Analysis of urine and seawater samples by ultrasonic nebulization with a high resolution ICP spectrometer

Tran T. Nham Varian Australia Pty Ltd Mulgrave, Victoria, Australia

Introduction

Because of the low detection limit capability of the ultrasonic nebulizer when combined with an ICP spectrometer, it has been widely used to determine trace and ultratrace levels in water samples1. The purpose of this work was to evaluate the ability of the ultrasonic nebulizer to handle samples with significant amounts of dissolved solids, such as urine and seawater.

Experimental

Instrumentation

All measurements were carried out on Varian Liberty 200 ICP spectrometer. The ultrasonic nebulizer was a Cetac U-5000 AT.

The operating conditions of the instrument system are listed in table 1.

For optimum operation, the use of a higher auxiliary Ar flow of 2.25 L/min lifts the plasma away from the injector tube to minimize the possibility of salt build up at the tip of the injector tube. The use of a 1.8 mm I.D. injector tube can also minimize salt build up at the injector tube, especially while running seawater samples.

Lowering the torch by 1 mm (compared with the normal position) can also minimize salting up at the injector tip. The distance from the injector tip to the top of the intermediate tube is normally set to 3 mm with a demountable torch.

Reagents

All chemicals used were analytical grade. The multielement stock solution was made up from single element standard stock solutions (Spectrosol, BDH Chemicals). The multi-element working standard was prepared by dilution of the multi-element stock solution and contained 0.14 M HNO3.

Matrix matched standards and blank were employed in this work. As an alternative, the method of internal standardization could also be used, to compensate for differences in transport efficiences between different sample types.

For Seronorm urine analysis, the standard and blank were prepared in synthetic urine solution. Synthetic urine solution2 contains 14.1 g NaCl, 2.8 g KCl, 17.3 g urea, 1.9 mL of 25% v/v ammonia solution, 0.6 g CaCl2 and 0.43 g MgSO4, made up to 1 L with 0.02 M HCl solution.

For seawater analysis, the standard and blank were prepared in artificial seawater solution. Artificial seawater solution3 contains 25.4 g NaCl, 5.083 g MgCl2, 1.104 g CaCl2, 0.722 g KCl, 0.026 g H3BO3 and 0.203 g SrCl2, made up to 1 L with distilled de-ionized water.

Results

1. Long term stability

Long term stability was evaluated by performing a calibration on a multi-element standard and a blank, both being made up in 1% w/v NaCl solution. The standard was then measured as a sample continuously for a period of an hour. After an hour, the reproducibility ranged between 0.8 to 2 %RSD for the measurements plotted in figure 1.

An additional stability test was performed on a seawater solution, as shown in figure 2. The reproducibility over an hour ranged from 2 to 5 %RSD. It was noted that after continuously aspirating the seawater solution for over an hour, the injector tube gradually showed signs of blockage due to salt deposition. Further experimentation confirmed that rinsing between samples is desirable to minimize any potential clogging of the tube and improve the long term stability of the analysis.

2. Urine analysis

Urine samples can be analyzed directly by ultrasonic nebulization. Analysis was performed on a Seronorm urine sample. The standard and blank were prepared in a synthetic urine solution. The results of the analysis and the calculated detection limits (2 sigma) are listed in table 2. The measured values are in good agreement with the certified values. The precision of the measurements ranged from 0.5 to 3 %RSD.

Wavelength scans of standard, blank and urine sample on Cd and Cr are illustrated in figure 3.

3. Seawater analysis

This study has shown that seawater samples can be analyzed directly following ultrasonic nebulization. To demonstrate the accuracy of the method, a known amount of analytes were added to the sample and the recovery calculated.

The dissolved solids content of seawater can be up to 3.5% w/v. Because of the relatively high salt content in seawater, it is recommended that longer washout times are required between samples to clean out any possible salt deposition in the injector tube and improve long term stability. The use of an auxiliary rinse, which delivers rinsing agent directly onto the transducer surface, can speed up the washout process and reduce the analysis time.

The results of the seawater analysis, the percentage spike recovery and the calculated detection limit (2 sigma) are given in table 3. The percentage recoveries ranged from 90103%. The precision of the measurements ranged from 0.5 to 3 %RSD.

Wavelength scans of 50 μg/L of Pb and Ni in seawater sample are shown in figure 4.

Summary

The Liberty ICP and ultrasonic nebulizer combination can be used for the direct determination of trace elements in samples such as urine and seawater. The measured values are in good agreement with the certified values and the percentage recoveries ranged from 90 to 103%.

References

1. ICP At Work No. 8, Nham T. T., Varian, Mulgrave, Victoria, 1992.

2. Determination of silicon, calcium, magnesium and phosphorus in urine using inductively coupled plasma emission spectrometry and a matrix matching technique, Tanaka T., Hayashi Y., Clin. Chim. Acta, 1986, 156, 109-114.

3. Water analysis by Atomic Absorption, C. R. Parker, Varian Techtron, Springvale, Victoria, 1972.


'"/>

Source:


Page: All 1 2 3 4

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
(Date:12/24/2014)... The report provides basic ... definition, classification, application and industry overview. This ... and product cost structure. Production is separated ... also covers upstream raw materials, equipment, downstream ... and proposals. In the end, the export ...
(Date:12/24/2014)... BEIJING , Dec. 23, 2014 ... or the "Company"), a leading fully integrated plasma-based ... today announced that its majority-owned subsidiary, Shandong Taibang ... Practices ("GMP") certification from the China Food and ... factor production facility. As previously disclosed in the ...
(Date:12/24/2014)... Fla. , Dec. 23, 2014  Rock Creek ... that it has filed a Clinical Trial Application (CTA) ... Healthcare products Regulatory Agency (MHRA) seeking regulatory approval to ... lead molecule, Anatabine Citrate. Contingent on the ... conduct a Phase I trial to assess the safety, ...
(Date:12/22/2014)... Dec. 22, 2014 /PRNewswire/ - RepliCel Life Sciences Inc. ... company focused on the development of autologous cell therapies, ... the University of Calgary in conjunction with co-authors from ... further validates the company,s ongoing clinical research using dermal ... pattern baldness. The paper entitled " Hair ...
Breaking Biology Technology:Worldwide Methyl Mercaptan Market 2015-2020 Forecasts on Development & Trends Now Available at ReportsnReports.com 2China Biologic Receives GMP Certification for New Coagulation Factor Facility 2China Biologic Receives GMP Certification for New Coagulation Factor Facility 3Rock Creek Pharmaceuticals Files Clinical Trial Application in the UK for Clinical Trials of Anatabine Citrate Formulations 2Rock Creek Pharmaceuticals Files Clinical Trial Application in the UK for Clinical Trials of Anatabine Citrate Formulations 3Newly Released Peer-Reviewed Publication Further Validates RepliCel's Use of Dermal Sheath Cup Cells to Treat Pattern Baldness 2Newly Released Peer-Reviewed Publication Further Validates RepliCel's Use of Dermal Sheath Cup Cells to Treat Pattern Baldness 3Newly Released Peer-Reviewed Publication Further Validates RepliCel's Use of Dermal Sheath Cup Cells to Treat Pattern Baldness 4Newly Released Peer-Reviewed Publication Further Validates RepliCel's Use of Dermal Sheath Cup Cells to Treat Pattern Baldness 5
... 2012  Ariosa Diagnostics, a molecular diagnostics company, today ... will be available in the United States and ... and Integrated Genetics, a member of LabCorp,s Specialty ... for detection of common fetal trisomies utilizing a ...
... and manipulating nanoparticles, researchers at the National Institute of ... useful life more than tenfold.* This new approach, which ... experimenters the trapping time they need to build nanoscale ... nanoparticles inside biological cells without damaging the cells with ...
... China, May 4, 2012 /PRNewswire-Asia-FirstCall/ -- China Pharma Holdings, Inc. ... manufactures, and markets specialty pharmaceutical products in China, today announced ... 2012 financial results before the market opens on Tuesday, May ... conference call at 8:30 am ET on May 15, 2012 ...
Cached Biology Technology:Ariosa Diagnostics Announces Nationwide Launch of the Harmony™ Prenatal Test Through LabCorp 2Light touch keeps a grip on delicate nanoparticles 2China Pharma Holdings, Inc. to Report First Quarter of Fiscal Year 2012 Financial Results on May 15, 2012 2
(Date:12/17/2014)... , Dec. 15, 2014 The Defense ... it easier to detect and prevent counterfeit microcircuits ... Today the agency started performing an in-house microcircuit ... validate the authenticity of purchased microcircuits while increasing ... quality control measures will be conducted at DLA,s ...
(Date:12/15/2014)... 2014 Research and Markets ( ... "Global Facial Recognition Market 2015-2019" report ... http://photos.prnewswire.com/prnh/20130307/600769 Facial recognition is a ... recognition system measures the overall facial feature of ... and the distance between eyes. Facial recognition is ...
(Date:12/10/2014)... , Dec. 8, 2014 You,ve been here ... account but can,t remember your password, site key or the ... was your first grade teacher? Today, ... the app that will finally put an end to ... 1U TM . 1U leverages a user,s smartphone ...
Breaking Biology News(10 mins):Defense Logistics Agency Launches DNA Marking Capability to Strengthen Microcircuit Supply Chain 2Defense Logistics Agency Launches DNA Marking Capability to Strengthen Microcircuit Supply Chain 3Global Facial Recognition Market 2015-2019: Key Vendors are 3M Cogent, Cognitec Systems, NEC and Safran Group 2Global Facial Recognition Market 2015-2019: Key Vendors are 3M Cogent, Cognitec Systems, NEC and Safran Group 3The Password is Finally Dead: Launch of 1U Mobile App Eliminates Need for All Usernames and Passwords 2The Password is Finally Dead: Launch of 1U Mobile App Eliminates Need for All Usernames and Passwords 3
... spinal cord injury is receiving attention in the research ... department of biomedical engineering at NJIT, recently has published ... stimulators for individuals with spinal cord injuries. ... of support from a four-year, $1.4 million National Institutes ...
... Inc. (Nasdaq: AWRE ), a global provider ... of the Next-Generation Universal Automated Booking Station, or "Universal ... of Justice. Universal ABS is a biometric enrollment application ... modular, COTS software framework that is far more easily ...
... S.C., Oct. 17, 2011 Integrated Biometrics, a South ... announced today that they have launched a new FBI ... device has been tested by The MITRE Corporation and ... Certified Product List. Watson is the world,s first non-optical ...
Cached Biology News:NJIT researcher testing micro-electronic stimulators for spinal cord injuries 2Aware Develops Next-Generation Biometric Enrollment Application for Department of Justice 2Aware Develops Next-Generation Biometric Enrollment Application for Department of Justice 3Aware Develops Next-Generation Biometric Enrollment Application for Department of Justice 4The World's First Non Optical FBI Appendix F Certified Fingerprint Sensor 2
LECHNER AND LaVECK MEDIUMfor the clonal growth of normal human bronchial and other epithelial cells. Requires additives. With glutamine.Recommended storage condition:oC to 8oCIntended use(s):research...
...
... System Modules Highly ... directional cloning into advanced ... vectors For ... Cloning Systems, the modules ...
... antibody-based immunoaffinity matrices, such as Sepharose ... without compromising antigen-antibody reactivity, the need ... has precluded the application of these ... CHEMICON Re-Blot Plus Western Blot Mild ...
Biology Products: