High resolution DNA analysis with the,,,DNA 500 and DNA 1000 LabChip kits

Slab gel analysis is the traditional technique used for DNA analysis. By comparing the separated bands of an unknown sample to a known standard (size or mass ladder), the size and quantitation of each fragment can be estimated. The Agilent 2100 bioanalyzer is an instrument, which utilizes LabChip technology to achieve DNA and RNA analysis on micro-fabricated chips. This approach entails significant advantages including increased speed of analysis, improved data precision, and ease of use.
In contrast to slab gel electrophoresis, the Agilent 2100 bioanalyzer in conjunction with LabChip kits performs chipbased separations under standardized conditions, resulting in more reproducible results. Using prepackaged reagents and an assay specific software script that defines all instrumental parameters, a user can always obtain equivalent results for a given sample. This approach simplifies sample preparation, minimizes user errors, and facilitates the exchange of data between different users and laboratories.
In the context of LabChip Technology, several DNA sizing kits are available to cover a wide range of DNA fragment sizes. In addition to the DNA 500 and DNA 1000 kits, two other DNA LabChip kits are available: the DNA 7500 and DNA 12000 LabChip kits. They allow separation of DNA fragments up to 12000 base pairs. The DNA 500 LabChip kit for analysis of DNA fragments, ranging in size from 25 to 500 base pairs, and the DNA 1000 LabChip kit for analysis of DNA fragments, ranging in size from 25 to 1000 base pairs, supplement these two kits. The size range of the DNA 500 and DNA 1000 kits is especially useful for small to medium sized PCR and RT-PCR products and can resolve primer dimers. Other relevant applications include competitive PCR and cleavage based applications such as RFLP. Also, the resolution in the lower range, i.e. below 100 bp, is significantly increased compared to the other two DNA LabChip kits, making it possible to distinguish between fragments that are very similar in size. Since the DNA 500 and the DNA 1000 LabChip kit show similar performance and are used for similar applications, they are presented jointly in this application note.
DNA 500 LabChip kit
The performance of the DNA 500 assay is demonstrated by the analysis of several mixtures of PCR products (figure 1). All chipbased separations were performed on the Agilent 2100 bioanalyzer according to the instructions provided with each kit. The conditions for the acrylamide gel separations are listed in the appropriate figure caption. Figure 1 shows the overlay of 3 different electropherograms acquired on the Agilent 2100 bioanalyzer that are mixtures of PCR samples ranging from 25 to 500 base pairs in size. The two closest eluting bands (50 bp and 53 bp) are partially separated and identified by the software as two separate peaks. The DNA 500 assay achieves a resolution of 5 base pairs from 25 to 100 base pairs and a 5 % resolution from 100 to 500 base pairs where the sizing error is less than 10 % over the entire size range. The digital data format allows accurate quantitation of the separated bands based on peak areas. A DNA mass ladder was used to verify quantitation accuracy. The ladder contains 3 fragments in the specified size range, with concentrations guaranteed by the supplier (Low DNA Mass ladder, Life Technologies, USA).

Table 1 lists the quantitative results obtained on the Agilent 2100 bioanalyzer. All separated bands were well within the specified margin of 70 % accuracy, with a standard deviation of 10 % or less. The DNA 500 assay relies, like the other DNA assays, on an intercalating dye for fluorescence detection. The dye incorporation depends on the basepair composition and the secondary structure of the DNA fragments. For larger fragments such sequence differences are usually statistically evened out. For shorter fragments, however, the sequence can make a difference in dye incorporation. Therefore, deviations from the sizing and quantitative specifications can occur for individual fragments below 70 bp. The same samples shown in the chip-based electropherograms in figure 1 were compared to a slab gel separation using a commercially available 4- 20 % gradient gel (Novex, USA).

As shown in figure 2, the Agilent 2100 bioanalyzer performs with equal or better resolution than the slab gel, delivering a clearer image of the sample.

DNA 1000 LabChip kit
In the size range below 500 bp, the DNA 1000 LabChip kit shows a performance that is comparable to the DNA 500 LabChip kit. The operating range was extended to 1000 bp, to better match the size range, in which most PCR fragments fall. From 500 to 1000 bp the resolution of the DNA 1000 assay is 10 %, while the maximum error for quantitation remains at 30 % (this corresponds to a 95 % confidence level).
The DNA 1000 kit was used to optimize PCR conditions for a competitive RT-PCR reaction of two genes: GAPDH and the heat shock protein hsp721 . The goal of the experiment was to achieve a co-amplification of GAPDH (housekeeping gene) and hsp72. The hsp72 was the gene of interest, the expression of which can be influenced by various factors. For the experiment, RNA was isolated from HepG2-cells and reverse transcribed into cDNA. The cDNA was amp lified using a specific set of primers to result in a 439 bp fragment for the amplification of GAPDH and a 650 bp fragment for the amplification of hsp72. A primer dropping method was employed because the two genes are expressed at different levels within the cell. In the present case this means that hsp72 was amplified for 40 cycles and the primers for the amplification of GAPDH were added at the beginning, after 2 cycles, 4 cycles and so on, resulting in 20 to 40 amplification cycles. Since the PCR reaction was carried out under competitive conditions, the PCR product of hsp72, though at a constant amplification of 40 cycles, disappears with increasing cycle number of GAPDH (figure 3).

Figure 3 shows the separation of the two PCR products as a gellike image and also shows a plot of the concentrations that were calculated by the bioanalyzer for the different amplification conditions. From the results it was concluded that a primer dropping method with 40 cycles for hsp72 and 36 cycles GAPDH would be preferable. Under these conditions a large GAPDH and a small hsp72 peak are visible. This allows monitoring up-regulation of the hsp72 gene in future experiments. For other experiments, a different number of cycles might be preferable.
The advantage of using the DNA 1000 assay for the analysis of competitive PCR products lies in the accurate absolute and relative quantitation of each amplified product. Small differences in the amplified amount, that cannot be detected using slab gel analysis, are easily analyzed with the DNA 1000 assay. This allows not only the optimization of PCR conditions but also detection of changes in gene expression via RT-PCR methods.
Conclusion
The DNA 500 and DNA 1000 assay perform DNA separations reliably in the range of 25 to 500 base pairs and 25 to 1000 bp, respectively. Both assays are optimized to give accurate sizing and quantitation of PCR products, which is needed for numerous PCR-based methods, such as multiplex PCR, RT-PCR, competitive PCR, and so on. They supplement the two other DNA assay kits to allow analysis of a wide range of samples. LabChip technology allows the molecular biologist to not only perform nucleic acid analyses in a more accurate way and with greater ease of use but also requires less time and sample. With additional automated data analysis and storage capabilities, a complete package is offered enabling precise and well-documented experiments. Considering the advantages of Lab-on-a-Chip technology, the chip-based approach outperforms slab gels and represents an indispensable tool for molecular biologists.



'"/>

Source:


Page: All 1 2 3 4 5

Related biology technology :

1. Analysis of urine and seawater samples by ultrasonic nebulization with a high resolution ICP spectrometer
2. High-throughput gene expression analysis to screen for anti-asthma drugs
3. Fully automated DNA purification and efficient multiplex PCR for analysis of microsatellite loci
4. Flow cytometric analysis of human primary cells using the Agilent 2100 bioanalyzer and on-chip staining
5. 2100 expert software Powerful software for the analysis of RNA, DNA, proteins, and cells with the Agilent 2100 bioanalyzer
6. Using the Agilent 2100 bioanalyzer for quality control of protein samples prior to MS-analysis
7. Microfluidic analysis of multiplex PCR products for the genotyping of Helicobacter pylori
8. Extended data analysis using DeCyder EDA
9. 2D-LC analysis of phosphopeptides in brain tissue using Ettan MDLC and Finnigan LTQ
10. Neurite outgrowth cell-by-cell analysis using the IN Cell Developer Toolbox
11. Laser Capture Microdissection of muscle fiber populations and expression analysis by RT-PCR
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:6/24/2016)... DIEGO , June 24, 2016 ... more sensitively detects cancers susceptible to PARP inhibitors ... circulating tumor cells (CTCs). The new test has ... HRD-targeted therapeutics in multiple cancer types. ... targeting DNA damage response pathways, including PARP, ATM, ...
(Date:6/23/2016)... ... June 23, 2016 , ... UAS ... the launch of their brand, UP4™ Probiotics, into Target stores nationwide. The company, ... proud to add Target to its list of well-respected retailers. This list includes ...
(Date:6/23/2016)... , June 23, 2016 Houston Methodist ... the Cy-Fair Sports Association to serve as their ... agreement, Houston Methodist Willowbrook will provide sponsorship support, ... connectivity with association coaches, volunteers, athletes and families. ... the Cy-Fair Sports Association and to bring Houston ...
(Date:6/23/2016)... SAN FRANCISCO , June 23, 2016   ... it has secured $1 million in debt financing from ... to ramp up automation and to advance its drug ... for its new facility. "SVB has been ... goes beyond the services a traditional bank would provide," ...
Breaking Biology Technology:
(Date:6/21/2016)... VANCOUVER, British Columbia , June 21, 2016 ... been appointed to the new role of principal ... has been named the director of customer development. ... , NuData,s chief technical officer. The moves reflect ... development teams in response to high customer demand ...
(Date:6/9/2016)... 9, 2016  Perkotek an innovation leader in attendance control systems is proud to ... hours, for employers to make sure the right employees are actually signing in, and ... ... ... ...
(Date:6/2/2016)... 2016 The Department of Transport Management ... 44 million US Dollar project, for the , ... Personalization, Enrolment, and IT Infrastructure , to ... and implementation of Identity Management Solutions. Numerous renowned international vendors ... Decatur was selected for the most compliant and ...
Breaking Biology News(10 mins):