Benchmark Plus Microplate Reader: Quantitation of Protein Concentration Using Two Different Colorimetric Assay Kits, Rev A

Theresa Redila-Flores, Bio-Rad Laboratories, 2000 Alfred Nobel Drive, Hercules, CA 94547 USA


The Bradford and Lowry assays are 2 colorimetric assays that are commonly used to quantitate protein concentration. Bio-Rads protein assay (catalog #500-0001), based on the Bradford dye-binding assay (Bradford 1976), measures total protein concentration. The standard assay is used for protein concentrations ranging from 50 g/ml to 1.0 mg/ml. This assay is based on the change of color of Coomassie Brilliant Blue G-250 dye with different protein concentrations, in which the dye binds to basic and aromatic amino acid residues. This assay is measured between 465 and 595 nm. Bio-Rads detergent compatible (DC) protein assay (catalog #500-0111) is similar to the Lowry assay (Lowry et al. 1951), but only requires a 15 min incubation. The reaction between protein and copper in the alkaline solution and the Folin reagent reduced by the copper-treated protein leads to color development. This protein assay is measured between 650 and 750 nm, with protein concentrations ranging from 200 g/ml to 1.0 mg/ml.


Procedure
Bradford Method (Bio-Rad Protein Assay) Using Clear 96- and 384-Well Plates
1. Dilute dye reagent with 1 part of dye reagent concentrate and 4 parts of distilled water. Filter through a Whatman #1 filter. The diluted solution is good for about 2 weeks when stored at room temperature.

2. Serially dilute standard bovine IgG protein 1:2, ranging from 31 g/ml to 1.0 mg/ml.

3. Aliquot 10 l of each diluted standard in triplicate into a clear 96-well microplate.

4. Add 200 l of diluted dye reagent. Mix thoroughly using a microplate mixer.

5. Incubate 5 min at room temperature.

6. Measure absorbance at 595 nm.

7. For a 384-well microplate, the protein standard and diluted dye reagent volumes are reduced to one quarter of the total volume used in a 96-well plate; that is, 50 l per well.

8. For 384-well measurements, aliquot 10 l of each diluted standard into a tube.

9. Add 200 l of diluted dye reagent. Vortex for about 5 sec.

10. Let incubate for 5 min at room temperature.

11. Aliquot 52.5 l per well. Make sure no bubbles are present, as this may cause reading errors.

12. Measure absorbance at 595 nm.


Lowry Method (Bio-Rad Detergent Compatible (DC) Protein Assay) using 96- and 384-Well Plates
1. Prepare working reagent as follows: Add 20 l of reagent S to 1 ml of reagent A as needed for assay.

2. Serially dilute the standard bovine IgG protein 1:2, from 125 g/ml to 2.0 mg/ml.

3. Aliquot 30 l of each of the diluted standards into a separate 2 ml tube.

4. Add 150 l of the working reagent to each tube and then add 1.2 ml of reagent B.

5. Vortex for about 5 sec.

6. Incubate at room temperature for 15 min.

7. Aliquot 230 l of each sample in triplicate into a clear 96-well microplate.

8. Measure absorbance at 750 nm.

9. For a 384-well microplate, aliquot 6.25 l of diluted standard into a 1.5 ml microfuge tube.

10. Add 31.25 l of working reagent and 250 l of reagent B.

11. Vortex for about 5 sec.

12. Incubate at room temperature for 15 min.

13. Aliquot 57.5 l of each sample in triplicate into a clear 384-well microplate.

14. Measure at 750 nm.


Results
Bio-Rad's protein assay kit gave linear results using the Benchmark Plus microplate reader to quantitate protein
concentration. The linear range for Bio-Rads protein assay was between 50 g/ml and 1.0 mg/ml in both 96- and 384-well plates (Figures 1 and 2). Readings were linear within their designated ranges with a correlation coefficient of ≥0.99.

Bio-Rad's detergent compatible (DC) protein assay kit gave linear results using the Benchmark Plus microplate reader to quantitate protein concentration. The linear range for both plates was between 125 g/ml and 2.0 mg/ml in both 96- and 384-well plates (Figures 3 and 4). Readings were linear within their designated ranges with a correlation coefficient of ≥0.99.


Conclusion
The Benchmark Plus microplate reader is an excellent instrument to quantitate concentrations of protein samples
using 2 of Bio-Rads protein assay kits. The correlation coefficients of all the graphs were ≥0.99, within the specified detection ranges of 31.25 g/ml to 1.0 mg/ml. Assays of both
96- and 384-well plate formats gave linear graphs, thus allowing a choice of using these 2 different plate formats with excellent results along with high throughput. The results
showed that the Benchmark Plus microplate reader exhibits excellent linearity, repeatability, and accuracy for use in 96- and 384-well plate formats.


References
Bradford MM, A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding, Anal Biochem 72, 248254 (1976)

Lowry OH et al., Protein measurement with the Folin phenol reagent, J Biol Chem 193, 265275 (1951)


Coomassie is a trademark of Imperial Chemical Industries PLC.


back to top
'"/>

Source:


Page: All 1 2 3 4

Related biology technology :

1. Benchmark Plus Microplate Reader Scan Well Feature
2. Cell Proliferation Assay Using the Benchmark Plus Microplate Spectrophotometer
3. Cerenkov Counting Performance on the TopCountTM Microplate Scintillation and Luminescence Counter
4. Counting Non-Aqueous Samples in the TopCount Microplate Scintillation and Luminescence Counter
5. Fluorometric Protease Assays in the SpectraMax Gemini Microplate Spectrofluorometer: Example Using Caspase3 (MaxLine Application Note #35)
6. LIVE/DEAD Viability/Cytotoxicity Assay for Animal Cells Using the SpectraMax Gemini XS Fluorescence Microplate Reader (MaxLine Application Note #43)
7. Using the OligoGreen Oligonucleotide Quantitation Reagent in the Gemini XS, Gemini EM and SpectraMax M2 Microplate Readers (MaxLine Application Note #21)
8. PathCheck Applied to Measurement of Protein Solutions in the SpectraMax Plus Microplate Spectrophotometer (MaxLine Application Note #26)
9. Microdetermination of Phosphorus Using the SpectraMax Plus Microplate Spectrophotometer: Choice of Microplate, Cuvette, or Test Tube Assay Formats (MaxLine Application Note #24)
10. UV Absorbance Measurements in SpectraMax Microplate Spectrophotometers MaxLine Application Note #32)
11. Ca++-Induced G Protein Coupled Receptor (GPCR) Activation Monitored via Aequorin Luminescence in the LMax Microplate Luminometer (MaxLine Application Note #42)

Post Your Comments:
(Date:11/25/2009)...Nov. 25 /PRNewswire/ - Cytochroma today announced ...f Directors. Dr. Moldt will replace Mr. Ulrik Spor...Board. The number of Board members remains unchang...elcome Peter Moldt to the board and look forward t...ytochroma continues to make progress in becoming a...
(Date:11/25/2009)...K, Nov. 25 Reportlinker.com announ... in its catalogue: ,, Brazil Bioethanol Market....reportlinker.com/p0165503/Brazil-Bioethanol-Marke...l Bioethanol Market Analysis and Forecasts to 2013...d analysis and forecasts of the global bioethanol ...
(Date:11/24/2009)...ble in German . , Whether it,s CD packaging...adio Frequency Identification) are increasingly fi...ossible to label objects or goods and identify the...e scanner can read and process the data contained ... under production conditions of up to 100 degrees ...
(Date:11/24/2009)... Ambassador to also Visit Detroit , ...David Appia, French Ambassador for International I...rance Agency, , , , ,What: Ambassador Appia is av...usiness initiatives adopted by France to further d...e for innovation and development., , ,When: Availa...
Breaking Biology Technology:Cytochroma Announces Appointment of Peter Moldt to Board of Directors 2Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 2Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 3Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 4Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 5Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 6Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 7Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 8Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 9Reportlinker Adds Brazil Bioethanol Market Analysis and Forecasts to 2013 10Intelligence inside metal components 2
...up, a life science-focused strategic communication...d as public relations agency of record by therapeu...nces, Inc. , , Donna L. LaVoie, President & ...to represent Profectus during an important stage i...ward to working with them to communicate their mes...
... work could help farmers protect against billions ... St. Louis, MO (PRWEB) April 29, 2009 -- In the pa..., has received more than $1.2 million in research ...om the National Science Foundation (NSF) to furthe...nematode control in crops such as corn. The Phase...
..., Inc. (Nasdaq: TELK ) announced results of stud...ere presented at the 100th annual meeting of the A...Denver, CO. The abstracts can be found on the Tel...haracterization and Mechanism of Action of Orally ... 1764 , Aurora kinases are Ser/Thr kinases that p...
Other Biology Technology:LaVoie Group Selected by Vaccine Developer, Profectus BioSciences, as Agency of Record 2Divergence Receives $1.2 Million in Grants, Including NSF Grant to Fund Continuing Nematode Research in Corn 2Divergence Receives $1.2 Million in Grants, Including NSF Grant to Fund Continuing Nematode Research in Corn 3Telik Reports Preclinical Results at AACR Annual Meeting 2Telik Reports Preclinical Results at AACR Annual Meeting 3
(Date:11/24/2009)... , Rocket science is opening new doors to underst... affect the hearing of a marine mammal or if they...rial sized X-ray scanners that NASA uses to detect...rockets is now allowing scientists to peek inside ... detailed three-dimensional replicas of a whale,s ...
(Date:11/24/2009)... , Springer and the American College of Medical T...borate on the publication of the Journal of Medic...ACMT, beginning in March 2010. It was previously ... , The Journal of Medical Toxicology , an inte...advancing the science and practice of medical toxi...
(Date:11/24/2009)... , Cambridge, Md. (November 24, 2009) New resear...rctic river ecosystems reveals predictable tempora...hese communities as markers for monitoring climate... this week in the Proceedings of the National Aca...ial communities in the six rivers shifted synchron...
Breaking Biology News(10 mins):Rocket science leads to new whale discovery 2Springer and the American College of Medical Toxicology to work together 2Researchers establish common seasonal pattern among bacterial communities in Arctic rivers 2New Survey Identifies Specialized Training and Respect for Culture as Key Factors in Healthcare Security Personnel Selection 54068 1New Survey Identifies Specialized Training and Respect for Culture as Key Factors in Healthcare Security Personnel Selection 54068 2Enfina Healthcare Solutions Implements Business Process Outsourcing Services Software from DST Health Solutions 54066 1Enfina Healthcare Solutions Implements Business Process Outsourcing Services Software from DST Health Solutions 54066 2Latest Holiday Migraine Survey Results Published 54065 1Latest Holiday Migraine Survey Results Published 54065 2Latest Holiday Migraine Survey Results Published 54065 3
...ug abusers are increasingly turning to a slow-rele...dangerous high, University of Florida researchers ...people dying from an overdose of the drug fentanyl...ises. , Addicts are misusing a clear patch that tr...skin into the bloodstream over the course of a few...
...new website with a Global Information System will ...mental hazards caused by Hurricane Katrina. The Na...(NIEHS), one of the National Institutes of Health,...e data to public health and safety workers on cont...try maps, as well as demographic information for l...
...new website with a Global Information System will ...mental hazards caused by Hurricane Katrina. The Na...(NIEHS), one of the National Institutes of Health,...e data to public health and safety workers on cont...try maps, as well as demographic information for l...
Other Biology News:Prescription pain patch abuse blamed for increase in deaths 2Prescription pain patch abuse blamed for increase in deaths 3FDA approves child-friendly AIDS medicine 2NIEHS launches website with information for assessing environmental hazards from Hurricane Katrina 2
... was tested on rat tissue extracts (liver, kidney,...) extract. It was also tested with microsomal frac...plication: A simple and reliable method for measur...d on the detection of the inorganic phosphate rele...
...ve stress is the condition in which a lack of bala...ious antioxidants. Oxidative stress appears to be... including artherosclerosis, chronic inflammatory ...s several antioxidant systems that are very import...
FlowJo from Tree Star Inc
...ytic degradation is critical to the maintenance of...proteins as important and diverse as those involve...sponse, antigen presentation, modulation of cell s...lation, transcription, and signalling factors. The...
Biology Products: