Multichannel Pipetting: How to Choose the,,,Correct Pipettor

by Sari Mannonen, Pasi Nieminen, Jukka Kaasinen and Kati Andersin

Multichannel pipettors are widely used in microplate work, including serological testing, molecular biology, immunology assays, and polymerase chain reaction (PCR). The explosion of specific markets such as genome projects and drug discovery has further increased the use of the microplate. Because microplates are currently being used in very labor-intensive applications worldwide, there is a need to increase productivity, maintain accuracy and precision, and reduce physical stress. Each of the approximately ten multichannel pipettor manufacturers emphasizes ergonomics, comfort, ease of use, and safety. Despite this, many users have problems with performance, tip mounting, tip sealing, and tip ejection. The key question is: How does one choose the correct multichannel pipettor? Following are a few criteria that should be considered when choosing a multichannel pipettor.

Ergonomics
All manufacturers claim that their pipettors are ergonomic. However, there exist clear differences in pipetting and tip ejection force as well as in weight, length, and design. For example, the difference between the lightest instrument, the mLINE (Biohit Plc., Finland) (Figure 1), and the heaviest is more than 100 g (137 vs 241 g) (Table 1). Usually, the shorter and lighter the pipettor, the more ergonomic it is to use. However, more important than weight is the balance and fitting of the pipettor to the hand of the user. This is because good stability considerably reduces the strain on the muscles of the hand.1 Most, but not all, multichannel pipettors have a finger support to secure better fitting. Table 1 lists available mechanical multichannel pipettors and their features.

Ease of use
An important feature in multichannel pipettors is the ability to turn the liquid end to an optimal position, for example, when pipetting to microplates (Figure 1). However, many pipettors allow rotation only in one direction, since rotation in the other direction will disassemble the liquid end from the handle. In some models, the connecting nut has to be tightened after adjusting. To avoid accidental disassembling, which leads to a loss of accuracy or even leakage during pipetting, one should ensure that the pipettor chosen allows 360 free rotation without a need for tightening. The mLINE instrument features this ability. Pipettors that are color-coded and have large, clear numbers are easier to read and use. This is also the case with the mLINE unit; in addition, the volume adjustment is stepless, extremely light, and fast to adjust. Most importantly, the volume is locked after adjustment to avoid accidental alteration of the volume during pipetting.2 To avoid this, many pipettors on the market have adjustment systems that are extremely heavy to turn, particularly while wearing gloves.

Pipetting force, accuracy, and precision
There exists much variation among different manufacturers in the specifications for instrument accuracy and precision (Table 1). This regards not only the guaranteed specifications but also the pipetting force, which has an influence on accuracy. The force needed to depress the plunger to the first position varies from 8.7 to 20 N, depending on the manufacturer and the pipettor in use; for the blowout, that number goes up to 45 N (4.5 kg).3 In conventional pipettors, the smaller the volume, the more thumb force that is demanded to operate the plunger. High force, together with small movements, cannot be repeated with the same accuracy after hundreds of pipettings. If one is pipetting more than 1000 samples a day with a conventional mechanical pipettor, it is no surprise that serious fatigue in the muscles of the hand occurs, which can ultimately lead to repetitive strain injuries (RSI).

The mLINE pipettor is lightweight, and the patented spring mechanism,4 combined with low-friction sealings and grease, ensures that less pipetting force is required (Figure 2). The spring mechanism allows the user to pipet with a constant and very low force, regardless of the volume, which has a direct effect on pipetting accuracy. The smaller the force needed for minute piston movements (small volumes such as 0.5 L), the easier it is to pipet accurately.

Due to the instrument's design, the maximum pipetting force of the mLINE is lower than the minimum pipetting force of pipettors produced by most other manufacturers. mLINE demands four times less work from the user than conventional pipettors. The low pipetting and tip ejection force ensures that precision values easily stay within specification, even after prolonged hours of pipetting. Moreover, because the operating mechanism is thermally insulated from the handle, the warmth of the hand does not affect pipetting results at all, which may be the case with other pipettors, especially after several hours of continuous pipetting.

Figure 1 The mLINE pipettor features very light pipetting and tip ejection force.

Figure 2 The mLINE pipettor features individual piston/ tip cone assemblies, spring-loaded tip cones, low-friction sealings and grease, and full autoclavability.

Tip sealing and tip ejection force
Tip mounting and ejection demands much force in multichannel pipetting, and is directly related to tip sealing and the force with which tips are being mounted to the pipettor. Since the pipettor and the tip are designed as a system to ensure a perfect fit, only tips specified by the pipettor manufacturer should be used. A compromise in tip selection may result in uneven sample pickup, especially with 12-channel pipettors.

However, even when the tips of the original manufacturer are being used, traditional multichannel pipettors require a high insertion force, up to 10 kg, because of the large sealing area. This, in turn, leads to massive force needed to eject the tips, especially when using 12-channel pipettors. The tip ejection force with multichannel pipettors can vary up to threefold depending on the number of channels, tip mounting force, manufacturer, and model. Some manufacturers even recommend rocking the pipettor over the tip rack and tightening the tips by hand to ensure secure tip fitting. This is both time-consuming and causes a possible risk of contamination. Without doubt, it results in laborious tip ejection requiring much force. O-rings or elastomer tip cones eliminate that problem, but require maintenance rather often, and wear particles can contaminate samples. The mLINE Optiload system features spring-loaded tip cones (Figures 2 and 3) that allow pipettor tips to be picked up easily with a constant low force. In practice, tips cannot be sealed too tightly to the tip cones, which prevents overinsertion and jamming of the tips. This system ensures that all tips are always sealed on the same height, which is particularly important in microplate pipetting, in which it is necessary that all 12 tips reach the well bottom evenly. Thus, the springs secure optimal, even tip sealing and consistent sample loading, and reduce tip ejection force considerably, up to 65% (Figure 2). Moreover, the tip ejector design has a large, rounded surface, which makes it convenient for users and allows onehanded, low-force tip ejection. Thus, when hundreds of tip ejections are performed daily, optimal tip loading and ejection significantly reduces the risk for RSI.

Volume range
Much variety exists in the volume ranges of different multichannel pipettors. For example, if 50 L is to be dispensed, the range can be as follows: 50-300 L, 30-300 L, 40-350 L, 20-200 L, or 20-300 L, depending on the manufacturer. However, a large volume range should not come at the expense of compromised accuracy and precision, which is often the case (Table 1). Even at low volumes, the piston travel of the mLINE instrument is longer than usual, which makes pipetting more accurate in small volumes. In addition, the minimized dead volume, i.e., the air space between the tip and the piston, further improves accuracy.

Protective tip cone filters and filter ejection
Pipettors are contaminated very easily if no protection is used (filter tips or filter in the pipettor tip cone). Even when working with buffers or aqueous solutions, small amounts of liquid may get inside the tip cone, causing contamination. Special attention should be paid to this problem when working with infectious, radioactive, or DNA samples. An economical and effective way to protect both the pipettor and the sample from contamination is to use a filter in the pipettor tip cone.5 Standard tips can be used with the tip cone filter in many applications, instead of the more expensive filter tips. This feature is offered on Biohit multichannel pipettors (Table 1). Moreover, the mLINE features a thumb-activated filter ejection system with which the filters can be removed safely without the user touching them (Figure 4).6 Very often, the filter is enough to protect the pipettor. However, only autoclaving ensures that the pipettor is virtually free of dangerous pathogens or the agent of concern.

Decontamination and autoclaving
Easy disassembling and autoclaving are regarded as important features for pipettor maintenance. To be able to autoclave the entire pipettor without disassembling the liquid end or having to change the O-rings that many multichannels feature is always an advantage, even if total disassembling is sometimes needed for cleaning. However, there are also convenient and effective decontamination solutions available on the market, which require simple spraying and wiping instead of time-consuming autoclaving.5

When choosing a multichannel pipettor, the user should make sure that it is fully autoclavable; in many pipettors, only the lower part can be autoclaved or the unit is not autoclavable at all. If autoclaving is done frequently, the pipettor, like the mLINE, should be fully autoclavable (numerous cycles) and should not need prior disassembling.

Calibration and in-house service
When purchasing a multichannel pipettor, the user should make sure that it is CE marked, and can easily be calibrated and serviced in-house when necessary. In many multichannel pipettors, service is time-consuming, requiring extra tools and many different parts to be disassembled, cleaned, and reassembled in the correct order. This makes self-service practically impossible, especially if one broken tip cone requires a time-consuming change of the whole liquid end. For the user to be able to service the pipettor quickly and effectively, the multichannel pipettor should have individual piston/tip cone assemblies, and only 1-2 parts to be cleaned, as is the case with the mLINE. The calibration of the instrument has been made to be as user-friendly as possible. The calibration tool, which also acts as a tube cap opener, enables fast and accurate calibration. Most importantly, it allows fine adjustment. Due to a special transmission mechanism, even very small increments in calibration can be done easily and safely.

Figure 3 The Optiload system features springloaded tip cones that allow pipettor tips to be picked up with a constant force. This, in turn, ensures even tip sealing and sample pickup, and considerably reduces tip ejection force compared to conventional pipettors.

Figure 4 With the instrument's filter ejection mechanism, filters can be removed simply by pressing the piston all the way down to the third position.

Conclusion
There are several types of multichannel air displacement pipettors on the market. The best way to find the right tool for each application is to go through the pipetting procedures and find out which pipetting techniques and volumes are needed. However, less force in pipetting and tip ejection means less fatigue and discomfort, which results in improved productivity and user satisfaction. Therefore, one should look for a light, low-force, ergonomic multichannel pipettor that does not compromise accuracy and precision. It is preferable that tips designed for the particular pipettor are used. Before choosing a multichannel pipettor, the user should ask for a trial run to test the instrument. The best decision comes through evaluation of one's own experience.

References

  1. David G, Buckle PA. A questionnaire survey of the ergonomic problems associated with pipettes and their usage with specific reference to work-related upper limb disorders. Appl Ergonom 1997; 28 (4):257-62.
  2. FI patents 86968, EP 0527170, JP 3168296, and U.S. 5,347,878.
  3. Mannonen S, Nieminen P, Kaasinen J, Andersin K. Raising the standard of mechanical pipetting. Int Labmate 2003; 28(5):34-5.
  4. FI patent 109407 and patents pending.
  5. Kolari M, Mannonen S, Takala T, Saris P, Suovaniemi O, Salkinoja-Salonen MS. The effect of filters on aseptic pipetting lifetime of mechanical and electronic pipettors and carryover during pipetting. Lett Appl Microbiol 1999; 29:123-9.
  6. FI patents 108281 and 104885 and patents pending.

Dr. Mannonen is Vice President, Sales and Marketing, Mr. Nieminen and Mr. Kaasinen are Mechanical Designers, and Ms. Andersin is Product Manager, Biohit Plc., Laippatie 1, 00880 Helsinki, Finland; tel.: +358 9 773 861; fax: +358 9 773 86 200; e-mail: sari.mannonen@biohit.com.
'"/>

Source:


Page: All 1 2 3 4 5 6 7 8

Related biology technology :

1. Verifying Multichannel Pipettor Performance with Standard Dispense Solutions in the SpectraMax Plus (MaxLine Application Note #28)
2. Choose the Right Transfection Agent for Your RNAi Experiment
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Multichannel Pipetting How Choose the Correct Pipettor

(Date:10/22/2014)... 2014 Shimadzu Scientific Instruments introduces its ... for the pain management and clinical markets. The software ... review by allowing users to highlight and review results ... QuantAnalytics allows analysts to filter results by group or ... by eliminating the need to sift through analytes of ...
(Date:10/22/2014)... Cranford, NJ (PRWEB) October 22, 2014 ... technologies and services – has been working with a ... Health and New Haven Farms Community Farming Collaborative to ... iPad-based ePRO solution. The study, which started in ... Wellness program that provides participants weekly exposure to and ...
(Date:10/22/2014)... (PRWEB) October 22, 2014 Grace Century, ... and healthcare projects, announces the addition of Dr. ... to its advisory team. Dr. Siddiqui will provide further ... , A graduate of University College Medical School ... degreed in medicine in 2001. With further certification as ...
(Date:10/22/2014)... (PRWEB) October 22, 2014 According ... Market by Service Type (Consulting, Integration, Deployment & ... by Deployment Model (Public, Private, Others (Hybrid & ... MarketsandMarkets, defines and segments the Cloud Professional Services ... and forecasting of revenues. It also identifies the ...
Breaking Biology Technology:Shimadzu Releases QuantAnalytics Open Access LCMS Software Package for Clinical Applications 2myPROpad™ Topic of Panel Discussion at the Yale Food Systems Symposium 2myPROpad™ Topic of Panel Discussion at the Yale Food Systems Symposium 3Dr. Yousef Siddiqui joins the Grace Century Advisory Team 2Cloud Professional Services Market by Public, Private, Hybrid & Community Expected to Reach $34.41 Billion by 2019 - New Report by MarketsandMarkets 2Cloud Professional Services Market by Public, Private, Hybrid & Community Expected to Reach $34.41 Billion by 2019 - New Report by MarketsandMarkets 3Cloud Professional Services Market by Public, Private, Hybrid & Community Expected to Reach $34.41 Billion by 2019 - New Report by MarketsandMarkets 4Cloud Professional Services Market by Public, Private, Hybrid & Community Expected to Reach $34.41 Billion by 2019 - New Report by MarketsandMarkets 5
... SAN DIEGO, Feb. 3 Gen-Probe Incorporated (Nasdaq: ... has joined the Company as senior vice president, research ... expertise and leadership to our already strong senior management ... officer. "We believe Eric will play an integral ...
... Primary and Key Secondary Endpoints -- CAPACITY 1 Misses ... Submit NDA and MAA -BRISBANE, Calif., Feb. 3 ... results from the two Phase 3 CAPACITY studies evaluating ... The primary endpoint of change in percent predicted Forced ...
... is like none other. To mark the observance, African ... promote science education, achievement motivation and positive psychological development among ... ... Washington, DC (PRWEB) February 3, 2009 -- African Ancestry, Inc., ...
Cached Biology Technology:Eric Lai Joins Gen-Probe as Senior Vice President, Research and Development 2InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 2InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 3InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 4InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 5InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 6InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 7InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 8InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 9InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 10InterMune Reports Results of Two Phase 3 CAPACITY Studies of Pirfenidone in IPF 11African Ancestry, Educators, Identity Experts Launch DNA Education Project during Historic Black History Month; 7th Graders Score High on Ultimate Self Test 2African Ancestry, Educators, Identity Experts Launch DNA Education Project during Historic Black History Month; 7th Graders Score High on Ultimate Self Test 3
(Date:10/15/2014)... has admitted Johannes Gutenberg University Mainz (JGU) as ... hub continues to maintain a significant nationwide standing ... "The acceptance of Johannes Gutenberg University Mainz as ... milestone in the national and international competition for ... of Rhineland-Palatinate can be proud that Mainz University ...
(Date:10/15/2014)... the 44th Rosenstiel Award for Distinguished Work in Biomedical ... the mechanisms of genomic instability and its implications for ... second alumnus to win the Rosenstiel Award; the first, ... Alt is the Charles A. Janeway Professor of Pediatrics ... an investigator at the Howard Hughes Medical Institute at ...
(Date:10/14/2014)... 2014)—It,s been millions of years since T. rex ... by Ohio University scientists is breathing life back into ... dinosaur snouts. The research has important implications for how ... to enhance the sense of smell and cool their ... Ohio University doctoral student Jason Bourke, lead author of ...
Breaking Biology News(10 mins):Johannes Gutenberg University Mainz joins Germany's Gauss-Allianz as a full member 2Johannes Gutenberg University Mainz joins Germany's Gauss-Allianz as a full member 3Brandeis awards 44th Rosenstiel Award to pioneering geneticist Fred Alt 2Dinosaur breathing study shows that noses enhanced smelling and cooled brain 2Dinosaur breathing study shows that noses enhanced smelling and cooled brain 3
... Bartness, Regents, Professor of Biology at Georgia State University, ... Institutes of Health to further research into the biological ... Research in Time (MERIT) award from the National Institute ... his lab,s investigation into the communication loop between fat ...
... new study suggests that it might be a genetic ... three thousand men - all pairs of male twins ... first time they had sex. Many participants in the ... and premature ejaculation at their first sexual encounter. These ...
... called a "Dream Team." Lombardi Comprehensive Cancer ... DSc, and 12 of the nation,s top breast cancer ... Stand Up To Cancer (SU2C) to form a scientific ... and brightest scientists will conduct new breast cancer research ...
Cached Biology News:Researcher garners major award from NIH for further exploration into the mechanisms of obesity 2Lombardi scientist brings 'dream team' breast cancer research effort to GUMC 2Lombardi scientist brings 'dream team' breast cancer research effort to GUMC 3
... BD Pathway 435 is an affordable, ... plates and slides, providing a broad ... system rapidly acquires high-resolution images with ... flexible computer-controlled excitation and emission filter ...
... include: Microarray hybridization chamber assembly (1 ... clamp, thumbscrew) Microarray chamber user guide ... intended to be cleaned and reused. The ... Catalog Oligo Microarray kits and all custom ...
... Blood Vessel Staining Kit is a very ... vascular endothelial cells, containing both anti-von Willebrand ... applicable to a variety of species. This ... for the staining of frozen and paraffin-embedded ...
... Coon's Modification of Ham's F-12 was ... produced by viral fusion. The modification consists ... and including ascorbic acid. The salt concentrations ... contains 0.863 mg zinc sulfate per liter, ...
Biology Products: