Analysis of Fuel Oxygenates by U.S. EPA Method 8260B Using Headspace Trap with GC/MS

Author:
Heidi Grecsek

PerkinElmer Life and Analytical Sciences
710 Bridgeport Avenue
Shelton, CT 06484

Introduction

The Clean Air Act amendments of 1990 require that oxygenated compounds be added to gasoline to produce fuel with increased oxygen content. Oxygenated fuel is desirable because it burns more completely, thereby reducing tailpipe emissions. The most commonly used fuel oxygenate additive is methyl tertiary butyl ether (MTBE).

The transportation, transfer and especially storage of oxygenated fuel in underground tanks present a risk to the environment. Leaks and spillage can contaminate ground, surface and waste waters, representing a significant threat to drinking-water supplies. Consequently, there is an increasing need for the analysis of MTBE in water by many state and federal agencies. In addition, there are new regulations requiring that MTBE be phased out, so the detection and quantification of other fuel oxygenates must also be considered.1

This field application report describes the use of a PerkinElmer TurboMatrix HS-110 Trap and Clarus 500 GC/MS optimized for low-level determination of fuel oxygenates by U.S. Environmental Protection Agency (EPA) Method 8260B. The headspace trap uses heat to extract (partition) the compounds out of the water into an equilibrium headspace, instead of purging. This transports a significant and highly reproducible fraction of the analytes from the sample onto a thermal desorption trap, where they are retained and focused on a multi-bed adsorbent packing before injection onto the GC column. The headspace trap offers operators low detection limits, excellent linearity and high throughput for fast sample turnaround.

Results

The heated headspace-trap transfer line is connected to the GC injector port with an injector adapter (PerkinElmer part number B0505266). The headspace-trap fused-silica transfer line (PerkinElmer part number N9301357) is connected through the injector port so approximately 20 cm resides in the oven. This is connected to the GC column by using a universal capillary-column connector (PerkinElmer part number N9302149). The rest of the method setup parameters are found in Table 1.

Following U.S. EPA 8260B method requirements, the system was tuned using 25 μg/L (ppb) of BFB (4-bromofluorobenzene). There are nine EPA BFB ion ratios that were checked for compliance (Figure 1). 19 compounds were analyzed and examined for detection limits, linearity and throughput. Linearity was checked using a five-level calibration curve in DI water. The range was 0.5-100 μg/L. Ethanol and tert-butyl alcohol were calibrated from 5-200 μg/L. Table 2 displays the list of components analyzed, demonstrates the quantitative precision and linearity of the method. The %RSD (based on relative response factors) for all analytes was within the 15% U.S. EPA 8260B method requirement mandated in environmental labs.

All 19 compounds were identified at all 5 calibration levels and a chromatogram of the 25 μg/L (mid-level calibration) standard is shown in Figure 2. All standards were prepared with appropriate Method 8260B internal standards and surrogates. Inserted into Figure 2 is an extracted ion chromatogram of 5 μg/L of ethanol (the lowest level standard), which demonstrates excellent signal to noise at this concentration. Ethanol is challenging at low concentration levels, when using traditional purge-and-trap technology. Fuel oxygenates such as ethanol are highly soluble in water and difficult to purge. This application illustrates the low detection limits achieved for these difficult components using the new headspace-trap technology.

The TurboMatrix Headspace Trap system offers a standard 12-position thermostatted chamber. This allows multiple samples to be heating at the same time so, as soon as the GC becomes ready, the next sample can be injected. The instrument automatically calculates PII (period from injection to injection) time, making this 12-position overlapped thermostatting the most efficient use of valuable time while running multiple samples.

The TurboMatrix Headspace Trap system uses elevated temperatures to partition the analytes of interest into the vapor phase for analysis. This offers a unique advantage over traditional purge-and-trap concentrators. The headspace trap only transfers the volatile analytes not the solid or liquid phase to the GC. In addition, the system automatically bakes out the flow path between injections. This results in very low carryover and minimum downtime, increasing laboratory productivity and eliminating the contamination problems that many environmental laboratories contend with when using purge-and-trap concentrators. Figure 3 is a demonstration of a typical environmental sample analyzed using a TurboMatrix Headspace Trap. This concentration level would typically cause significant carryover problems with purge-and-trap concentrators. The TurboMatrix Headspace Trap demonstrated less than 0.2% carryover on the next injection.

Conclusions

U.S. EPA 8260B is a challenging method used in the majority of U.S. and many international environmental laboratories for the analysis of VOCs. It is a complex method that requires multiple quality criteria be passed throughout the analysis. Fuel oxygenates have traditionally been difficult to analyze using purge-and-trap concentrators. As demonstrated here, the PerkinElmer Clarus 500 GC/MS configured with a TurboMatrix HS-110 Trap meets all the 8260B method requirements for the analysis of fuel oxygenates. In addition, the TurboMatrix HS Trap offers the following advantages over a conventional purge-and-trap concentrator:

Easy and convenient sample preparation
Good response with poor purging compounds
Increased sample throughput by overlapped thermostatting
Eliminates many of the carryover contamination problems seen with the purge-and-trap concentrators
No cross-contamination of samples from using the same purge vessel

References

1. http://www.arb.ca.gov/fuels/gasoline/ cbgmtbe.htm

2. Test Methods for Evaluating Solid Waste; U.S. Environmental Protection Agency, National Technical Information Service: Washington, DC, Rev. 2, Vol. B. (1996).


'"/>

Source:


Page: All 1 2 3 4

Related biology technology :

1. A Positive Selection Assay for Mutation Analysis in Big Blue Animals
2. Epitope-Tagging Vectors for Functional Analysis in Yeast
3. Map and Link Human Genetic Disorders with SSLP Analysis
4. CastAway Precast Gels for Rapid Automated DNA Sequence Analysis
5. Single-cell RT-PCR Analysis of Paramecium Primaurelia with the Eppendorf cMaster RTplusPCR System
6. Structural Analysis of Glycosylated Peptides in Complex Mixtures with Ion Trap MSn
7. Use of Transcriptor Reverse Transcriptase in Microarray Analysis
8. Increased Sensitivity in Microarray Analysis
9. Gene-Expression Analysis of TP, DPD, and TS Using the LightCycler mRNA Quantification KitsPLUS
10. Agilent 2100 Bioanalyzer Automated Analysis System
11. Analysis of Microarray Data
Post Your Comments:
*Name:
*Comment:
*Email:
TAG: Analysis Fuel Oxygenates EPA Method Using Headspace Trap with

(Date:12/19/2014)... 19, 2014 Bina Technologies, Inc. (Bina, ... they have been acquired by Roche (SIX: RO, ROG; ... held company that provides a big data platform for ... data for the academic and translational research markets.  Bina ... will continue to focus on development of their innovative ...
(Date:12/19/2014)... Altamonte Springs, FL (PRWEB) December 18, 2014 ... one of the nation’s leading specialty pharmacies, has ... who serve as clinical pharmacy consultants for prescriber ... disease therapy expertise. , “We have a ... disease therapies. For example, those specializing in hepatitis ...
(Date:12/19/2014)... (PRWEB) December 18, 2014 Ed Bilsky, ... of the Center for Excellence in the Neurosciences (CEN), ... England, has been selected as a member of the ... the first neuroscientist from Maine to receive this honor. ... organization supported entirely by the Dana Foundation. Its ...
(Date:12/19/2014)... contrast to traditional LED lights where the spectrum is ... LED technology to incorporate their spectra in a single ... in low proximity installations. The Valoya L-series offers high ... is up to four times longer than commonly used ... clearly higher when compared to traditional fluorescent or LED ...
Breaking Biology Technology:Bina Technologies is acquired by Roche. 2Bina Technologies is acquired by Roche. 3BioPlus Specialty Pharmacy Expands ‘Regional Pharmaceutical Care Liaison’ Team 2BioPlus Specialty Pharmacy Expands ‘Regional Pharmaceutical Care Liaison’ Team 3UNE's Ed Bilsky Named Member of Dana Alliance for Brain Initiatives 2UNE's Ed Bilsky Named Member of Dana Alliance for Brain Initiatives 3
... PCOP ), an innovator in the discovery and ... company plans to,release its financial results for the first ... on Wednesday, May 7, 2008. The company will,conduct a ... the same day., Dr. Joseph A. Mollica, Chairman ...
... find a noninvasive method to determine the amount of oxygen ... in a tumor have been linked to a poor outcome. ... surrounding organs and tissues at the same time has been ... Krishna and colleagues, at the National Institutes of Health, Bethesda. ...
... April 22 , HIGHLIGHTS: 2008 Results (all percentage comparisons are to comparable ... Record: Q1 2008 sales reached a new ... four business units also achieving new quarterly sales records ... Organic sales growth of 6.5% was ...
Cached Biology Technology:Pharmacopeia to Host First Quarter 2008 Financial Results Webcast and Conference Call 2Pharmacopeia to Host First Quarter 2008 Financial Results Webcast and Conference Call 3Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 2Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 3Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 4Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 5Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 6Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 7Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 8Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 9Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 10Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 11Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 12Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 13Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 14Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 15Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 16Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 17Sigma-Aldrich (Nasdaq: SIAL) Raises 2008 Diluted EPS Forecast by $.05 to $2.57-$2.67. Reports Q1 2008 Diluted EPS Up 14.3%; Sales Gain of 14.9%. 18
(Date:12/11/2014)... Dec. 09, 2014 Research and Markets ... ( http://www.researchandmarkets.com/research/twhjll/biometrics_market ) has announced the ... APAC Region 2015-2019"  report to their offering. ... market is advances in technology. With continuous advances ... solutions to the latest standard that meets the ...
(Date:12/11/2014)... 2014  That blood pressure plays a role in human ... the medical term for high blood pressure – was first ... inflatable cuff that,s used in measuring blood pressure was invented ... new about hypertension, its triggers and its effects. In fact, ... and the best ways to treat it. For ...
(Date:12/10/2014)... You,ve been here before: you desperately need to sign into ... key or the answer to your secret question. What,s your ... Today, Hoyos Labs , a digital infrastructure security ... an end to the frustration that comes with usernames, passwords ... a user,s smartphone to acquire his or her biometrics to ...
Breaking Biology News(10 mins):Biometrics Market in the APAC Region 2015-2019: Key Vendors are 3M Cogent, NEC, Safran and Suprema 2Research points to need for new approaches to treatment of high blood pressure 2Research points to need for new approaches to treatment of high blood pressure 3Research points to need for new approaches to treatment of high blood pressure 4The Password is Finally Dead: Launch of 1U Mobile App Eliminates Need for All Usernames and Passwords 2The Password is Finally Dead: Launch of 1U Mobile App Eliminates Need for All Usernames and Passwords 3
... in humans it is detected in approximately 0.5% ... a kidney abnormality known as hydronephrosis. Hydronephrosis arises because ... bladder is impeded. By studying kidney development in mice, ... Children, Toronto, have identified a new cellular mechanism underlying ...
... BOSTON Congenital heart diseases affect approximately one in ... birth defect and the number-one cause of pediatric deaths. ... drug rapamycin can reverse cardiac muscle damage in a mouse ... the first possible medical treatment for this rare condition, but ...
... choices an individual makes about environmental issues are affected ... network. Michigan State University scientists are studying how to ... sustainability science. "Instead of trying to communicate with ... using the structure of social networks to spread information," ...
Cached Biology News:Study shows rapamycin reverses myocardial defects in mouse model of LEOPARD syndrome 2Study shows rapamycin reverses myocardial defects in mouse model of LEOPARD syndrome 3Sustainability solutions need the power of networks 2
Extremely high efficiency packaging extracts useful for transfecting our EpiFOS™ or CopyControl™ Fosmids, or any DNA vector with cos sites. A tremendous value when conpared to the competition...
... Workstation for PCR set-up Easy ... to hold 0.2 ml PCR tubes, strips or Nunc ... handling systems Removable tray can be fitted ... V-bottom 0.2 ml tube block thermal cycler formats of ...
... in DMEM and harvested at the log phase ... antigens in their native forms, cells were fixed ... a 12-well (5 mm) adhesive coated slide, with ... attachment and to minimize background staining. Each ...
...
Biology Products: