Navigation Links
Benchmark Plus Microplate Reader Scan Well Feature

Theresa Redila-Flores and Li Tian, PhD, Bio-Rad Laboratories, 2000 Alfred Nobel Drive, Hercules, CA 94547 USA


The Benchmark Plus microplate reader is a new monochromator- based absorbance reader from Bio-Rad that offers superior convenience, accuracy, and flexibility. The Benchmark Plus is controlled by Microplate Manager software. A new feature has been designed in Microplate Manager software (version 5.1 and higher) that utilizes the monochromator design of Benchmark Plus and allows the user to prescan a well in order to determine the optimal wavelength prior to running a full plate of samples. In this note, a prescan procedure is described to show how to implement such a function into real-life applications.

In the following procedure, standard protein (bovine IgG) is added to cell culture medium without serum and the Scan Well function in Benchmark Plus is used to determine the best wavelength for protein assay with cell culture medium. Bio-Rads protein assay kit (catalog # 500-0001) is used to check for measurement inconsistencies in culture medium.


Procedure
1. Serially dilute (1:2) standard bovine IgG protein ranging from 62.5 g/ml to 1 mg/ml in F-12 Hams medium without serum.

2. Aliquot 10 l (in triplicate) of each diluted standard protein into a 96-well clear bottom black microplate; i.e., A1, B1, and C1 for 1 mg/ml protein sample. Aliquot 10 l (in triplicate) of F-12 Hams medium into 3 wells (A7, B7, C7) to be used as blanks.

3. Add 200 l of diluted dye reagent (from Bio-Rad protein assay kit) to each well. Mix thoroughly using a microplate mixer for about 5 sec. Make sure that no bubbles are present. Bubbles may cause incons istencies in absorbance reading.

4. Incubate 5 min at room temperature.

5. Place microplate into Benchmark Plus microplate reader.

6. Open Microplate Manager. Under File, select Endpoint protocol.

7. Select 96-well format. Under Advanced Options, open Scan Well.

8. Scan well A1 at 340800 nm using a 5 nm step size. Save data.

9. Scan remaining wells (A2A7) and save data.

10. For bovine IgG at optimal wavelength, select Endpoint protocol, 96-well format, and single wavelength reading at 590 nm.

11. Highlight the appropriate wells to be read; i.e., A1 to C7 for the concentration range from 62.5 g/ml to 1 mg/ml and the blank. 12. Read absorbance. Save and analyze data.


Results
The most concentrated protein sample shows a higher peak at 590 nm than the least concentrated protein samples (Figure 1). The least concentrated samples and the blank showed higher peaks at 470 nm, whereas the more concentrated protein samples showed similar peak heights at around 470 and 590 nm. From the figure, one can conclude that whereas a peak at 470 nm is apparently caused by the cell culture medium used here, the optimal wavelength for bovine IgG is at 590 nm.

Further reading of bovine IgG concentration at 590 nm on Benchmark Plus yields a linear graph with a correlation coefficient of 0.9993 (Figure 2).


Conclusion
The Benchmark Plus Scan Well feature is used to ascertain the best wavelength for reading a desired assay. It allows scanning of a well from 340800 nm at various wavelength increments. A graph and data of the scan area are provided and can be exported into Microsoft Excel software for furt her analysis. The Scan Well feature minimizes the problem of finding out at what wavelength samples should be read, thus saving time and effort. At the optimal wavelength for the bovine IgG tested, the correlation coefficient is ≥ 0.99 and linear over the dilution range of 62.5 g/ml to 1 mg/ml. The dilution in F-12 Hams medium without serum did not affect the measurement of the protein. The results show that the Benchmark Plus microplate reader is an excellent instrument to use for absorbance studies, as well as an instrument to save time in determining the optimal wavelength to read a sample.


Bibliography
Bradford MM, A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding, Anal Biochem 72, 248254 (1976)

Lowry OH et al., Protein measurement with the Folin phenol reagent. J Biol Chem 193, 265275 (1951)


back to top
'"/>

Source:


Page: All 1 2 3

Related biology technology :

1. Cell Proliferation Assay Using the Benchmark Plus Microplate Spectrophotometer
2. Benchmark Plus Microplate Reader: Quantitation of Protein Concentration Using Two Different Colorimetric Assay Kits, Rev A
3. Cerenkov Counting Performance on the TopCountTM Microplate Scintillation and Luminescence Counter
4. Counting Non-Aqueous Samples in the TopCount Microplate Scintillation and Luminescence Counter
5. Fluorometric Protease Assays in the SpectraMax Gemini Microplate Spectrofluorometer: Example Using Caspase3 (MaxLine Application Note #35)
6. LIVE/DEAD Viability/Cytotoxicity Assay for Animal Cells Using the SpectraMax Gemini XS Fluorescence Microplate Reader (MaxLine Application Note #43)
7. Using the OligoGreen Oligonucleotide Quantitation Reagent in the Gemini XS, Gemini EM and SpectraMax M2 Microplate Readers (MaxLine Application Note #21)
8. PathCheck Applied to Measurement of Protein Solutions in the SpectraMax Plus Microplate Spectrophotometer (MaxLine Application Note #26)
9. Microdetermination of Phosphorus Using the SpectraMax Plus Microplate Spectrophotometer: Choice of Microplate, Cuvette, or Test Tube Assay Formats (MaxLine Application Note #24)
10. UV Absorbance Measurements in SpectraMax Microplate Spectrophotometers MaxLine Application Note #32)
11. Ca++-Induced G Protein Coupled Receptor (GPCR) Activation Monitored via Aequorin Luminescence in the LMax Microplate Luminometer (MaxLine Application Note #42)
Post Your Comments:
*Name:
*Comment:
*Email:


(Date:2/11/2016)... Feb. 11, 2016  Neurocrine Biosciences, Inc. (NASDAQ: NBIX ) ... December 31, 2015. --> --> ... net loss of $29.3 million, or $0.34 loss per share, compared ... share for the same period in 2014. For the year ended ... million, or $1.05 loss per share, as compared to a net ...
(Date:2/11/2016)... and GERMANTOWN, Maryland , ... Frankfurt Prime Standard: QIA) today announced the introduction of ... gene expression profiling, expanding QIAGEN,s portfolio of Sample to ... researchers to select from over 20,000 human genes and ... between genes, cellular phenotypes and disease processes. --> ...
(Date:2/11/2016)... ATLANTA , Feb. 11, 2016  Spectra BioPharma ... Organization (CSO) that provides biopharma companies the experience, ... implement and deploy outsourced sales teams. Created in ... team addresses both the strategic and tactical needs ... innovative sales solutions through both personal and non-personal ...
(Date:2/11/2016)... (PRWEB) , ... February 11, ... ... business-to-business publication dedicated to delivering cutting-edge information focused on the development and ... Sciences to become a premier sponsor of the 2016 BioProcess International Awards ...
Breaking Biology Technology:
(Date:2/11/2016)... 2016  Vigilant Solutions announces today that its license plate ... by Lee,s Summit Police Department to ... arrest of a homicide suspect. Kansas City ... 65 square miles and is home to roughly 100,000 residents. ... single mobile license plate reader system and also leverages Vigilant,s ...
(Date:2/8/2016)... PRAGUE, Czech Republic , February 8, 2016 ... first EU-regulated global payment platform which presents innovation ... Voice Biometrics Authentication feature called VoiceKey. --> ... payment platform which presents innovation for clients, comfort ... feature called VoiceKey. --> Worldcore ...
(Date:2/3/2016)... , February 3, 2016 ... market research report "Automated Fingerprint Identification System Market by ... Search), Application (Banking & Finance, Government, Healthcare, and Transportation) ... MarketsandMarkets, the market is expected to be worth USD ... 21.0% between 2015 and 2020. The transformation and technology ...
Breaking Biology News(10 mins):