How to select the best spectrofluorometer


Sensitivity: Its more than concentration
Be sure that sensitivity calculations are identical for a valid comparison. SPEX conservatively quotes a minimum signal-to-noise ratio (S/N) of 2500:1 for the FluoroMax. In 2001, the average FluoroMax S/N was closer to 5000:1, and no instrument was shipped with a S/N below 3000:1. Compare this to other manufacturers who quote 500:1.

Let us compare experimental conditions. SPEX calculates S/N via a water-Raman test with 5-nm bandpass. Other companies use a 10nm bandpass on both emission and excitation monochromators. This lets twice the light into each monochromator and twice the light out, a factor of four for each monochromator, and 16 for the entire instrument. Thus, the other companies data are inflated by more than an order of magnitude. The S/N varies as the square root of the signal. Therefore, under similar conditions, the seemingly respectable number of 500:1 is actually closer to 100:1!


So, what does this mean for you?
Certainly the main indicator of sensitivity is the ability to examine lower concentrations. For expensive samples, you need less sample or you consume less probe. Yet there are two other advantages of sensitivity: accuracy and measurement time.
Measurement time: When comparing data with other firms, consider that the square root of the measured number of photons determines the S/N. If two instruments have a S/N difference of a factor of two, one will actually take four times longer than the other to measure the same signal.
Accuracy: The final measurement is only as good as the amount of data going in. To count photons with 1% accuracy, you need at least 10 000 photons. We assume random error and Gaussian statistics, so that the first standard deviation equals the square root of the measured number1,

Quality of electronics does not matter: without 10 000 photons, the accuracy cannot be 1%. (Analog detection has other problems, e.g., hidden noise in the signal, and response times that skew the data.) Hence, an instrument with twice the S/N is twice as accurate under the same conditions.


Where does sensitivity come from?
If everyone needs sensitivity, why doesnt everyone have it? Its a question of economics and quality. Photon-counting electronics, reflective (mirror-based) rather than refractive (lensbased) optics, planar instead of concave gratings, full instrument automation, and all the other high-quality components of a SPEX spectrofluorometer are more expensive. Thus, an accurate, time-saving, sensitive instrument is worth more than one that is not.


How much is your research worth?


1 Strobel, H.A. and Heineman, W.R., Chemical Instrumentation: A Systematic Approach, 3rd ed., Wiley, New York, 1989, p. 345.


back to top


'"/>

Source:


Page: All 1 2

Related biology technology :

1. Tired of the spin? ExpertVoter website makes candidate selection easier
2. San Diego firm selected for Madison Wi-Fi project
3. Miller Brewing selects Procuri software
4. Uniden selects Catalyst RFID software
5. Hammes Co. selects SiteStuff as procurement company
6. Meat-snack maker selects Prescient Applied Intelligence tech package
7. UW Colleges selects Allot Communications to manage multi-campus network
8. ProHealth Care selects NetScaler delivery system
9. Merkle selects Firstlogic data quality software
10. We Energies selects Frictionless SRM Software
11. Southwestern lab selects Third Wave reagents for Hepatitis tests
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:1/14/2017)... , Jan. 13, 2017  The Alliance for Safe ... response to FDA final guidance on biologic ... continued leadership in emphasizing the importance of distinct naming ... aware of the benefits biosimilars will bring to patients, ... Yet the portion of the Guidance dealing with ...
(Date:1/13/2017)... NY (PRWEB) , ... January 13, 2017 , ... ... meet every cuvette-related demand that it has found among its diverse customer base. ... for use in most brands electroporators including BTX and Bio-Rad. FireflySci is ...
(Date:1/12/2017)... ... January 12, 2017 , ... ... performing routine electrochemical biosensing has increased dramatically. Primarily driven by the medical ... quantification of various analytes in complex samples. , Screen-printed electrodes ...
(Date:1/12/2017)... Conn. , Jan. 12, 2017   ... company and maker of Flublok Influenza Vaccine ... Zika vaccine candidate had good safety results and ... in preclinical studies. The product is expected to ... few months.  In addition, the Institute of Technology ...
Breaking Biology Technology:
(Date:12/15/2016)... Germany , December 15, 2016 ... today announced an agreement with NuData Security, an award-winning ... The partnership will enable clients to focus on good customer ... data protection regulation. ... In order to provide a one-stop fraud prevention ...
(Date:12/15/2016)... , Dec 15, 2016 ... Research and Markets has announced the addition ... offering. The report forecasts the global military biometrics market ... The report has been prepared based on an in-depth market ... and its growth prospects over the coming years. The report also includes ...
(Date:12/15/2016)... 2016  There is much more to innovative access ... engine. Continental will demonstrate the intelligence of today,s solutions ... . Through the combination of the keyless entry and ... elements, the international technology company is opening up new ... "The integration of biometric elements brings our ...
Breaking Biology News(10 mins):