This Application Note describes how Gene Logic, Inc., has improved its
ability to analyze the quality of total RNA and fragmented cRNA by using
the Agilent 2100 bioanalyzer. The Eukaryote Total RNA Nano and Pico
assays are currently being utilized to examine the 28S/18S ratio in the
total RNA as well as detection of possible degradation. The mRNA Smear
Nano assay is used to confirm that a high percentage of the cRNA sample
is fragmented to lengths in the optimal range necessary to be hybridized
onto Affymetrix microarrays.
Gene Logic Inc. is a leading
provider of integrated genomics
information and bioinformatics
products and services related to
gene activity in human disease
and toxicity that enable global
pharmaceutical and biotechnology
companies to optimize the time,
risk and cost of drug discovery
and drug development. In order to
do this, the quality control methods
must ensure that only the
highest quality samples proceed
through our high-throughput
process, which in turn help to
reduce the downstream costs of
cDNA synthesis, In-Vitro Transcription
and, finally, hybridization
onto microarrays. For qualitative
purposes, many laboratories continue
to analyze their RNA samples
using MOPS gels with gel
electrophoresis. However, an
internet search of Microarray
facilities around the country
shows an increase in the usage of
the Agilent bioanalyzer.
When using gel electrophoresis to
examine the quality of samples,
pre-cast MOPS gels were used.
Every now and then a problem
with the gels was encountered,
either user related or gel related.
Some general problems occurring
were sample migration, mold
spots in the pre-cast gels and
staining difficulties. The key issue
associated with the RNA quality
control gel was that at times the 18S and/or 28S bands of the total
RNA were not clearly visible on
the gels making the interpretation
difficult. When running fragmented
cRNA samples on MOPS gels it
was often difficult to distinguish
whether or not the samples were
allowed a sufficient amount of
time to fragment because of poor
resolution. Problems with MOPS
gels would lead to re-running of
the samples on new gels, therefore
wasting valuable time and money.
The samples that should have
failed would be caught by quality
control measures later on in the
process, but would ultimately cost
the company money on reagents
in the steps leading up to that.
The Agilent 2100 bioanalyzer combined
with the RNA 6000 Nano
LabChip Kit provides key advantages
to running MOPS gels. The
Agilent bioanalyzer data is easy to
interpret and the chip runs on the
bioanalyzer are less time consuming.
It can take up to an hour and
half to prepare the samples, run a
gel, stain the gel, and take a picture.
The whole Agilent process
(sample preparation to saving the
file) can be done in half that time.
There is no improper staining and
no shortened or lengthened run
times. Also, because each chip has
the capacity to run 12 samples, at
approximately 30 minutes per
chip, this makes the Agilent 2100
bioanalyzer a useful tool for high
throughput screenings, especially
with multiple bioanalyzers.
The goal was to find a new assay
to be used to evaluate the quality
of the total RNA and fragmented
cRNA. This was needed because
there was occasional inconsistency
with the use of MOPS gels. The
qualitative analyses should be as
definitive and dependable as possible.
Problems associated with
MOPS gels in a high throughput
environment may be due to user
error. Common mistakes seen in
the laboratory are: accidentally
piercing the gel while loading, not
loading enough of sample into the
well, improperly securing the gel
to the gel rig, not running the gel
on the appropriate voltage/time or
not staining the gel properly.
These errors are reflected in the
gel images below (Figures 1-6).
These problems would be eliminated
using the Agilent 2100 bioanalyzer.
MOPS gels may be suitable to get
an idea of the quality of samples,
but the gels do not always clearly
reveal a detailed image of the sample.
It is important that the
genomics information GeneLogic,
Inc. provides to its customers is of
the highest quality. Therefore,
there needs to be an accurate, reliable
way of assessing the quality
of the samples.
Agilent 2100 bioanalyzer vs MOPS
gels for Total RNA
The Agilent 2100 bioanalyzer was
tested as a tool for quality control
for total RNA and compared
against the MOPS gels for the
same samples over a series of
weeks. During this time a grading
system was derived for the sample
data using the Eukaryote Total
RNA Nano assay with the bioanalyzer.
It was determined that the
samples were either given a
pass, fail, or Pass needs
additional QC rating. Pass-Needs
additional QC would essentially mean that the sample was
passed, but that it would from
then on be closely monitored
for its quality in further steps.
Additional Quality Control
assays would be employed to
further evaluate the quality of
these samples. For the MOPS
gels a system was already in
place to either pass or fail a
sample. Because of the clarity
of the electropherograms produced
using the bioanalyzer,
distinguishing between the
high quality samples and poor
quality samples became easier.
For example, when samples
were analyzed using both the
bioanalyzer and the MOPS gels,
some of them that were failed
using the Agilent software
were passed using gel electrophoresis,
and vice versa.
Some samples that were
passed using the bioanalyzer
and failed on the MOPS gel are
shown in Figures 7 and 8.
Based upon the electropherograms,
the acceptable quality
of the samples is clearly determined,
yet on the MOPS gel
images a ribosomal band is
missing from each.
Just as the MOPS gel is capable
of giving false negatives, it is
also capable of giving false
positives. Some examples of
samples that were passed on
the MOPS gel but failed on the
bioanalyzer are shown in Figures
9 and 10. (Note: Samples
that were passed using the
MOPS gels were failed by QC
measures further along in the
process)
Based upon these observations,
there are clearly more
distinguishable results on the
electropherograms produced using the bioanalyzer then there
are in the MOPS gel images. There
is a lot more room for human and
mechanical error using the MOPS
gels, which may contribute to the
apparent differences.
Agilent 2100 bioanalyzer vs MOPS
gels for fragmented cRNA
When analyzing fragmented cRNA,
what is examined is that the sample
has been allowed to heat long
enough and is fully fragmented to
a desirable level. The size of the
fragments can be estimated using
the ladder on the MOPS gel, but
this is not conclusive and determination
of fragmentation is very
subjective due to the poor resolution
of low molecular weight components
(refer to Figures 1-3). The
mRNA Smear Nano assay used
with the bioanalyzer is capable of
showing the desired fragment
length range and percentage of the
total area for each sample. This is
an extremely valuable tool
because it takes the subjectivity
away from the determinations.
The software allows the user to
isolate the range of fragment
lengths necessary for proper
hybridization onto microarrays,
and gives the user the percent of
the total area that the range covers.
These tools are utilized to
ensure that the samples have been
fully fragmented to the proper size
range. If the samples do not have
a high enough percentage of total
area in the proper size range then
they are heated for an additional
amount of time to allow adequate
fragmentation. Examples of varying
levels of fragmentation are
shown below (Figures 11-15).
The mRNA Smear Nano assay is
clearly a useful means for examining
the fragmented cRNA samples.
In a MOPS gel image it is not
always easy to tell whether or
not a sample is fragmented to
sufficient lengths necessary for
hybridization. Samples that may
need to be heated longer may not
be recognized, and their hybridization
onto the arrays may be affected.
By using the Agilent 2100 bioanalyzer
it is easy to see which samples
are not fully fragmented.
Conclusion
The Agilent 2100 bioanalyzer and
the RNA 6000 LabChip Kit were
used to analyze the quality of the
total RNA and fragmented cRNA.
The Eukaryote Total RNA Nano
and mRNA Smear Nano assay
gave clear, distinct results. It is
easy to draw conclusions from the
electropherograms created using
the Agilent software. The whole
Agilent process is less time consuming,
less subjective, more
reliable, and is suitable for use in
a high throughput environment.
Money could also be saved on
reagents, microarrays and samples
because more concrete data is
being generated with the Agilent
2100 bioanalyzer. Use of the
Agilent 2100 bioanalyzer also
helps ensure Gene Logic, Inc.s
customers that the data they are
providing to them is of superior
quality.
'"/>Source:
Page: All 1 2 3 4 5 6 Related biology technology :1.
Using the Agilent 2100 bioanalyzer for
quality control of protein samples prior to
MS-analysis2.
Is Z factor the best assessment for the quality of cellular assays delivering higher content?3.
Evaluation of RNA isolated from human tissues: quality control for surgical and postmortem samples4.
Quality control of antibodies using the
2100 bioanalyzer and the Protein 200
Plus assay5.
Homogeneous, chemiluminescent kinase assay kits for assessing the functional activity of a broad range of serine-threonine and tyrosine kinases6.
Easy preparation of total RNA from cultured cells with TRIzol and
Eppendorf Phase Lock Gel