Analysis of urine and seawater samples by ultrasonic nebulization with a high resolution ICP spectrometer

Tran T. Nham Varian Australia Pty Ltd Mulgrave, Victoria, Australia

Introduction

Because of the low detection limit capability of the ultrasonic nebulizer when combined with an ICP spectrometer, it has been widely used to determine trace and ultratrace levels in water samples1. The purpose of this work was to evaluate the ability of the ultrasonic nebulizer to handle samples with significant amounts of dissolved solids, such as urine and seawater.

Experimental

Instrumentation

All measurements were carried out on Varian Liberty 200 ICP spectrometer. The ultrasonic nebulizer was a Cetac U-5000 AT.

The operating conditions of the instrument system are listed in table 1.

For optimum operation, the use of a higher auxiliary Ar flow of 2.25 L/min lifts the plasma away from the injector tube to minimize the possibility of salt build up at the tip of the injector tube. The use of a 1.8 mm I.D. injector tube can also minimize salt build up at the injector tube, especially while running seawater samples.

Lowering the torch by 1 mm (compared with the normal position) can also minimize salting up at the injector tip. The distance from the injector tip to the top of the intermediate tube is normally set to 3 mm with a demountable torch.

Reagents

All chemicals used were analytical grade. The multielement stock solution was made up from single element standard stock solutions (Spectrosol, BDH Chemicals). The multi-element working standard was prepared by dilution of the multi-element stock solution and contained 0.14 M HNO3.

Matrix matched standards and blank were employed in this work. As an alternative, the method of internal standardization could also be used, to compensate for differences in transport efficiences between different sample types.

For Seronorm urine analysis, the standard and blank were prepared in synthetic urine solution. Synthetic urine solution2 contains 14.1 g NaCl, 2.8 g KCl, 17.3 g urea, 1.9 mL of 25% v/v ammonia solution, 0.6 g CaCl2 and 0.43 g MgSO4, made up to 1 L with 0.02 M HCl solution.

For seawater analysis, the standard and blank were prepared in artificial seawater solution. Artificial seawater solution3 contains 25.4 g NaCl, 5.083 g MgCl2, 1.104 g CaCl2, 0.722 g KCl, 0.026 g H3BO3 and 0.203 g SrCl2, made up to 1 L with distilled de-ionized water.

Results

1. Long term stability

Long term stability was evaluated by performing a calibration on a multi-element standard and a blank, both being made up in 1% w/v NaCl solution. The standard was then measured as a sample continuously for a period of an hour. After an hour, the reproducibility ranged between 0.8 to 2 %RSD for the measurements plotted in figure 1.

An additional stability test was performed on a seawater solution, as shown in figure 2. The reproducibility over an hour ranged from 2 to 5 %RSD. It was noted that after continuously aspirating the seawater solution for over an hour, the injector tube gradually showed signs of blockage due to salt deposition. Further experimentation confirmed that rinsing between samples is desirable to minimize any potential clogging of the tube and improve the long term stability of the analysis.

2. Urine analysis

Urine samples can be analyzed directly by ultrasonic nebulization. Analysis was performed on a Seronorm urine sample. The standard and blank were prepared in a synthetic urine solution. The results of the analysis and the calculated detection limits (2 sigma) are listed in table 2. The measured values are in good agreement with the certified values. The precision of the measurements ranged from 0.5 to 3 %RSD.

Wavelength scans of standard, blank and urine sample on Cd and Cr are illustrated in figure 3.

3. Seawater analysis

This study has shown that seawater samples can be analyzed directly following ultrasonic nebulization. To demonstrate the accuracy of the method, a known amount of analytes were added to the sample and the recovery calculated.

The dissolved solids content of seawater can be up to 3.5% w/v. Because of the relatively high salt content in seawater, it is recommended that longer washout times are required between samples to clean out any possible salt deposition in the injector tube and improve long term stability. The use of an auxiliary rinse, which delivers rinsing agent directly onto the transducer surface, can speed up the washout process and reduce the analysis time.

The results of the seawater analysis, the percentage spike recovery and the calculated detection limit (2 sigma) are given in table 3. The percentage recoveries ranged from 90103%. The precision of the measurements ranged from 0.5 to 3 %RSD.

Wavelength scans of 50 μg/L of Pb and Ni in seawater sample are shown in figure 4.

Summary

The Liberty ICP and ultrasonic nebulizer combination can be used for the direct determination of trace elements in samples such as urine and seawater. The measured values are in good agreement with the certified values and the percentage recoveries ranged from 90 to 103%.

References

1. ICP At Work No. 8, Nham T. T., Varian, Mulgrave, Victoria, 1992.

2. Determination of silicon, calcium, magnesium and phosphorus in urine using inductively coupled plasma emission spectrometry and a matrix matching technique, Tanaka T., Hayashi Y., Clin. Chim. Acta, 1986, 156, 109-114.

3. Water analysis by Atomic Absorption, C. R. Parker, Varian Techtron, Springvale, Victoria, 1972.


'"/>

Source:


Page: All 1 2 3 4

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
(Date:1/22/2015)... Crystal Diagnostics (CDx) Xpress System, a rapid ... AOAC-PTM Certifications for the six non-O157 Shiga toxin-producing E. coli ... to as STEC or the “Big-6”) as well as Escherichia ... per 325 g of raw ground beef and raw beef ...
(Date:1/22/2015)... Dr. Greg Leyer of UAS Labs recently was invited ... seminar on probiotics in San Diego, CA. , The ... for health care professionals. This year’s pre-conference seminar was held ... health. Dr. Leyer spoke about the emerging topics and science ...
(Date:1/22/2015)... Denville, NJ (PRWEB) January 22, 2015 ... shearing technologies such as the Bioruptor® and complete ... semi-automation system for chromatin immunoprecipitation, alleviating the need ... reagents and controls needed for ChIP of histones ...
(Date:1/22/2015)... 2015  Varian Medical Systems (NYSE: VAR ), world ... its commitment to sustainability with inclusion on a prestigious list ... ranked healthcare equipment company among the Corporate Knights Global 100 ... Forum at Davos, Switzerland . ...
Breaking Biology Technology:Crystal Diagnostics Awarded AOAC-PTM Accreditation for the Rapid Detection of “Big 6” E.coli Food Pathogens 2Diagenode Launches Unique Semi-Automated Solution for Chromatin Immunoprecipitation 2Varian Honored Among World's 100 Most Sustainable Corporations 2
... the American Society of Hematology Annual Meeting , ... Corporation (Nasdaq: ACEL ) today announced that American ... (2008 112: Abstract 4205 & 2008 112: Abstract 4010) ... in chronic lymphocytic leukemia (CLL) and acute myeloblastic leukemia (AML) ...
... (OTC Bulletin Board: VSPC) announced today that it has planted 1.2 ... the IPA factory in Guangdong province in China. Giant King ... seedlings rather than seeds. In tropical and subtropical areas such ... up to 156 tons of grass per acre per year. , ...
... Year-End Results Show 2008 Revenue Growth Of 23% , ... In a news release today, Microbix announced results for the ... core business are continuing to fund Microbix Biosystems Inc.,s (TSX:MBX) ... in the short and long term, according to William J. ...
Cached Biology Technology:Alfacell's ONCONASE(R) and R-Amphinase Show Anti-Tumor Activity in Chronic Lymphocytic Leukemia and Acute Myeloblastic Leukemia Cells 2Alfacell's ONCONASE(R) and R-Amphinase Show Anti-Tumor Activity in Chronic Lymphocytic Leukemia and Acute Myeloblastic Leukemia Cells 3Alfacell's ONCONASE(R) and R-Amphinase Show Anti-Tumor Activity in Chronic Lymphocytic Leukemia and Acute Myeloblastic Leukemia Cells 4Alfacell's ONCONASE(R) and R-Amphinase Show Anti-Tumor Activity in Chronic Lymphocytic Leukemia and Acute Myeloblastic Leukemia Cells 5VIASPACE Plants Biofuel & Animal Feed Grass and Signs Land Lease in China 2VIASPACE Plants Biofuel & Animal Feed Grass and Signs Land Lease in China 3Microbix Uses Core Business Cashflow To Fund Late-Stage Pipeline To Build Shareholder Value In 2009 2Microbix Uses Core Business Cashflow To Fund Late-Stage Pipeline To Build Shareholder Value In 2009 3Microbix Uses Core Business Cashflow To Fund Late-Stage Pipeline To Build Shareholder Value In 2009 4
(Date:1/22/2015)... , Jan. 20, 2015 NXT-ID, Inc. (NASDAQ: NXTD ... commerce market, reports on the recent success of the Wocket™ smart wallet ... Wocket smart wallet was named as one of the "11 Hot ... "5 Best Products Launched At CES So Far" by Newseveryday.com and "The ...
(Date:1/22/2015)... Jan. 22, 2015 Research and Markets ... the "Global Wearable Technologies Market and Applications, ... Trends and Forecast 2014-2020" report to their ... smart devices that can be worn on the ...
(Date:12/22/2014)... 2014  NXT-ID, Inc. (NASDAQ: NXTD and NXTDW) ... the growing mobile commerce market, announces it has retained famous ... the 2015 International CES debut of the Wocket™ biometric smart ... at the NXT-ID booth January 6th and 7 th , ...
Breaking Biology News(10 mins):CES Response for NXT-ID's Wocket Smart Wallet Kicks Off 2015 2CES Response for NXT-ID's Wocket Smart Wallet Kicks Off 2015 3Global Wearable Technologies Market and Applications, Opportunities, Industry Analysis, Segmentation, Size, Growth, Share, Overview, Trends and Forecast 2014-2020 2NXT-ID Hires Master Pickpocket and Security Consultant Apollo Robbins for Wocket Smart Wallet CES 2015 Debut 2NXT-ID Hires Master Pickpocket and Security Consultant Apollo Robbins for Wocket Smart Wallet CES 2015 Debut 3NXT-ID Hires Master Pickpocket and Security Consultant Apollo Robbins for Wocket Smart Wallet CES 2015 Debut 4
... carbon dioxide than it can bind. Local variations have increased the ... a study of how carbon dioxide flows between the water of ... the University of Gothenburg, Sweden. "The capacity of the ... acidity of the water has changed in recent centuries. In the ...
... Several NASA satellites have been following Hurricane Jova since ... "winked" at them. Satellite imagery from NASA,s Aqua ... only sometimes visible and other times appeared cloud covered, ... as NOAA,s GOES-11 satellite captured Jova,s "winks." In ...
... back. What many might consider a "dinosaur" may be ... said Dean Riechers, University of Illinois associate professor of ... 2,4-D or other auxin herbicides," Riechers said. "But herbicide ... bring it back. That illustrates how severe this problem ...
Cached Biology News:The Baltic Sea contributes carbon dioxide to the atmosphere 2NASA gets an icy cold wink from Hurricane Jova's eye 2Which direction are herbicides heading? 2Which direction are herbicides heading? 3
... may be carried out using this high ... System control and data processing is provided ... The elegant, powerful and very easy to use ... and 2000., The supplied automatic sample injector has ...
This CLS number is a new product number, created to easily match Cornings product number. If showing no availability yet, please order under the old Sigma-Aldrich number (F1676) or contact customer s...
... a new product number, created to ... showing no availability yet, please order ... or contact customer service for assistance. ... 11.25 6.40 in. Mfr ...
A convenience pack containing one each of the most popular sizes of Cat. No. 4980 flask. Designed for the low volume user, a case contains one each of five sizes: 50mL, 125mL, 250mL, 500mL, and 1L....
Biology Products: