FlashPlates and ScintiPlates are popular platforms for running a variety of radiometric screening assays. The Wallac MicroBeta effectively measures both plate types.
MicroBeta detectors comprise two photomultiplier tubes, one above the sample and one below.
FlashPlate is a solid white plate and so MicroBeta measurements are conducted with the upper PM tube only. A comparison is made of MicroBeta with a Packard TopCount, an instrument that employs single PM tube detectors.
ScintiPlate has clear wells so when measured on MicroBeta both upper and lower PM tubes are used. A comparison of MicroBeta performance is made when measuring the same assay with both FlashPlate and ScintiPlate.
APPLICATION I
A Reverse Transcriptase Assay was performed using a Streptavidin coated FlashPlate (SMP103). In this assay, a biotinylated DNA oligonucleotide acts as a primer for reverse transcriptase (RT) on an RNA template. The signal is generated by incorporation of labeled dTTP into the newly synthesized strand of DNA. Plates were read on both a TopCount and a MicroBeta, and as shown, the MicroBeta effectively reads with only the top PM tubes in use.
APPLICATION II
To offer as diverse a range of applications as possible, there are a variety of FlashPlate and ScintiPlate coatings and surface treatments available.
The following experiment was devised to show that although using both PM tubes gives ultimate sensitivity, for example, when measuring adherent cells in ScintiPlates, MicroBeta still provides excellent performance with upper tube only counting.
[3H]Biotin was titrated on both Streptavidin coated FlashPlates and ScintiPlates, then bound activity was read on the MicroBeta. The mass bound was determined by the differential between the known mass/DPM added and the unbound mass/DPM, using the known specific activity for the compound and the known efficiency of a LSC counter. The MicroBeta proved to read the opaque plate from the top as effectively as the clear well plate read from the top and bottom.
Instructions for counting FlashPlates on the MicroBeta. Software version 2.7
Create a new normalization
Select Other for the label
Select Background Sample
Click on Extended
Click on PMT Usage
Select Upper and click on OK
Change the window setting to:
H-3, 175 360
I-125, 175 530
C-14, 175 650
S-35, 175 650
P-33, 175 - 960
Click on OK
Click on New in the Plate/Filter section
Enter the name of the plate (i.e. FlashPlate). No other changes from the default protocol are necessary.
Important! Change the window setting after selecting the PMT usage. Otherwise the program will reset the window to 150 - 1024
Instructions for counting FlashPlates on the MicroBeta. Software version 3.0
Create a new normalization
Select Other for the label
Corrections Tab
Select Background sample
Other Tab
Select Upper tube only
Change the window setting to:
H-3, 175 360
I-125, 175 530
C-14, 175 650
S-35, 175 650
P-33, 175 - 960
Save the normalization protocol and then check the plate map. Two plates are required for FlashPlate normalization. The first one is an empty plate that is used for determining the background of each of the detectors. For example, a six-detector instrument will store six background values and subtract the appropriate value from each unknown - detector specific background subtraction. The second plate has the active sample in G11 (96-well plate) or M22 (384-well plate). It is advisable to acquire at least 10,000 counts for each reading.
Next, create a counting protocol. The key selections, Other, upper PM, and window setting must be identical to the normalization or the normalization will not be visible from within the counting protocol.
Now, when unknowns are selected, the CPM1 is the raw value and the CCPM1 is corrected for background and efficiency.
To avoid phosphorescence, both normalization plates and assay plates should be dark-adapted for a couple of minutes before counting.
FlashPlate is a registered trademark of Packard Instrument Company, exclusively licensed to PerkinElmer Life Sciences, Inc. FlashPlate is protected under US Patent 5,496,502 and foreign equivalents, to all of which PerkinElmer Life Sciences holds an exclusive worldwide license. TopCount is a registered trademark of Packard Instrument Co.