| HOME >> BIOLOGY >> TECHNOLOGY |
Elizabeth P. Roquemore, Sam Murphy, *Stephen J. Capper, Suzanne M. Hancock, Elaine Adie, Molly Price-Jones, Stephen Game & Stuart Swinburne
Amersham Biosciences UK Limited, Amersham Place, Little Chalfont, Buckinghamshire HP7 9NA England
Abstract
As high throughput imaging systems, analysis algorithms and associated
cellular assays have begun to be used in profiling and secondary analysis,
scientists have used tools derived from primary screening such as Z factor to
assess assay quality. Image-based cellular assays may be better suited to
lead profiling, where the needs are different and modified criteria should be
applied in assessing assay quality. We ask the question what statistical
metrics most accurately reflect the nature and use of cell-based assays in
lead profiling? We present data from a variety of such assays we have
developed, such as GFP translocation assays, including simple Z factor
analysis and propose multiple statistical approaches to assess assay quality
that may be better suited to high information content assays.
Introduction
We have developed a number of live-cell translocation assays
that are compatible with high-throughput micro-imaging platforms
such as the IN Cell Analyzer 3000 and the IN Cell Analyzer
1000. To assess assay quality during assay development and for
subsequent QA, we initially used the well known measure of
assay performance, Z factor. While Z factor is of value in
assessing the size of the reading window for a screening assay,
we observed that in some cases Z factor analysis obscured
meaningful differences in assay performance. Consequently, we
have explored the use of signal-to-noise as a statistical metric
more appropriate to the needs of image-based cellular assays.
Results
A common characteristic of cell-based assays (in contrast to
many in vitro assays) is that the statistical variation associated
with control and treated sample populations is often significantly
different (Figure 1). This phenomenon may be both cell type- and
target-dependent. The causes may be manifold, but inherent
cell population heterogeneity, particularly with respect to cell
cycle postion, and methods of image analysis have been noted
as potential contributory factors.
Assay performance assessments based on a signal-to-noise (S:N) method that does not take into account the standard deviation of both control and treated samples will be prone to error. For example, a signal-to-noise metric still cited in screening literature for assessment of in vitro assays involves dividing the magnitude of assay response by the standard deviation of the control (untreated) sample:
We used this S:N calculation to assess assay performance of eight replicate plates imaged and analyzed using the IN Cell Analyzer 1000. The S:N values obtained using this method varied greatly (Figure 2, S:N, method A).
By contrast, S:N values were much more consistent between replicate plates (Figure 2, method B) when we used an alternative method for S:N calculation that takes into account variation of both the control and responding sample populations:
During assay optimization, S:N (method B) can be a more sensitive indicator of assay performance than the Z factor. This is demonstrated by the data shown in Figure 3, where Z factor and S:N values of the same assay performance data are compared.
Since both S:N (method B) and Z factor are based on the same variables, there is a defined relationship between the two metrics, assuming for simplicity that standard deviation of the positive and negative controls are equal:
Although the assumption that standard deviations of positive and negative controls are equal is not always true, we have observed that actual data correspond fairly well to this model relationship. A plot of the relationship between S:N (method B) and Z factor is shown in Figure 4 together with the actual data from an EGFPNFATc1 translocation assay we have developed. The actual data show a very good correlation to the theoretical curve.
As the results in Figure 5 demonstrate, this relationship holds true for a variety of cell-based assays, including FYVE, AKT-1, PLCd-PH, Rac-1, MAPKAP-k2, SMAD2 and NFATc1 (all GFPbased assays performed in live cells).
It can be seen from Figures 4 and 5 that, while Z factor may be a sensitive indicator of assay performance at the lower end of the performance scale (i.e. Z < 0.5, S:N < 8), S:N may be a more sensitive indicator at the higher end of the performance scale. As S:N increases, Z factor approaches 1 asymptotically, making it an increasingly less sensitive measure of performance improvement. For example, the arrows in Figure 5 indicate two assays whose performance could not confidently be distinguished on the basis of Z factor, but which have distinctly different S:N values. Both metrics of assay performance (Z factor and S:N) therefore may be useful in optimizing assay performance.
Ref: Zhang et al J. Biomol. Screening 4, 67-73, 1999.
Acknowledgement
The authors gratefully acknowledge members of the Molecular Cell Biology
Department, Amersham Biosciences, Cardiff, for providing data from a range
of cell-based asssays. We also extend our thanks to Bob Nadon (McGill University)
for his expert technical advice.
CONCLUSION
High information-content image-based cellular assays have
properties distinct from more traditional screening assays,
and their evaluation may therefore require adoption of
additional statistical metrics.
S:N calculations that take into account standard deviation of both positive and negative controls are less prone to error than those derived from the standard deviation of negative controls only.
Both S:N and Z factor may provide valuable information during assay and image analysis optimization.
The theoretical relationship between Z factor and S:N (method B) has been shown to be well supported by the results of several cell-based assays performed using IN Cell Analyzer 3000 and IN Cell Analyzer 1000.
back to top
'"/>
Source: