Mark Churchill
1, Mark Harrison
1,
John Henninge,
2 and Hanne Lund
2
1 Thermo Finnigan, Stafford House, 1 Boundary Park, Boundary Way, Hemel
Hempstead, HP2 7GE, UK;
2 Aker University Hospital, Hormone Laboratory, Section for Doping Analysis,Trondheimsveien,
N-0514, Oslo, Norway
Monitoring Clenbuterol in Humans
The data presented here was acquired on a Thermo
Finnigan TSQ Quantum mass spectrometer
Introduction
Clenbuterol (Figure 1) is a beta-2-adrenergic
agonist, an effective bronchodilator drug used for the treatment of human
asthma. Clenbuterol also relieves bronchial airway smooth muscle contractions
caused by Chronic Obstructive Pulmonary Disease (COPD) and allergyinduced
respiratory distress.
Clenbuterol also has significant anabolic effects and could be used as a
drug of abuse in athletes and livestock for its muscle growth stimulant
properties. It also raises the body temperature and hence facilitates fat
tissue catabolism. Due to Clenbuterol having these anabolic properties,
it must be routinely monitored in biological samples by veterinary and human
doping control laboratories.
Figure 1: Clenbuterol
Goal
One of the limitations to quantitation is the
unequivocal identification of analytes in biological samples due to endogenous
matrix interferents.
This report describes the use of high resolution on the TSQ Quantum to exploit
the negative mass defect of a compound containing Chlorine, such as Clenbuterol,
and hence improve the selectivity of the quantitative assay.
TSQ Quantum Tune View Optimisation of Clenbuterol
Clenbuterol (C
12H
18C
l2N
2O,
molecular weight 276.08) was infused, 0.1 ng/L, into the ESI source and
the four most abundant product ions for the MS/MS breakdown were determined
using the automated compound optimisation procedure in the TSQ Quantum (Figure
2).
The transition yielding the most abundant product ion (m/z 203.0) was selected
for the analysis of Clenbuterol.
Figure 2: Automated optimisation of MS/MS
parameters for Clenbuterol
Experimental Conditions
Sample Preparation: Human urine extracts were
prepared using a Solid Phase Extraction technique (Varian Bond-Elut LRC-18,
methanol as eluent). The extracted urine was spiked with Clenbuterol in
the concentration range 0.1, 0.5, 1, 5, 10, 50 and 100 pg/L for the calibration
standards. No internal standard was used in this study.
Sample Analysis: The spiked urine extracts were chromatographed (Thermo
Finnigan Surveyor System) on a C18 Elite 100 mm 2.1 mm column (Thermo
Hypersil-Keystone) at a flow rate of 300 L/min with a linear gradient of
10% solvent B (Methanol/Ammonium acetate [10mM] 90/10 v/v) to 100% B over
5 minutes. Solvent A was Ammonium acetate (10 mM). The calibration standards
were injected in duplicate at volumes of 10 L.
MS Conditions:
Mass spectrometer:
Thermo Finnigan TSQ Quantum
Ionisation mode:
Electrospray (ESI), positive ion
SRM:
Clenbuterol 277.1 > 203.0 +/- 0.3 Da, 22 V Collision energy
Resolution:
Experiment 1 0.7 Da FWHM on Q1 and Q3
Experiment 2 0.1 Da FWHM on Q1, 0.7 Da FWHM on Q3
Two separate quantitative analyses were performed
at peak widths of 0.1 Da and 0.7 Da Full Width Half Maximum (FWHM) on Q1
in SRM mode. A peak width of 0.7 Da FWHM was used on Q3 for all analyses.
Results
The chromatogram of a pure standard of Clenbuterol
in aqueous solvent demonstrates the retention time at 5.8 minutes (Figure
3).
Figure 3: Determination of Clenbuterol retention
time
Experiment 1: Quantitative Analysis Performed
at 0.7 Da FWHM
The data below shows the quantitative analysis
of Clenbuterol in Human urine at peak width settings of 0.7 Da FWHM on Q1
and Q3. Chromatograms are shown for blank urine (Figure 4) and urine containing
Clenbuterol at 0.1 pg/L (Figure 5).
Figure 4: Urine blank, 0.7 Da FWHM
Figure 5: Clenbuterol, 0.1 pg/L in urine,
0.7 Da FWHM
Experiment 1: Quantitative Analysis Performed
at 0.7 Da FWHM
A calibration curve of Clenbuterol analysed
at 0.7 Da FWHM was constructed using linear fit of peak area plotted against
concentration, weighted 1/X (Figure 6). A correlation coefficient of r2=0.9990
with an equation of Y = 8496.82+266143*X was obtained for the curve.
Figure 6: Clenbuterol curve at 0.7 Da FWHM
The peak area, back-calculated values and precision
of all calibration standards are shown in Figure 7.
Figure 7: Calculated standards at 0.7 Da
FWHM
Experiment 2: Quantitative Analysis Performed
at 0.1 Da FWHM
The data below shows the quantitative analysis
of Clenbuterol in Human urine at peak width settings of 0.1 Da FWHM on Q1
and 0.7 Da FWHM on Q3. Chromatograms are shown for blank urine (Figure 8)
and urine containing Clenbuterol at 0.1 pg/L (Figure 9).
Figure 8: Urine blank, 0.1 Da FWHM
Figure 9: Clenbuterol, 0.1 pg/L in urine,
0.1 Da FWHM
Experiment 2: Quantitative Analysis Performed
at 0.1 Da FWHM
A calibration curve of Clenbuterol analysed
at 0.1 Da FWHM was constructed using linear fit of peak area plotted against
concentration, weighted 1/X (Figure 10). A correlation coefficient of r2=0.9994
with an equation of Y = 2661.76+85951.1*X was obtained for the curve.
Figure 10: Clenbuterol curve at 0.1 Da FWHM
The peak area, back-calculated values and precision
of all calibration standards are shown in Figure 11.
Figure 11:
Calculated standards
at 0.1 Da FWHM
Discussion
Analysis, in SRM mode, of the spiked urine samples at a resolution setting of 0.7
Da FWHM resulted in a Clenbuterol peak eluting from the column upon a broad
chemical noise background signal containing interferent peaks from the urine.
The same urine samples analysed at a peak resolution setting of 0.1 Da FWHM
resulted in elimination of the interfering isobaric mass peaks and the broad
background chemical noise previously seen in the analysis at a peak width
setting of 0.7 Da FWHM. The selected reaction monitoring analysis performed
at a higher resolution setting of 0.1 Da FWHM resulted in increased selectivity
of the assay and hence an increase in the precision could be achieved.
The increase in selectivity at a peak width setting of 0.1 Da FWHM is due to
the fact that Clenbuterol is a chlorinated compound and thus the negative mass
deficiency can be used to eliminate interferents from the urine matrix in SRM
mode. This increased selectivity can be achieved without detrimental loss of
transmission. Typically only a factor of two to three fold decrease in peak area
is observed between analyses performed at 0.7 and 0.1 Da FWHM, however,
greater selectivity could then be achieved.
The calibration curves for Clenbuterol concentrations of between 0.1 to 100
pg/L at resolution settings of 0.1 and 0.7 Da FWHM both demonstrate excellent
linearity. The calibration line at 0.7 Da FWHM showed a high intercept due to
chemical background in the urine blank. This was significantly reduced by the
use of high resolution.
The use of higher resolution to increase selectivity and precision could enable
the limit of quantitation of an assay to be lowered and achieves a higher degree
of confidence in identification of analytes in biological matrices.
'"/>Source:
Page: All 1 2 3 4 5 Related biology technology :1.
The Best Enzyme for the Toughest PCR Challenges Improved with New Hot-Start Feature2.
Improved Sensitivity for Staining RNA with SYBR Gold Stain3.
Improved Vectors for PathDetect Trans-Reporting Systems4.
Gene Expression Arrays:
Highly Sensitive Detection of Expression Patterns
with Improved Tools for Target Amplification5.
The Worlds Best DNase Improved TURBO DNA-free6.
Mx4000 Multiplex Quantitative PCR System*,PPP7.
The LightCycler is the only System for Quantitative PCR that is as Flexible
as
You are!8.
Quantitative Measurement of Cell Proliferation
Using the BrdU ELISA: A Comparison
Between Colorimetric and Chemiluminescent Detection9.
Quantitative end-point RT-PCR gene
expression measurement using the
Agilent 2100 Bioanalyzer and
standardized RT-PCR10.
Detection of Genetically Modified Soybean in Processed Foods Using Real-Time Quantitative PCR with SYBR Green I Dye on the DNA Engine Opticon 2 System11.
Qualitative and Quantitative Assessment of Plasmid DNA Isolation, Rev A