CleanSEQ on the ABI 3100


Removal of unincorporated dye-terminators is required prior to sequence delineation on capillary electrophoresis instrumentation. CleanSEQ has proven an efficient and highly effective dye-terminator removal system with the ABI 3700 and 3730xl sequencers. Samples run on the ABI 3100 are more prone to evaporation and exposure to air before machine injection which can have deleterious effects on sample integrity including the appearance of dye blobs. This application note will demonstrate the effective use of CleanSEQ for dye-terminator removal in combination with the ABI 3100.


Introduction
CleanSEQ has proven to outperform many other dye-terminator removal technologies when used in conjunction with the high throughput ABI PRISM 3700 and 3730xl sequencers. The ABI 3100 is a lower throughput instrument that unlike the 3700 and 3730xl systems, has historically given more reproducible results with formamide elution and injection. Formamide acts to reduce sample evaporation which helps maintain sample integrity. Formamide, however, is well known to significantly reduce signal strengths.

CleanSEQ is a magnetic bead dye-terminator removal system based on SPRI chemistry. Experimentation during CleanSEQ development indicated that samples eluted in water were of high quality and produced sequences with long Phred20 read lengths and much higher signal intensities than samples eluted in formamide. Our standard CleanSEQ protocol, therefore, recommends the elution of sequencing products in water (Figure 1). We believe the septa seals on the 3100 may not effectively prevent exposure to air, potentially s ubjecting samples eluted in water to evaporation,

degradation and the appearance of dye blobs in sequence traces. We tested a number of CleanSEQ elution buffers and procedures in an effort to inhibit sample evaporation and prevent sample degradation, thereby improving sequence quality on the 3100 (data not shown). As result of these studies, we recommend the elution of samples from CleanSEQ microparticles using water followed by application of mineral oil.

We collectively ran over 1000 samples using our recommended oil overlay protocol and found significantly increased signal strengths and extended protection of samples from degradation compared with formamide elution. Importantly, no decrease in capillary performance was observed due to mineral oil usage. This simple procedure results in high quality sequence reads and eliminates the appearance of dye blobs (Figure 2).


Experimental Methods
pGEM was sequenced using v3.1 BigDye Terminator chemistry at 1/4x dilution. Sequencing reactions were purified using the standard CleanSEQ 96 protocol. Samples were eluted in either 50l of 100% formamide, 50l of water, or 50l of water followed by application of 20l of mineral oil. No sample transfers were required* from the time of sequencing set up through loading onto the 3100 with POP6 polymer. *Plate transfer required when using POP4 polymer.


Higher Signal Strengths using Water Elution
We compared signal strengths and signal to noise ratios for samples eluted in water, water with mineral oil application and formamide. Signal strengths are significantly higher using water elution com pared to those observed using formamide and even higher if water elution is followed by the addition of mineral oil (Table 1). We propose that mineral oil prevents evaporation and blocks oxygen exchange with the sample, thereby slowing dye degradation and maintaining high signal to noise ratios. The increased signal to noise ratios produced using CleanSEQ with water elution allow extensive dilution of BigDye Terminators without affecting Phred20 read lengths (data not shown).


Extended Sample Stability
We assessed the stability of samples eluted in water and overlaid with mineral oil by comparing sequencing results from plates run immediately following CleanSEQ elution (fresh) versus samples reinjected 6 days later. Samples eluted in formamide were used as references, because they are known to be refractory to sample degradation. Both signal strength and Phred20 read length were unaffected by prior injection regardless of elution buffer (Figure 3). Signal to noise ratios were, however, far greater for samples eluted in water. These results clearly demonstrate that overlaying mineral oil stabilizes samples as well as formamide and allows flexible injection schedules.


Instrument Compatibility
Because much is proprietary about the manufacturing of ABI capillaries, Agencourt conducted thorough testing to confirm the compatibility of mineral oil and the 3100 capillary array. We injected samples eluted with water and overlaid with mineral oil in the same capillaries for 99 consecutive runs. Formamide samples were run in parallel in alternate capillaries as a base line of capil lary function. Samples loaded in formamide or in water overlaid with mineral oil had identical patterns of Phred20 read lengths over time (Figure 4). Thus, mineral oil had no detrimental effects on capillary performance.


Conclusions
We recommend eluting samples from CleanSEQ beads with 50l of water followed by application of 20l of mineral oil for use with ABI 3100 sequencers.
CleanSEQ samples eluted in water and overlaid with oil show no sample degradation after 6 days at room temperature, providing injection flexibility and eliminating the appearance of dye blobs.
Increased signal strengths using CleanSEQ and water elution allow increased dilution of BigDye Terminators resulting in substantial cost savings.
Injection of samples containing mineral oil does not adversely effect either short or long-term capillary performance.


Available For Use In Your Laboratory
To Place an Order or Receive Technical Assistance: In the U.S.and Canada call us at 800-361-7780. Outside the U.S. and Canada call us at 978-867-2603. Visit us at www.agencourt.com.


back to top
'"/>

Source:


Page: All 1 2 3 4

Related biology technology :

1. Genomic DNA Extraction from Buccal Swabs Using the Perfect gDNA Blood Mini Kit
2. Genomic DNA Extraction from Buccal Swabs Using the Perfect gDNA Blood Mini Kit
3. Protocol for GenomePlex Whole Genome Amplification from Buccal Swab
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:12/5/2016)... HAMILTON, Bermuda , Dec. 5, 2016 ... leading clinical-stage biopharmaceutical company focused on the treatment of ... of intepirdine for the treatment of Alzheimer,s disease will ... Disease (CTAD) Meeting on Friday, December 9, 2016 in ... CTAD will show results of both simple and complex ...
(Date:12/5/2016)... , ... December 05, 2016 , ... ... to soon resume cervical and lumbar disc production, company President, Jake Lubinski will ... who are implanting the AxioMed disc in Cologne and Karlsruhe to discuss the ...
(Date:12/5/2016)... NxGen MDx announced today that it brought its NxGen Informed ... been able to improve customer service through shortened turnaround times and at ... , CEO of NxGen MDx. ... A decrease in turnaround times by ... job opportunities at the Grand Rapid headquarters. The NxGen Informed ...
(Date:12/4/2016)... ... December 02, 2016 , ... A proposed five-year extension ... federally funded research and development is welcome news for the photonics community, say ... , As part of the National Defense Authorization Act (NDAA) compromise agreement finalized ...
Breaking Biology Technology:
(Date:11/15/2016)... 15, 2016  Synthetic Biologics, Inc. (NYSE MKT: ... on the gut microbiome, today announced the pricing ... of its common stock and warrants to purchase ... price to the public of $1.00 per share ... from the offering, excluding the proceeds, if any ...
(Date:6/27/2016)... Research and Markets has announced the addition of the "Biometrics ... The report forecasts the biometrics market ... CAGR of 12.28% during the period 2016-2020. The ... inputs from industry experts. The report covers the market landscape and ... a discussion of the key vendors operating in this market. ...
(Date:6/21/2016)... , June 21, 2016 NuData ... the new role of principal product architect and ... the director of customer development. Both will report ... technical officer. The moves reflect NuData,s strategic growth ... response to high customer demand and customer focus ...
Breaking Biology News(10 mins):