 3 Remove the plate, pour out the water and blot the plate dry.

Making pathlength-corrected absorbance measurements
Note: If you have elected to use the pre-programmed Water Constant, you may skip Step 1.
Step 1 Place a clean quartz cuvette containing your sample buffer or distilled water into the SPECTRAmax PLUS cuvette port.

Step 2 Transfer aliquots of your samples and blanks into their designated wells in the microplate. Accurate pipetting is not necessary, though the volumes should between 100 L and 300 L for best pathlength-corrected results.

Step 3 Put the plate into the SPECTRAmax PLUS drawer, then click the Read button in SOFTmax PROs toolbar to make the absorbance measurements. (The cuvette will be read in the near-infrared at the same time the plate is read, and its absorbance values used in the PathCheck calculations.)

To avoid evaporation error, make the readings within a few minutes of putting the samples in the microplate. If the readings must be delayed, cover the plate with an adhesive seal. Remove the seal immediately prior to reading the plate.

Storing and re-using pre-read data
Quartz plates are so uniform that it is not necessary to subtract the plate background on a well-by-well basis. Instead, you can pre-read one plate, then store the pre-read data in a SOFTmax PRO file for subsequent re-use, being careful to save the data under a different file name each time, to avoid over-writing the pre-read data.

If you use Costar UV plates, you may or may not want to use stored pre-read values, depending on the precision you require. At 280 nm, the average absorbance of Costar wells containing water is approximately .050 + .002 OD. Thus, samples having raw absorbance values of 1 OD would be subject to less than 1% potential error due plate background variation. However, the lower the absorbance in the wells, the greater the maximum potential error due to plate variability.

Example 1: Demonstration of the utility of PathCheck using a protein standard curve
To illustrate the performance of the instrument, a protein standard curve was prepared. Bovine serum albumin (BSA) was dissolved in phosphate-buffered saline (PBS), pH 7.4, filter-sterilized and diluted to 6, 5, 4, 3, 2, 1, 0.5 and 0.25 mg/ mL. Absorbance measurements made in a microplate were compared with measurements made on the same solutions in a 1 cm cuvette.

SOFTmax PRO was launched and a Group entitled Plate Standard was created and given a Standards format in the Template Editor of a Plate section. The standards were assigned to wells as illustrated in Figure 3. Standard 01 was PBS with no added protein (0 mg/mL). An identical template was set up in the Template Editor of a Cuvette Set, except the group was named Cuvette Standard.

The instrument settings and data reduction for the Plate section were set up as shown in Figures 1 and 2 above. The CuvetteSet section was set up for an endpoint read at 280 nm.

A cuvette containing PBS was placed in the cuvette port to serve as the cuvette reference for PathCheck. Pre-read measurements were made with 200 L water in each well of a Costar UV plate, afte r which the plate was emptied and blotted dry. Aliquots of the protein standard solutions were transferred into their designated wells using disposable transfer pipets. The volumes in each well were deliberately varied in order to demonstrate the utility of the PathCheck feature. The plate was placed into the drawer of the microplate reader and read.

In the Cuvette Set, a reference read was taken on the PBS-containing cuvette in the cuvette port. Then, one by one, each of the protein standard solutions was read in the cuvette.

The Group Table associated with the Plate section is shown in Figure 4.

All columns except the far right column in the Group Table were created automatically by SOFTmax PRO when the Plate Standard group was created in the template. The numbers in the Values column are absorbance values corrected to a 1.0 cm pathlength. The numbers in the far right column titled Raw Values are raw absorbance values (pre-read subtracted) before pathlength correction. These values were obtained by back-calculation using the custom formula shown in the Calculation dialog box for that column ( Figure 5). Basically, the Raw Values were calculated by multiplying the corrected values by their corresponding pathlengths and taking the average of the replicates.

The Group Table (Cuvette Standard) displays the data from the absorbance readings in the cuvette ( Figure 6). The column Values contains the absorbance values at 280 nm.

Next, the standard curves from the plate and cuvette were plotted in a single graph. A Graph section was created, then the Plots dialog box was used to create three plots by selecting the corresponding Group tables from the Groups pop-up menu, then specifying the x and y variables for each plot ( Figure 7). In the example shown in the fi gure, the highlighted plot is Plate with PathCheck. The data came from the group Plate Standard, the x variable is concentration, and the y variable is MeanValue (i.e. the mean pathlength-corrected absorbance value).

The resulting graph ( Figure 8 ) shows the three plots (plate values with and without PathCheck and the cuvette values). The plot of raw plate absorbance values without PathCheck (bottom line) is irregular because the volumes in the wells (and therefore the pathlengths in the wells) were not uniform. The plot of plate values with PathCheck is linear through the highest concentration (6 mg/ mL). The corrected plate values are superimposable on the cuvette values up through approximately 3 mg/ml, above which the cuvette plot begins to deviate from linearity. The corrected plate plot extends above the cuvette plot because the values are calculated values and their raw absorbance values (bottom plot) are still well within the linear range of the curve.

Example 2: Using PathCheck to evaluate column fractions during purification of antibody-enzyme conjugate on a gel filtration column
Fractions were collected during purification of an antibody-enzyme conjugate on a Pharmacia S-100 gel filtration column. The fractions in the void volume were discarded and fractions 21 through 61 were each read at 280 nm in a Costar UV plate and also in a standard quartz cuvette for comparison. Fractions 24 through 29 were too concentrated and therefore were diluted 1 to 5 to bring them into the linear range of absorbency.

Initial setup of the Group Table
There are two ways in SOFTmax PRO to set up a Group Table for samples (e.g. column fractions) that have dilution factors, but are not going to be quantitated against a standard curve:

1. Assign the Unknowns(dilution) format to the group. SO FTmax PRO will automatically activate the Dilution Factor field and include columns for Dilution (Factor) and Adj.Result in the Group Table. The software also will assume a standard curve is being created, and will create 5 columns that are unnecessary for the present example (calculations of concentrations based on a standard curve and statistical information). Because these columns will not be needed in any calculations, you can delete them from the Group Table.

2. Assign the Basic format to the group, then modify the group to include the dilution factor information, either via the Group Settings dialog box, or via the Series dialog box. Two columns must then be added to the Group Table: one that contains the dilution factors, and a second that multiplies the absorbance values by the dilution factors.

In this example, the second method was used to create the Group Table and assign sample names to the wells. A new group named Plate was created in the Plate section Template Editor and assigned the Basic format. Rows A though D were highlighted and the Series button clicked to access the Series dialog box ( Figure 9 ).

The starting sample was set at 21 (to correspond to its fraction number) and the series was set to proceed from left to right with one replicate. The Sample Descriptor box was checked to activate it, then dilution factor typed into the field. The Units field was cleared and left blank. After clicking the OK button to create the series, the template displayed the sample names (fraction numbers) assigned to the wells ( Figure 10). Samples with dilution factors other than 1 had the dilution factor added by clicking the well in the template to select it, then typing in the appropriate factor into the Dilution Factor field. In the figure, well A5 (fraction #25) is highlighted and its sample name (25) and its dilution factor (5) are displayed in their respective fields. Th e buffer blanks were assigned to wells D6 and D7.

Note: the series feature initially filled all of Row D because SOFTmax PRO by default fills complete rows or columns. Wells D6 through D12 were cleared so that the series would stop at sample #61.

In an analogous manner, the same sample names were assigned to a CuvetteSet section.

Experimental details and customization of the group table
The instrument settings were identical to those used in Example 1 above. Plate pre-read measurements were made in a Costar UV plate. Aliquots of each column fraction and the buffer blank were transferred into their designated wells using disposable transfer pipets, then read in the SPECTRAmax PLUS. As in the example above, the volumes in each well were deliberately varied in order to demonstrate the utility of the PathCheck feature. The column fractions were also read in a quartz cuvette as described above.

The Group Table for the Plate section (after customizing) is shown in Figure 11. Of the seven columns, only three (Sample, Wells and Correct.Values) were created automatically by SOFTmax PRO. (The Correct.Values column was originally entitled Values, but was renamed in this example to distinguish pathlength-corrected values from raw absorbance values. The other four columns were given custom formulas. The Fraction# column was created so that the samples could be given numeric values corresponding to their fraction numbers, then graphed. (In the Sample column, the sample numbers are treated as names and therefore cannot be graphed.)

The custom formula for the Fraction # column ( Figure 12) is Index+20 so that the first entry is 21.

The Raw Values column contains the absorbance values before pathlength correction and were obtained by back-calculation from the cor rected values using the custom formula: Correct.Values*!WellPathLength ( Figure 13).

The custom formula for the Dilution column is shown in Figure 14. The term !factor refers to the sample descriptor defined in the template.

The column Final Values contains the corrected absorbance values multiplied by the dilution factor. The custom formula is Correct.Values* Dilution (Figure 15). In the Cuvette Group Table, an analogous column was created to multiply the absorbance values by the dilution factors.

Figure 16 shows the elution profiles from the gel column obtained in the plate and the cuvette. The lower plot (plate values not pathlength corrected), not surprisingly, does not strictly mirror the upper profiles because the volumes in the wells were not uniform. However, when PathCheck was applied to the plate values, the plate profile was essentially superimposable on the cuvette profile.

The above examples illustrate the utility of the PathCheck feature of the SPECTRAmax PLUS. Protein solutions can be read in a microplate and the absorbance values are identical to those obtained by reading the same solutions in a standard 1 cm cuvette. Accurate pipetting is not necessary. An additional benefit of PathCheck is that the linear range is extended above the cuvette linear range (as long as the raw absorbance values in the wells are within the linear range of the instrument).

back to top



Page: All 1 2 3 4 5 6 7 8 9

Related biology technology :

1. Advantages of Roche Applied Science amplification products
2. Matched siRNAs and Assays: Ambion + Applied Biosystems = RNAi Success
3. Amplified Fragment Length Polymorphism (AFLP) Analysis on Applied Biosystems Capillary Electrophoresis Systems
4. Quantitative Measurement of Cell Proliferation Using the BrdU ELISA: A Comparison Between Colorimetric and Chemiluminescent Detection
5. Cell Proliferation and Viability Measurement
6. LabVIEW for Measurement and Data Analysis
7. Use of FlashPlate Technology for In Vitro Measurement of [35S]-GTPγS Binding in CHO Cells Expressing the Human 5-HT1B Receptor
8. Use of FlashPlate Technology for In Vitro Measurement of 125I-Labeled TGF-1 Binding on Chimeric Extracellular Domain of Type II Transforming Growth Factor Receptor
9. In Vitro Measurement of the Second Messenger cAMP: RIA vs. FlashPlate
10. Measurement of proteasome inhibition in live cells in Molecular Devices microplate fluorometers
11. Measurement of Green Fluorescent Protein in the SpectraMax Gemini XS Spectrofluorometer (MaxLine Application Note #44)
Post Your Comments:

(Date:11/30/2015)... - BioAmber Inc. (NYSE: BIOA ), a leader in renewable ... Act on Climate Pledge, alongside more than 140 companies from ... Administration to demonstrate an ongoing commitment to climate action and ... Paris climate negotiations. ... . --> BioAmber uses biotechnology to convert ...
(Date:11/30/2015)... , Nov. 30, 2015  Culprits beware, a ... chemistry professor Jan Halámek, is taking crime scene ...   -->   ... --> --> ... have discovered a straightforward concept for identifying whether ...
(Date:11/30/2015)... Nov. 30, 2015 /PRNewswire/ - Zenith Epigenetics Corp. ("Zenith" or ... Norman C.W. Wong to its Board of Directors to ... Zenith with a wealth of experience as co-founder of Resverlogix, ... --> --> Dr. Wong ... board of directors. Zenith,s long standing expertise in epigenetics and ...
(Date:11/30/2015)... (PRWEB) , ... November 30, 2015 , ... ... new globally touring exhibition Jurassic World: The Exhibition, opening in March 2016 at ... embark on a worldwide tour including several North American tour dates. The Exhibition ...
Breaking Biology Technology:
(Date:10/29/2015)... YORK , Oct. 29, 2015 ... technology, announced a partnership with 2XU, a global ... to deliver a smart hat with advanced bio-sensing ... and other athletes to monitor key biometrics to ... the strategic partnership, the two companies will bring together ...
(Date:10/27/2015)... October 27, 2015 Munich, ... Gaze Mapping technology (ASGM) automatically maps data from mobile ... Glasses , so that they can be quantitatively ... Munich, Germany , October 28-29, 2015. ... data from mobile eye tracking videos created with ...
(Date:10/26/2015)... 26, 2015 ... adds Biometrics Market Shares, Strategies ... well as Emerging Biometrics Technologies: Global ... its collection of IT and Telecommunications ... --> . ...
Breaking Biology News(10 mins):