Dispensing 0.5μL to "Dry" 384 Plates with the Sciclone ALH

Mike Benway, Larry Simonson and John Ung
Caliper Life Sciences Inc, Applications Laboratory, Hopkinton, Massachusetts, USA

Introduction

"Assay-Ready" plates prepared from compound or library stock plates in real time greatly accelerates the High Throughput Screening (HTS) process. The ability to transfer and dilute 0.5μL volumes is a prerequisite for this function. Here we present preliminary data on 0.5μL dry dispensing using the Caliper Life Sciences Sciclone Advanced Liquid Handler (ALH) Workstation, configured with a Low Volume Head and 384/1536 Fixed-Cannula Array.

Materials and Methods

Instrumentation

All testing was performed using a Sciclone ALH with a Low Volume Dispensing Head, and a Bulk Reagent Dispenser (Caliper Life Sciences). Absorbance was measured at 450nm using a Wallac Victor2 1420 Multilabel Counter (PerkinElmer).

Reagents

Tartrazine was obtained from Sigma Chemical Company, Catalogue number T-0388. A stock dispensing solution was prepared by dissolving 2.8g of tartrazine into 1L of DMSO. The stock solution was prepared for dispensing volumes of 0.5μL. Deionized water (DI) was used as a diluent after dispensing into the dry plate.

Accuracy Determination

Dispense accuracy is determined before measuring dry dispense precision. A calibration is performed to determine the volume dispensed. For the calibration, 50μL of deionized water is pipetted in to a 384 Well Flat Bottom, Untreated Polystyrene Assay Plates (COSTAR, Catalogue # 3702). Next, 0.5μL of the tartrazine solution is manually injected into each of 20 wells, successively, from a filled 10μL Hamilton micro syringe. The wells are manually mixed by multiple aspirate- dispense cycles using a 50μL hand pipettor. Any air bubbles are removed, and the plate is read for absorbance. The average absorbance reading for these manual standards is used in a single point calibration routine.

Sciclone Protocols

Here a single aspiration step is performed, followed by priming dispenses, then incremental dispenses into dry plates as follows:

Data Acquisition and Processing

Absorbance data is acquired on a Wallac Victor2 reader using the same read protocol as for the calibration plate. The raw results are exported to ExcelTM and %CV is calculated. The volume dispensed is calculated by comparison of signal to a manually prepared calibration standard.

Results

Testing has been done for three and five incremental plates using the method described above. For each run, the volume dispensed and %CV for each plate is calculated and summarized (Table 1). The results for the two runs are represented graphically (Figure 3).

Discussion

In this preliminary work we have found that the Sciclone ALH equipped with a Low Volume Head is capable of delivering 0.5μL of liquid into dry microplates with a precision of less than 6%CV. This is performed with no direct contact between the cannulas and the plate. In the above results, a "first plate" effect is observed, in which the first plate in a series has a higher volume. We believe this is due to residual tartrazine solution of the external surfaces of the cannulas that is removed as the first plate is dispensed.

Initial dry plate work was carried out using a 384 Cannula Array. The benefits of this array are that it can access both deep well plates and standard microplates. However, the larger cannulas do not permit formation of well formed droplets. Without such droplets, physical contact of the cannulas with the plates is required when transferring sub-microliter volumes to dry plates, potentially resulting in variable %CV's (unpublished data). For the above results, we used a shorter, finer 384/1536 Cannula Array designed for accessing 384 well plates or 1536 plates (by dispensing into four quadrants). This type of a cannula array can fully access standard microplates, or half height deep well plates. The main benefit of the smaller cannulas is that uniform droplets can form on the ends of the finer cannulas, allowing the dispense of the droplet onto the plate without contacting the plate.

The success of dry dispensing depends on assuring the Cannula Array and destination plates are parallel. Therefore, it is necessary to level any position in which sub-microliter dry dispensing is attempted. The deck of the Sciclone ALH is designed to be adjustable at each corner. The individual plate locators must also be level. It is also necessary to determine a precise dispense height for any non-contact dispense operation.

Physical factors are an important determinant of overall dispensing performance. Such factors include microplate composition, microplate surface treatments, solvent concentrations and compositions, solute (compound) characteristics and solute concentration. The current study represents 2.8g/L tartrazine dissolved with DMSO and dispensed into untreated, flat-bottomed, polystyrene plates only. Under these conditions we demonstrated that it is feasible to dispense sub microliter volumes to dry plates with very good precision.


'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. Counting Colonies or Plaques Directly from Images of Agar Plates
2. Counting FlashPlates on the Wallac MicroBeta
3. Direct Counting of Millipore MultiScreen Filtration Plates.
4. Sciclone ALH Bulk Dispense Precision Testing
5. 96 Well Pipetting Precision Testing Using the Sciclone ALH with a 384 Fixed-Cannula Array
6. Sciclone ALH 3000 Liquid Handler: Functional Characterization of the High Volume Head - Part 1: Plate-to-Plate Precision Testing
7. Scalable Automation of the TempliPhi DNA Sequencing Template Amplification and DYEnamic ET Terminator Sequencing Kits on the Sciclone ALH 3000
8. Precision Testing for the Sciclone Advanced Liquid Handler Z8 Option
9. 384 Well Pipetting Precision and Accuracy Testing Using the Sciclone Advanced Liquid Handler
10. Automation of the Eppendorf Perfectprep BAC 96 DNA Kit on the Caliper Life Sciences Sciclone ALH 3000
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Dispensing Dry Plates with the Sciclone ALH

(Date:8/31/2015)... Aug. 31, 2015 LabStyle Innovations Corp. ... Solution, today announced it recently closed a round ... be used to support the global rollout and ... mobile, cloud-based, diabetes management solution that includes novel ... blood glucose monitoring device, and for working capital ...
(Date:8/31/2015)... ... August 31, 2015 , ... There’s a new treatment ... is the first commercially available new FDA approved system in the last 14 years, ... 92%, the high Amenorrhea (zero bleeding) rate of 66%, the high Patient Satisfaction rate ...
(Date:8/28/2015)... , August 28, 2015 ... Product (Media, Sera, Cryogenic Storage Systems, Thawing Equipment, Alarms), ... Regenerative Medicine, Clinical Trials), End User - Global Forecast ... is expected to reach USD 3,731.03 Million by 2020 ... CAGR of 11.65% between 2015 and 2020. ...
(Date:8/28/2015)... ... 28, 2015 , ... Riordan-McKenna Institute founders, stem ... that the use of sterile, dehydrated amniotic tissue AlphaPATCH™ developed by Amniotic ... surgical knee wound. , The case involved a 78-year-old male, who presented with ...
Breaking Biology Technology:LabStyle Innovations Closes $2.5 Million Funding to Support Global Rollout of Dario Diabetes Management Solution 2LabStyle Innovations Closes $2.5 Million Funding to Support Global Rollout of Dario Diabetes Management Solution 3LabStyle Innovations Closes $2.5 Million Funding to Support Global Rollout of Dario Diabetes Management Solution 4U.S. FDA Approves Minerva Endometrial Ablation System 2U.S. FDA Approves Minerva Endometrial Ablation System 3Biopreservation Market Worth 3,731.03 Million USD by 2020 2Biopreservation Market Worth 3,731.03 Million USD by 2020 3Biopreservation Market Worth 3,731.03 Million USD by 2020 4Biopreservation Market Worth 3,731.03 Million USD by 2020 5Dallas-Fort Worth Stem Cell Researchers Use Amniotic Tissue To Successfully Treat Non-Healing Surgical Wound 2Dallas-Fort Worth Stem Cell Researchers Use Amniotic Tissue To Successfully Treat Non-Healing Surgical Wound 3Dallas-Fort Worth Stem Cell Researchers Use Amniotic Tissue To Successfully Treat Non-Healing Surgical Wound 4
... June 24 The Board of,Directors of Schering-Plough Corporation ... 6.5 cents per common share. Payment will be made ... close of business on Aug.,1, 2008. As of March ... Board of Directors today also declared a quarterly dividend ...
... Sun Pharmaceutical,Industries Ltd. (Reuters: SUN.BO, Bloomberg: SUNP IN, NSE: ... Levitt, Chairman of the Board,of Directors of Taro Pharmaceutical ... Levitt,s letter of June 19, 2008 to Taro,s,shareholders, his ... the Initiating,Motion filed by Taro in the Tel-Aviv District ...
... etiquette tips via live, interactive ... ... a new way for life science post graduates to enhance their skills,and ... a dual-purpose service that,provides employers a chance to meet potential candidates, and ...
Cached Biology Technology:Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 2Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 3Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 4Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 5Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 6Sun Pharmaceutical Sends Second Letter to Taro Pharmaceutical Regarding Taro's Irish Subsidiary 7Online Career Tool Offers Live 'Expert Advice' Component 2
(Date:8/11/2015)... GOTHENBURG, Sweden , August 11, 2015 ... sensor FPC1155. Already received as well as expected revenues in 2015 ... included in FPC,s communicated revenue guidance of approximately 2,200 MSEK for ... is a prominent smartphone manufacturer ... ZUK selected FPC1 155 ...
(Date:8/6/2015)... Calif. , Aug. 6, 2015 ... developer of human interface solutions, today announced collaboration ... technologies on Windows 10. Microsoft leveraged Synaptics, deep ... the operating system. Through stringent ... are fully certified with Microsoft,s Precision TouchPad (PTP) ...
(Date:8/4/2015)...  AMRI (NASDAQ: AMRI ) today reported financial and operating ... Highlights: , Second quarter contract revenue of $ ... , Adjusted contract margins of 26 % ... $0.05 decrease in EPS from royalties in the current quarter ... "We are very pleased to present another strong financial quarter, ...
Breaking Biology News(10 mins):FPC's Touch Fingerprint Sensor FPC1155 in ZUK's Smartphone Z1 2Synaptics Chosen for Human Interface Co-Development on Windows 10 2AMRI Announces Second Quarter 2015 Results 2AMRI Announces Second Quarter 2015 Results 3AMRI Announces Second Quarter 2015 Results 4AMRI Announces Second Quarter 2015 Results 5AMRI Announces Second Quarter 2015 Results 6AMRI Announces Second Quarter 2015 Results 7AMRI Announces Second Quarter 2015 Results 8AMRI Announces Second Quarter 2015 Results 9AMRI Announces Second Quarter 2015 Results 10AMRI Announces Second Quarter 2015 Results 11AMRI Announces Second Quarter 2015 Results 12AMRI Announces Second Quarter 2015 Results 13AMRI Announces Second Quarter 2015 Results 14AMRI Announces Second Quarter 2015 Results 15AMRI Announces Second Quarter 2015 Results 16AMRI Announces Second Quarter 2015 Results 17AMRI Announces Second Quarter 2015 Results 18AMRI Announces Second Quarter 2015 Results 19AMRI Announces Second Quarter 2015 Results 20AMRI Announces Second Quarter 2015 Results 21AMRI Announces Second Quarter 2015 Results 22AMRI Announces Second Quarter 2015 Results 23AMRI Announces Second Quarter 2015 Results 24
... has comparable efficacy to chemotherapy, and is better ... has progressed quickly after treatment with first-line therapy, ... show. Dr Tudor Ciuleanu from the Institute ... finding from the international TITAN study at the ...
... have found that when growing corn crops for ethanol, more ... edition of the American Chemical Society,s journal Environmental Science ... fertilizer while they improve their production of feedstock for ethanol ... implications for an industry that has grown dramatically in recent ...
... York University,s Clinical and Translational Science Institute (CTSI) has ... grant to assess the feasibility of using gingival crevicular ... a blood glucose measurement -- as a means of ... between New York University,s Langone Medical Center and the ...
Cached Biology News:Erlotinib effective and with fewer side-effects after first-line treatment 2Overfertilizing corn undermines ethanol 2Overfertilizing corn undermines ethanol 3NYU nursing-dental team receives grant to assess effectiveness of A1C diabetes screening technique 2NYU nursing-dental team receives grant to assess effectiveness of A1C diabetes screening technique 3
... is a 20 amino acid synthetic peptide whose sequence corresponds to ... is (amino to carboxy terminus): V115 - L - P ... P - Q - V - P - N - E ... (The C-terminal cysteine has been added to facilitate conjugation and is ...
... The Rapid-Screen cDNA Library Panels were ... just three sets of PCRs to identify ... analysis is performed in a 96-well "Master ... from 5,000 clones. Having identified the positive ...
... Library Panels were designed for full-length gene ... PCRs to identify the desired cDNA clone. ... a 96-well "Master Plate," where each well ... identified the positive well(s) by gel electrophoresis ...
... Pathway 435 is an affordable, compact ... and slides, providing a broad array ... rapidly acquires high-resolution images with high-performance ... computer-controlled excitation and emission filter changers, ...
Biology Products: