Benchmark Plus Microplate Reader Scan Well Feature

Theresa Redila-Flores and Li Tian, PhD, Bio-Rad Laboratories, 2000 Alfred Nobel Drive, Hercules, CA 94547 USA


The Benchmark Plus microplate reader is a new monochromator- based absorbance reader from Bio-Rad that offers superior convenience, accuracy, and flexibility. The Benchmark Plus is controlled by Microplate Manager software. A new feature has been designed in Microplate Manager software (version 5.1 and higher) that utilizes the monochromator design of Benchmark Plus and allows the user to prescan a well in order to determine the optimal wavelength prior to running a full plate of samples. In this note, a prescan procedure is described to show how to implement such a function into real-life applications.

In the following procedure, standard protein (bovine IgG) is added to cell culture medium without serum and the Scan Well function in Benchmark Plus is used to determine the best wavelength for protein assay with cell culture medium. Bio-Rads protein assay kit (catalog # 500-0001) is used to check for measurement inconsistencies in culture medium.


Procedure
1. Serially dilute (1:2) standard bovine IgG protein ranging from 62.5 g/ml to 1 mg/ml in F-12 Hams medium without serum.

2. Aliquot 10 l (in triplicate) of each diluted standard protein into a 96-well clear bottom black microplate; i.e., A1, B1, and C1 for 1 mg/ml protein sample. Aliquot 10 l (in triplicate) of F-12 Hams medium into 3 wells (A7, B7, C7) to be used as blanks.

3. Add 200 l of diluted dye reagent (from Bio-Rad protein assay kit) to each well. Mix thoroughly using a microplate mixer for about 5 sec. Make sure that no bubbles are present. Bubbles may cause incons istencies in absorbance reading.

4. Incubate 5 min at room temperature.

5. Place microplate into Benchmark Plus microplate reader.

6. Open Microplate Manager. Under File, select Endpoint protocol.

7. Select 96-well format. Under Advanced Options, open Scan Well.

8. Scan well A1 at 340800 nm using a 5 nm step size. Save data.

9. Scan remaining wells (A2A7) and save data.

10. For bovine IgG at optimal wavelength, select Endpoint protocol, 96-well format, and single wavelength reading at 590 nm.

11. Highlight the appropriate wells to be read; i.e., A1 to C7 for the concentration range from 62.5 g/ml to 1 mg/ml and the blank. 12. Read absorbance. Save and analyze data.


Results
The most concentrated protein sample shows a higher peak at 590 nm than the least concentrated protein samples (Figure 1). The least concentrated samples and the blank showed higher peaks at 470 nm, whereas the more concentrated protein samples showed similar peak heights at around 470 and 590 nm. From the figure, one can conclude that whereas a peak at 470 nm is apparently caused by the cell culture medium used here, the optimal wavelength for bovine IgG is at 590 nm.

Further reading of bovine IgG concentration at 590 nm on Benchmark Plus yields a linear graph with a correlation coefficient of 0.9993 (Figure 2).


Conclusion
The Benchmark Plus Scan Well feature is used to ascertain the best wavelength for reading a desired assay. It allows scanning of a well from 340800 nm at various wavelength increments. A graph and data of the scan area are provided and can be exported into Microsoft Excel software for furt her analysis. The Scan Well feature minimizes the problem of finding out at what wavelength samples should be read, thus saving time and effort. At the optimal wavelength for the bovine IgG tested, the correlation coefficient is ≥ 0.99 and linear over the dilution range of 62.5 g/ml to 1 mg/ml. The dilution in F-12 Hams medium without serum did not affect the measurement of the protein. The results show that the Benchmark Plus microplate reader is an excellent instrument to use for absorbance studies, as well as an instrument to save time in determining the optimal wavelength to read a sample.


Bibliography
Bradford MM, A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding, Anal Biochem 72, 248254 (1976)

Lowry OH et al., Protein measurement with the Folin phenol reagent. J Biol Chem 193, 265275 (1951)


back to top
'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. Cell Proliferation Assay Using the Benchmark Plus Microplate Spectrophotometer
2. Benchmark Plus Microplate Reader: Quantitation of Protein Concentration Using Two Different Colorimetric Assay Kits, Rev A
3. Cerenkov Counting Performance on the TopCountTM Microplate Scintillation and Luminescence Counter
4. Counting Non-Aqueous Samples in the TopCount Microplate Scintillation and Luminescence Counter
5. Fluorometric Protease Assays in the SpectraMax Gemini Microplate Spectrofluorometer: Example Using Caspase3 (MaxLine Application Note #35)
6. LIVE/DEAD Viability/Cytotoxicity Assay for Animal Cells Using the SpectraMax Gemini XS Fluorescence Microplate Reader (MaxLine Application Note #43)
7. Using the OligoGreen Oligonucleotide Quantitation Reagent in the Gemini XS, Gemini EM and SpectraMax M2 Microplate Readers (MaxLine Application Note #21)
8. PathCheck Applied to Measurement of Protein Solutions in the SpectraMax Plus Microplate Spectrophotometer (MaxLine Application Note #26)
9. Microdetermination of Phosphorus Using the SpectraMax Plus Microplate Spectrophotometer: Choice of Microplate, Cuvette, or Test Tube Assay Formats (MaxLine Application Note #24)
10. UV Absorbance Measurements in SpectraMax Microplate Spectrophotometers MaxLine Application Note #32)
11. Ca++-Induced G Protein Coupled Receptor (GPCR) Activation Monitored via Aequorin Luminescence in the LMax Microplate Luminometer (MaxLine Application Note #42)
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:2/11/2016)... 11, 2016  Wellcentive today announced it has ... Oregon -based community care organization (CCO) with ... analytics, quality reporting and care management solutions and ... team of quality managers, analysts and care managers ... groups serving FamilyCare members. ...
(Date:2/11/2016)... 2016  Dovetail Genomics™ LLC today announced that it ... a planned metagenomic genome assembly service. Richard Green ... assembly method in a talk on Friday, February 12 ... conference in Orlando, Fla. ... is difficult. Using its proprietary Chicago ...
(Date:2/11/2016)... 2016   BioInformant announces the February 2016 ... Products, Opportunities, Tools, and Technologies – Market Size, Segments, ... The first and only ... industry, BioInformant has more than a decade of historical ... by stem cell type. This powerful 175 page global ...
(Date:2/11/2016)... ... February 11, 2016 , ... Global Stem Cells Group, ... Bankok,Thailand-based Global Stem Cells Network (GSCN) to distribute exosome injection and other biological ... Mexico, Costa Rica, Dominican Republic, Colombia, Argentina, Nicaragua, Panama, El Salvador, Venezuela, Peru, ...
Breaking Biology Technology:
(Date:2/2/2016)... NEW YORK , Feb. 2, 2016 /PRNewswire/ ... Potentials of that Rising Market Are you ... new analysis forecasts revenues for checkpoint inhibitors. Visiongain,s ... world market, submarket, product and national level. ... Instead discover what progress, opportunities and revenues those ...
(Date:2/1/2016)... 1, 2016 Rising sales of ... global touchfree intuitive gesture control market size ... sales of consumer electronics coupled with new technological advancements ... size through 2020   --> ... new technological advancements to drive global touchfree intuitive gesture ...
(Date:1/27/2016)... WEST CHESTER, Ohio , Jan. 27, 2016 /PRNewswire/ ... services supplier based in West Chester, Ohio ... and their award winning service staff, based in ... Track,s technical capacity and ability to provide modifications, installations ... John Dovalina , CEO of PLUS, commented, "PLUS ...
Breaking Biology News(10 mins):