Bioanalytical Method Intraday Validation for the Quantitation of Paroxetine ,,, in Bovine Plasma using the TSQ Quantum Mass Spectrometer

Mark Churchill and Mark Harrison, Thermo Finnigan, Hemel Hempstead, UK

Bioanalytical Method Validation

The data presented here can be acquired using a Thermo Finnigan TSQ Quantum mass spectrometer Introduction Guidelines on Bioanalytical Method Validation acceptance criteria were established by the US Food and Drug Administration (FDA) from a conference on Analytical Methods Validation Bioavailability, Bioequivalence and Pharmacokinetic Studies held in Arlington,VA, December 3-5, 1990.1 These acceptance criteria were reviewed ten years on at Bioanalytical Methods Validation A Revisit with a Decade of Progress, Arlington,VA, January 12-14, 2000,2 a meeting sponsored by the American Association of Pharmaceutical Scientists and the FDA. This led to the issue of the FDA Guidance for Industry document on Bioanalytical Method Validation in May 20013. A full method validation should assess accuracy, precision, selectivity, sensitivity, reproducibility and stability of an analyte in matrix samples.

Paroxetine is an orally administered antidepressant which acts as a selective serotonin reuptake inhibitor (SSRI) but has only minimal effects on the neuronal reuptake of norepinephrine and dopamine. Paroxetine is administered for the treatment of depression, social anxiety disorder, obsessive compulsive disorder, panic disorder and generalized anxiety disorder. Goal This report describes the ease of use of the TSQ Quantum to validate bioanalytical methods in accordance with the guidelines issued by the US Food and Drug Administration.

The aim of this investigation was to demonstrate the sensitivity, linearity, and robustness of the TSQ Quantum to allow researchers to achieve excellent accuracy and precision for highthroughput quantitation of analytes in biological matrices. Paroxetine spiked into bovine plasma was injected as an intra-day batch consisting of blanks, calibration standards, and QCs at eight replicates. TSQ Quantum Tune View Optimization of Paroxetine Paroxetine (C19H20FNO3, molecular weight 329.37) was infused, 0.1 ng/L, into the ESI source and the four most abundant product ions for the MS/MS breakdown were determined using the automated compound optimization procedure in the TSQ Quantum (Figure 1).

The transition yielding the most abundant product ion (m/z 192.1) was used for the analysis of Paroxetine. Figure 1: Automated optimization of MS/MS parameters for Paroxetine Experimental Conditions Sample Preparation: Aliquots (1 milliliter) of Bovine Plasma (Sigma P4639) were spiked with Paroxetine (Figure 2) in the concentration range of 0.1, 0.2, 0.5, 1, 2, 5, 10, 50, 100, 200, 500 and 1000 pg/L for the calibration standards and at levels of 0.3, 25 and 750 pg/L for quality control samples. De-Fluoro-Paroxetine (Internal Standard, Figure 3) was spiked at a concentration of 25 pg/L. Blank Bovine plasma and plasma containing internal standard were also prepared. Eight replicate plasma aliquots of 100 L were taken from each spiked level and protein precipitated using 250 L of Acetonitrile, centrifuged at 13,000 rpm for 10 minutes and the supernatants decanted. The extracts were evaporated under a stream of nitrogen gas and reconstituted in 100 L of Water/Methanol/Acetic acid (80/20/0.1 v/v/v).

Sample Analysis: The plasma extracts were chromatographed (Thermo Finnigan Surveyor System) on a C18 Elite 100 mm 2.1 mm column (Thermo Hypersil-Keystone) at a flow rate of 250 L/min with a linear gradient of 10% Solvent B (methanol containing 0.1% acetic acid) to 95% B over five minutes. Solvent A was water containing 0.1% acetic acid. Injection volumes of 10 L were used. MS Conditions: Mass spectrometer: Thermo Finnigan TSQ Quantum
Ionization mode: Electrospray (ESI), positive ion
Resolution: 0.7 Dalton Full Width Half Maximum on Q1 and Q3
Selected Reaction Monitoring:
Paroxetine 330.20 > 192.1 +/- 0.3 Da, 22 V collision energy
De-Fluoro-Paroxetine 312.20 > 174.1 +/- 0.3 Da, 20 V collision energy
Argon collision gas pressure 1.5 mTorr

Figure 2: Paroxetine

Figure 3: De-Fluoro Paroxetine

LC/MS/MS Chromatograms of Bovine Plasma Spiked with Paroxetine and Internal Standard Blank Plasma,Plasma containing Internal Standard,Calibration Standards at 0.1 and 0.5 pg/L Figure 4a: Blank Plasma
Figure 4b: Plasma containing Int Std


Figure 4c: Plasma spiked at 0.1 pg/L
Figure 4d: Plasma spiked at 0.5 pg/L
LC/MS/MS Chromatograms of Bovine Plasma Spiked with Paroxetine and Internal Standard Quality Control samples at 0.3 and 25 pg/L Figure 5a: Quality Control at 0.3 pg/L

Figure 5b: Quality Control at 25 pg/L Results The chromatography (Figure 4a4d and 5a5b) was capable of resolving Paroxetine and De-Fluoro-Paroxetine from any matrix peaks and was suitable for the quantitation of unknowns.

A calibration curve was constructed from the first and eighth replicate set of calibration standards (Figure 6). The linear fit calibration curve, weighted 1/X, of area ratio plotted against concentration was used to calculate the concentrations of all calibration standards and quality control samples throughout the analytical batch.

A correlation coefficient of r2=0.9994 with an equation of Y=0.00346891 +0.0312025*x was obtained for the curve.

The internal standard peak area response was 3.1% cv for 121 injections (Figure 7). Figure 6: Paroxetine/Int Std area ratio

Figure 7: Internal Standard area response The calibration curve was used to calculate the values of the eight replicates at each level of quality control sample (Table 1) and of the remaining six replicates of standards (Table 2). Table 1: Intra-batch accuracy and precision of quality control data for Paroxetine in Bovine plasma

Table 2: Intra-batch accuracy and precision of calibration data for Paroxetine in Bovine Plasma
*Calibration data sets 1 and 8 used to construct curve Discussion The lowest calibration standard selected for this assay at 0.1 pg/L had a minimum signal-to-noise of greater than 5:1 and variability of 3% from eight determinations. This was well within the acceptance criteria set by the guidelines which defines that the Limit of Quantitation (LOQ) of an assay should have less than 20% variability.

The Intra-batch variability of back-calculated concentrations of Paroxetine in Bovine Plasma was less than 3% across the whole calibration range. This falls well within the expected method validation acceptance criteria as set by Shah et al of 20% variability at the LOQ and less than 15% variability at all other concentrations of the calibration standards.

The validated method accuracy and precision were determined from the backcalculated values of the Quality Control samples. The intra-batch accuracy was 103.6%, 104.0% and 99.7% at concentrations of 0.3, 25 and 750 pg/L respectively. The intra-batch precision was 2.2%, 0.6% and 0.6% at concentrations of 0.3, 25 and 750 pg/L respectively.

The validation data indicate that the TSQ Quantum is sensitive, robust and linear (four orders of dynamic range demonstrated here) and is an ideal instrument for the quantitation of analytical substances in Biological matrices. References 1 Shah, V.P., et al, Pharmaceutical Research, Vol 9, No 4, 588-592, 1992
2 Shah, V.P., et al, Pharmaceutical Research, Vol 17, No 12, 1551-1557, 2000
3 Guidance for Industry, Bioanalytical Method Validation, U.S. Food and Drug Administration, Center for Drug Evaluation and Research (CDER), May 2001
'"/>

Source:


Page: All 1 2 3 4 5 6

Related biology technology :

1. A New PCR-based Mycoplasma Detection Method
2. DNA-free A NEW Method to Remove DNA
3. RNase Activity in Mouse Tissue: Classification, Hierarchy, and Methods for Control
4. Comparison of Different Methods of mRNA Quantitation
5. Rapid Method Development With the BioLogic DuoFlow Chromatography System for the Purification of His-Tagged Proteins
6. A Rapid Method for the Purification of Analytical Grade Proteins From Plants Using Preparative SDS-PAGE and Preparative IEF
7. Comparison of PCR Kleen Spin Columns to Traditional Methods for Purification of PCR Products Prior to Sequencing, Rev A
8. Commercial Plasmid Preparation Methods Yield DNA With Different Transfection Capabilities
9. Bio-Plex Cytokine Immunoassays and ELISA: Comparison of Two Methodologies in Testing Samples From Asthmatic and Healthy Children, Rev A
10. A Rapid Method for the Purification of Analytical Grade Proteins From Plants Using Preparative SDS-PAGE and Preparative IEF
11. A Robust LC/MS/MS Method for the Quantitation of Very Long Chain Fatty Acids for Research into the Characterization of Peroxisomal Disorders
Post Your Comments:
(Date:8/21/2014)... at the Institute for Molecular Medicine Finland (FIMM), ... developed a novel "man and machine" decision support ... aid was described in PLOS One ... based on computer vision algorithms similar to those ... of only the diagnostically most relevant areas. Tablet ...
(Date:8/21/2014)... , August 21, 2014 ... http://www.researchandmarkets.com/research/3zh3q8/global_hormone ) has announced the addition of ...  report to their offering.       ... as hormone therapy, is a means of ... produced by the body. This type of ...
(Date:8/21/2014)... 21, 2014 The College of New ... college in the Mid-Hudson Region to be designated as ... accepting applications from qualified “high-technology” businesses that align with ... pleased to have been selected for the START-UP NY ... initiative to stimulate economic development in New York by ...
(Date:8/21/2014)... , Aug. 21, 2014 OTC ... Pharmaceuticals Inc., a biotechnology company, on its approval ... previously traded on OTCQX®, the best marketplace for ... Markets Group. Logo - http://photos.prnewswire.com/prnh/20110118/MM31963LOGO ... the successful execution of its growth strategy and ...
Breaking Biology Technology:A novel 'man and machine' decision support system makes malaria diagnostics more effective 2Global Hormone Replacement Therapy Market 2014-2018: Key Vendors are Actavis, Bayer, Merck & Co, Novartis and Novo Nordisk 2The College of New Rochelle Becomes First Private College in Mid-Hudson Region to Earn START-UP NY Designation 2The College of New Rochelle Becomes First Private College in Mid-Hudson Region to Earn START-UP NY Designation 3OTC Markets Group Congratulates Immune Pharmaceuticals on NASDAQ Listing 2
... ... , ... (Vocus) February 4, 2010 -- Qteros, Inc. , a leading developer of second-generation ... announced today that Mick Sawka has joined the company as Vice President, Business Development, ...
... PARK, N.C. , Feb. 4 ... antiviral therapeutics, announced today that Gwendolyn Powell Painter , ...   , Dr. Painter has more than 20 years of ... clinical safety and pharmacovigilance, and occupational and environmental medicine.  In ...
... Mass. , Feb. 4 Interleukin Genetics, Inc. (NYSE Amex: ... at the 2010 BIO CEO & Investor Conference. The 12th Annual BIO ... the Waldorf-Astoria in New York . , Interleukin Genetics Chief Executive ... Company and discuss recent milestones at 1:00 p.m. ET in ...
Cached Biology Technology:Biofuels Technology Leader Qteros Names Mick Sawka Vice President, Business Development 2Biofuels Technology Leader Qteros Names Mick Sawka Vice President, Business Development 3Chimerix Names Wendy Painter, M.D., M.P.H., as Chief Medical Officer 2Chimerix Names Wendy Painter, M.D., M.P.H., as Chief Medical Officer 3Interleukin Genetics to Present at the 12th Annual BIO CEO Conference 2
(Date:8/22/2014)... (Edmonton) Thanks to important discoveries in basic and ... cancer has mobilized into a complex offensive spanning ... of Alberta chemistry lab could help find new ... developed a compound that targets a specific enzyme ... activity in cells from brain tumours. , Chemistry ...
(Date:8/22/2014)... is a multifunctional enzyme that localizes ... neuroprotective and cholesterol-synthesizing activities. DHCR24 overexpression ... amyloid β deposition. Dr. Xiuli Lu ... China constructed two recombinant adenoviruses (Ad-rSYN1-DHCR24-myc ... specifically in neuronal cells. They also ...
(Date:8/21/2014)... a significant reduction in smog-producing toxins in past ... A new study shows that while the Greater ... the toxins that contribute to smog, the city ... air pollution., Smog, which can cause or aggravate ... bronchitis, is produced by a set of complex ...
Breaking Biology News(10 mins):New enzyme targets for selective cancer therapies 2New enzyme targets for selective cancer therapies 3Despite a significant reduction in smog-producing toxins, the Greater Toronto Area still violates Canada's standards for ozone air pollution 2
... largest study of migraines, researchers have found 5 genetic regions ... onset of migraine. This study opens new doors to understanding ... The team identified 12 genetic regions associated with migraine susceptibility. ... known to play a role in controlling brain circuitries and ...
... study shows that the predator-prey relationship can affect the ... influence on the environment may offer a new way ... climate change. The study, conducted by researchers ... comes out this week in the Proceedings of ...
... - 250 million years ago, a mammal forerunner and ... Africa, Australia and France have discovered a world first ... burrow from the Karoo Basin of South Africa. ... and fossilized after being trapped by a flash flood ...
Cached Biology News:Getting to grips with migraine 2New study shows predators affect the carbon cycle 2Oddest couple ever found 2Oddest couple ever found 3Oddest couple ever found 4
... The ImageXpress 5000A automated cellular imaging and ... software system designed specifically for rapid acquisition ... throughput cell-based screening. ImageXpress is fast. ... resolution (1280 x 1024 pixels), with two ...
...
... Custom Peptide Services offer a full range ... research needs. Services include: Peptide synthesis, ... Peptide synthesis, purified product (70-95% pure) ... amidation, biotinylation, fluorescein, myristylation, phosphorylation, and rhodamine) ...
... CLS number is a new ... match Cornings product number. If ... order under the old Sigma-Aldrich ... service for assistance. Material: ...
Biology Products: